GlucoMen Day continuous glucose monitoring system: a screening for enzymatic and electrochemical interferents.

Fausto Lucarelli1, Francesco Ricci, Felice Caprio, Francesco Valgimigli, Cosimo Scuffi, Danila Moscone, Giuseppe Palleschi.   

Abstract

BACKGROUND: While most of the common drugs with the potential to interfere with continuous glucose monitoring (CGM) systems are accessible over the counter and can be assumed by CGM patients without medical supervision, many other chemicals are frequently used to treat critically ill patients. Continuous glucose monitoring reading accuracy may also be compromised in patients characterized by abnormally high concentrations of physiological interferents. In this article, 22 species selected from endogenous and exogenous chemicals were screened as possible interferents of GlucoMen®Day (GMD), the new microdialysis-based CGM system from A. Menarini Diagnostics.
METHOD: Interference testing was performed according to the EP7-A2 guideline (Clinical and Laboratory Standards Institute 2005). Interference was evaluated at two levels of glucose, with each interferent additionally tested at two concentrations. Furthermore, two configurations of the GMD disposable sensor kit--one designed for subcutaneous application, the other for direct intravascular CGM--were challenged with interferent-spiked serum and blood samples, respectively.
RESULTS: With the exception of dopamine (however, at very high, nonphysiological concentrations), no interference was observed for all the tested substances. Interestingly, none of the common electrochemical interferents (including ascorbic acid, acetaminophen, and salicylic acid, which represent the major specificity issue for the competing CGM systems) significantly affected the system's output.
CONCLUSIONS: These results provide clear insights into the advantages offered by the use of a microdialysis-based CGM system that additionally relies on the detection of hydrogen peroxide at low operating potential. GlucoMen Day may become the CGM system of choice for those patients who require either regular administration of drugs or their glycemia to be tightly controlled in the intensive care unit or similar environments.
© 2012 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23063044      PMCID: PMC3570852          DOI: 10.1177/193229681200600522

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  13 in total

1.  Evaluating the clinical accuracy of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Authors:  Francesco Valgimigli; Fausto Lucarelli; Cosimo Scuffi; Sara Morandi; Iolanda Sposato
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  A continuous glucose sensor based on wired enzyme technology -- results from a 3-day trial in patients with type 1 diabetes.

Authors:  Ben Feldman; Ronald Brazg; Sherwyn Schwartz; Richard Weinstein
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

Review 3.  In vitro, in vivo and post explantation testing of glucose-detecting biosensors: current methods and recommendations.

Authors:  Heidi E Koschwanez; William M Reichert
Journal:  Biomaterials       Date:  2007-04-19       Impact factor: 12.479

Review 4.  Electrochemical glucose biosensors.

Authors:  Joseph Wang
Journal:  Chem Rev       Date:  2007-12-23       Impact factor: 60.622

5.  The chemistry of commercial continuous glucose monitors.

Authors:  Geoffrey McGarraugh
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

6.  Intensive insulin therapy in the intensive care unit: assessment by continuous glucose monitoring.

Authors:  Christophe De Block; Begoña Manuel-Y-Keenoy; Luc Van Gaal; Peter Rogiers
Journal:  Diabetes Care       Date:  2006-08       Impact factor: 19.112

7.  Lack of specific renal haemodynamic effects of different doses of dopamine after infrarenal aortic surgery.

Authors:  A R Girbes; A G Lieverse; A J Smit; D J van Veldhuisen; J H Zwaveling; S Meijer; W D Reitsma
Journal:  Br J Anaesth       Date:  1996-12       Impact factor: 9.166

8.  Measurement of oxidized/reduced glutathione ratio.

Authors:  Joshua B Owen; D Allan Butterfield
Journal:  Methods Mol Biol       Date:  2010

9.  Toward continuous glucose monitoring with planar modified biosensors and microdialysis. Study of temperature, oxygen dependence and in vivo experiment.

Authors:  Francesco Ricci; Felice Caprio; Alessandro Poscia; Francesco Valgimigli; Dimitri Messeri; Elena Lepori; Giorgio Dall'Oglio; Giuseppe Palleschi; Danila Moscone
Journal:  Biosens Bioelectron       Date:  2006-09-26       Impact factor: 10.618

10.  Novel planar glucose biosensors for continuous monitoring use.

Authors:  F Ricci; D Moscone; C S Tuta; G Palleschi; A Amine; A Poscia; F Valgimigli; D Messeri
Journal:  Biosens Bioelectron       Date:  2005-04-15       Impact factor: 10.618

View more
  13 in total

Review 1.  Wireless and battery-free platforms for collection of biosignals.

Authors:  Tucker Stuart; Le Cai; Alex Burton; Philipp Gutruf
Journal:  Biosens Bioelectron       Date:  2021-01-23       Impact factor: 10.618

2.  Minimizing the impact of time lag variability on accuracy evaluation of continuous glucose monitoring systems.

Authors:  Cosimo Scuffi; Fausto Lucarelli; Francesco Valgimigli
Journal:  J Diabetes Sci Technol       Date:  2012-11-01

3.  Interference testing: why following standards is not always the right thing to do.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

Review 4.  Pitfalls in hemoglobin A1c measurement: when results may be misleading.

Authors:  Michael S Radin
Journal:  J Gen Intern Med       Date:  2013-09-04       Impact factor: 5.128

5.  Interstitium versus Blood Equilibrium in Glucose Concentration and its Impact on Subcutaneous Continuous Glucose Monitoring Systems.

Authors:  Cosimo Scuffi
Journal:  Eur Endocrinol       Date:  2014-02-28

Review 6.  Microdialysis techniques and microdialysis-based patient-near diagnostics.

Authors:  Mark Thomas O'Connell; Jan Krejci
Journal:  Anal Bioanal Chem       Date:  2022-01-14       Impact factor: 4.142

Review 7.  Semi-Implantable Bioelectronics.

Authors:  Jiaru Fang; Shuang Huang; Fanmao Liu; Gen He; Xiangling Li; Xinshuo Huang; Hui-Jiuan Chen; Xi Xie
Journal:  Nanomicro Lett       Date:  2022-05-28

8.  Design of the micropump and mass-transfer compartment of a microfluidic system for regular nonenzymatic glucose measurement.

Authors:  Armita Najmi; Mohammad Said Saidi; Siamak Kazemzadeh Hannani
Journal:  Biotechnol Rep (Amst)       Date:  2022-03-24

9.  In Vitro Sugar Interference Testing With Amperometric Glucose Oxidase Sensors.

Authors:  Ryan Boehm; John Donovan; Disha Sheth; Andrew Durfor; Jason Roberts; Irada Isayeva
Journal:  J Diabetes Sci Technol       Date:  2018-08-03

Review 10.  Italian contributions to the development of continuous glucose monitoring sensors for diabetes management.

Authors:  Giovanni Sparacino; Mattia Zanon; Andrea Facchinetti; Chiara Zecchin; Alberto Maran; Claudio Cobelli
Journal:  Sensors (Basel)       Date:  2012-10-12       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.