Performance evaluation of a continuous glucose monitoring system under conditions similar to daily life.

Stefan Pleus1, Christina Schmid, Manuela Link, Eva Zschornack, Hans-Martin Klötzer, Cornelia Haug, Guido Freckmann.   

Abstract

BACKGROUND: This study aimed at evaluating and comparing the performance of a new generation of continuous glucose monitoring (CGM) system versus other CGM systems, under daily lifelike conditions.
METHODS: A total of 10 subjects (7 female) were enrolled in this study. Each subject wore two Dexcom G4™ CGM systems in parallel for the sensor lifetime specified by the manufacturer (7 days) to allow assessment of sensor-to-sensor precision. Capillary blood glucose (BG) measurements were performed at least once per hour during daytime and once at night. Glucose excursions were induced on two occasions. Performance was assessed by calculating the mean absolute relative difference (MARD) between CGM readings and paired capillary BG readings and precision absolute relative difference (PARD), i.e., differences between paired CGM readings.
RESULTS: Overall aggregate MARD was 11.0% (n = 2392). Aggregate MARD for BG <70 mg/dl was 13.7%; for BG between 70 and 180 mg/dl, MARD was 11.4%; and for BG >180 mg/dl, MARD was 8.5%. Aggregate PARD was 7.3%, improving from 11.6% on day 1 to 5.2% on day 7.
CONCLUSIONS: The Dexcom G4 CGM system showed good overall MARD compared with results reported for other commercially available CGM systems. In the hypoglycemic range, where CGM performance is often reported to be low, the Dexcom G4 CGM system achieved better MARD than that reported for other CGM systems in the hypoglycemic range. In the hyperglycemic range, the MARD was comparable to that reported for other CGM systems, whereas during induced glucose excursions, the MARD was similar or slightly worse than that reported for other CGM systems. Overall PARD was 7.3%, improving markedly with sensor life time.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23911164      PMCID: PMC3879747          DOI: 10.1177/193229681300700405

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  28 in total

1.  Evaluating the clinical accuracy of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Authors:  Francesco Valgimigli; Fausto Lucarelli; Cosimo Scuffi; Sara Morandi; Iolanda Sposato
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

3.  Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial.

Authors:  Satish Garg; Howard Zisser; Sherwyn Schwartz; Timothy Bailey; Roy Kaplan; Samuel Ellis; Lois Jovanovic
Journal:  Diabetes Care       Date:  2006-01       Impact factor: 19.112

4.  Analysis of time lags and other sources of error of the DexCom SEVEN continuous glucose monitor.

Authors:  Apurv Kamath; Aarthi Mahalingam; James Brauker
Journal:  Diabetes Technol Ther       Date:  2009-11       Impact factor: 6.118

5.  Evaluation of a novel continuous glucose measurement device in patients with diabetes mellitus across the glycemic range.

Authors:  Linda Morrow; Marcus Hompesch; Ann M Tideman; Jennifer Matson; Nancy Dunne; Scott Pardo; Joan L Parkes; Holly C Schachner; David A Simmons
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

6.  Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis.

Authors:  William L Clarke; Stacey Anderson; Leon Farhy; Marc Breton; Linda Gonder-Frederick; Daniel Cox; Boris Kovatchev
Journal:  Diabetes Care       Date:  2005-10       Impact factor: 19.112

7.  Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Authors:  W L Clarke; D Cox; L A Gonder-Frederick; W Carter; S L Pohl
Journal:  Diabetes Care       Date:  1987 Sep-Oct       Impact factor: 19.112

8.  Prevention of hypoglycemia by using low glucose suspend function in sensor-augmented pump therapy.

Authors:  Thomas Danne; Olga Kordonouri; Martin Holder; Holger Haberland; Sven Golembowski; Kerstin Remus; Sara Bläsig; Tanja Wadien; Susanne Zierow; Reinhard Hartmann; Andreas Thomas
Journal:  Diabetes Technol Ther       Date:  2011-08-09       Impact factor: 6.118

9.  Sensor-augmented pump therapy lowers HbA(1c) in suboptimally controlled Type 1 diabetes; a randomized controlled trial.

Authors:  J Hermanides; K Nørgaard; D Bruttomesso; C Mathieu; A Frid; C M Dayan; P Diem; C Fermon; I M E Wentholt; J B L Hoekstra; J H DeVries
Journal:  Diabet Med       Date:  2011-10       Impact factor: 4.359

10.  Comparison of accuracy and safety of the SEVEN and the Navigator continuous glucose monitoring systems.

Authors:  Satish K Garg; James Smith; Christie Beatson; Benita Lopez-Baca; Mary Voelmle; Peter A Gottlieb
Journal:  Diabetes Technol Ther       Date:  2009-02       Impact factor: 6.118

View more
  18 in total

1.  Extensive Assessment of Blood Glucose Monitoring During Postprandial Period and Its Impact on Closed-Loop Performance.

Authors:  Lyvia Biagi; Arthur Hirata Bertachi; Ignacio Conget; Carmen Quirós; Marga Giménez; F Javier Ampudia-Blasco; Paolo Rossetti; Jorge Bondia; Josep Vehí
Journal:  J Diabetes Sci Technol       Date:  2017-06-21

2.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

3.  Benefits and Limitations of MARD as a Performance Parameter for Continuous Glucose Monitoring in the Interstitial Space.

Authors:  Lutz Heinemann; Michael Schoemaker; Günther Schmelzeisen-Redecker; Rolf Hinzmann; Adham Kassab; Guido Freckmann; Florian Reiterer; Luigi Del Re
Journal:  J Diabetes Sci Technol       Date:  2019-06-19

4.  Analysis of the Accuracy and Performance of a Continuous Glucose Monitoring Sensor Prototype: An In-Silico Study Using the UVA/PADOVA Type 1 Diabetes Simulator.

Authors:  Marc D Breton; Rolf Hinzmann; Enrique Campos-Nañez; Susan Riddle; Michael Schoemaker; Guenther Schmelzeisen-Redeker
Journal:  J Diabetes Sci Technol       Date:  2016-12-13

5.  Use of microdialysis-based continuous glucose monitoring to drive real-time semi-closed-loop insulin infusion.

Authors:  Guido Freckmann; Nina Jendrike; Stefan Pleus; Harvey Buck; Steven Bousamra; Paul Galley; Ajay Thukral; Robin Wagner; Stefan Weinert; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2014-09-09

6.  Rate-of-Change Dependence of the Performance of Two CGM Systems During Induced Glucose Swings.

Authors:  Stefan Pleus; Michael Schoemaker; Karin Morgenstern; Günther Schmelzeisen-Redeker; Cornelia Haug; Manuela Link; Eva Zschornack; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2015-04-07

7.  A Personalized Week-to-Week Updating Algorithm to Improve Continuous Glucose Monitoring Performance.

Authors:  Stamatina Zavitsanou; Joon Bok Lee; Jordan E Pinsker; Mei Mei Church; Francis J Doyle; Eyal Dassau
Journal:  J Diabetes Sci Technol       Date:  2017-10-16

8.  Assessing sensor accuracy for non-adjunct use of continuous glucose monitoring.

Authors:  Boris P Kovatchev; Stephen D Patek; Edward Andrew Ortiz; Marc D Breton
Journal:  Diabetes Technol Ther       Date:  2014-12-01       Impact factor: 6.118

9.  Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm.

Authors:  Timothy S Bailey; Anna Chang; Mark Christiansen
Journal:  J Diabetes Sci Technol       Date:  2014-11-03

10.  The impact of continuous glucose monitoring on low interstitial glucose values and low blood glucose values assessed by point-of-care blood glucose meters: results of a crossover trial.

Authors:  Norbert Hermanns; Beatrix Schumann; Bernhard Kulzer; Thomas Haak
Journal:  J Diabetes Sci Technol       Date:  2014-02-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.