Periodic extraction of interstitial fluid from the site of subcutaneous insulin infusion for the measurement of glucose: a novel single-port technique for the treatment of type 1 diabetes patients.

Werner Regittnig1, Stefan Lindpointner, Stefan Korsatko, Dina Tutkur, Manfred Bodenlenz, Thomas R Pieber.   

Abstract

BACKGROUND: Treatment of type 1 diabetes patients could be simplified if the site of subcutaneous insulin infusion could also be used for the measurement of glucose. This study aimed to assess the agreement between blood glucose concentrations and glucose levels in the interstitial fluid (ISF) that is extracted from the insulin infusion site during periodic short-term interruptions of continuous subcutaneous insulin infusion (CSII). SUBJECTS AND METHODS: A perforated cannula (24 gauge) was inserted into subcutaneous adipose tissue of C-peptide-negative type 1 diabetes subjects (n=13) and used alternately to infuse rapid-acting insulin (100 U/mL) and to extract ISF glucose during a fasting period and after ingestion of a standard oral glucose load (75 g).
RESULTS: Although periodically interrupted for extracting glucose (every hour for approximately 10 min), insulin infusion with the cannula was adequate to achieve euglycemia during fasting and to restore euglycemia after glucose ingestion. Furthermore, the ISF-derived estimates of plasma glucose levels agreed well with plasma glucose concentrations. Correlation coefficient and median absolute relative difference values were found to be 0.95 and 8.0%, respectively. Error grid analysis showed 99.0% of all ISF glucose values within clinically acceptable Zones A and B (83.5% Zone A, 15.5% Zone B).
CONCLUSIONS: Results show that ISF glucose concentrations measured at the insulin infusion site during periodic short-term interruptions of CSII closely reflect blood glucose levels, thus suggesting that glucose monitoring and insulin delivery may be performed alternately at the same tissue site. A single-port device of this type could be used to simplify and improve glucose management in diabetes.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23126579      PMCID: PMC3540899          DOI: 10.1089/dia.2012.0173

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  31 in total

1.  Transdermal monitoring of glucose and other analytes using ultrasound.

Authors:  J Kost; S Mitragotri; R A Gabbay; M Pishko; R Langer
Journal:  Nat Med       Date:  2000-03       Impact factor: 53.440

2.  Evaluating the clinical accuracy of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Authors:  Francesco Valgimigli; Fausto Lucarelli; Cosimo Scuffi; Sara Morandi; Iolanda Sposato
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 3.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

4.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

5.  The chemistry of commercial continuous glucose monitors.

Authors:  Geoffrey McGarraugh
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

6.  Proposal for standardizing direct-reading biosensors for blood glucose.

Authors:  N Fogh-Andersen; P D'Orazio
Journal:  Clin Chem       Date:  1998-03       Impact factor: 8.327

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Analytical characterization of electrochemical biosensor test strips for measurement of glucose in low-volume interstitial fluid samples.

Authors:  M E Collison; P J Stout; T S Glushko; K N Pokela; D J Mullins-Hirte; J R Racchini; M A Walter; S P Mecca; J Rundquist; J J Allen; M E Hilgers; T B Hoegh
Journal:  Clin Chem       Date:  1999-09       Impact factor: 8.327

9.  FreeStyle Mini blood glucose results are accurate and suitable for use in glycemic clamp protocols.

Authors:  Ohad Cohen; Sigal Shaklai; Ephraim Gabis; Michael A Pani
Journal:  J Diabetes Sci Technol       Date:  2008-09

10.  Use of the site of subcutaneous insulin administration for the measurement of glucose in patients with type 1 diabetes.

Authors:  Stefan Lindpointner; Stefan Korsatko; Gerd Köhler; Hans Köhler; Roland Schaller; Ruthy Kaidar; Ofer Yodfat; Lukas Schaupp; Martin Ellmerer; Thomas R Pieber; Werner Regittnig
Journal:  Diabetes Care       Date:  2009-12-29       Impact factor: 17.152

View more
  5 in total

Review 1.  AP@home: The Artificial Pancreas Is Now at Home.

Authors:  Lutz Heinemann; Carsten Benesch; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2016-06-28

Review 2.  Moving Toward a Unified Platform for Insulin Delivery and Sensing of Inputs Relevant to an Artificial Pancreas.

Authors:  Anneke Graf; Sybil A McAuley; Catriona Sims; Johanna Ulloa; Alicia J Jenkins; Gayane Voskanyan; David N O'Neal
Journal:  J Diabetes Sci Technol       Date:  2016-12-13

Review 3.  Can glucose be monitored accurately at the site of subcutaneous insulin delivery?

Authors:  W Kenneth Ward; Jessica R Castle; Peter G Jacobs; Robert S Cargill
Journal:  J Diabetes Sci Technol       Date:  2014-02-18

4.  Measuring glucose at the site of insulin delivery with a redox-mediated sensor.

Authors:  Peter G Jacobs; Nichole S Tyler; Scott M Vanderwerf; Clara Mosquera-Lopez; Thomas Seidl; Robert Cargill; Deborah Branigan; Katrina Ramsey; Kristin Morris; Sheila Benware; W Kenneth Ward; Jessica R Castle
Journal:  Biosens Bioelectron       Date:  2020-04-29       Impact factor: 12.545

Review 5.  Insulin delivery and nocturnal glucose control in children and adolescents with type 1 diabetes.

Authors:  Martin Tauschmann; Roman Hovorka
Journal:  Expert Opin Drug Deliv       Date:  2017-08-18       Impact factor: 6.648

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.