Contrast-enhanced spectral mammography with a photon-counting detector.

Erik Fredenberg1, Magnus Hemmendorff, Björn Cederström, Magnus Aslund, Mats Danielsson.   

Abstract

PURPOSE: Spectral imaging is a method in medical x-ray imaging to extract information about the object constituents by the material-specific energy dependence of x-ray attenuation. The authors have investigated a photon-counting spectral imaging system with two energy bins for contrast-enhanced mammography. System optimization and the potential benefit compared to conventional non-energy-resolved absorption imaging was studied.
METHODS: A framework for system characterization was set up that included quantum and anatomical noise and a theoretical model of the system was benchmarked to phantom measurements.
RESULTS: Optimal combination of the energy-resolved images corresponded approximately to minimization of the anatomical noise, which is commonly referred to as energy subtraction. In that case, an ideal-observer detectability index could be improved close to 50% compared to absorption imaging in the phantom study. Optimization with respect to the signal-to-quantum-noise ratio, commonly referred to as energy weighting, yielded only a minute improvement. In a simulation of a clinically more realistic case, spectral imaging was predicted to perform approximately 30% better than absorption imaging for an average glandularity breast with an average level of anatomical noise. For dense breast tissue and a high level of anatomical noise, however, a rise in detectability by a factor of 6 was predicted. Another approximately 70%-90% improvement was found to be within reach for an optimized system.
CONCLUSIONS: Contrast-enhanced spectral mammography is feasible and beneficial with the current system, and there is room for additional improvements. Inclusion of anatomical noise is essential for optimizing spectral imaging systems.

Mesh:

Substances:

Year:  2010        PMID: 20527535     DOI: 10.1118/1.3371689

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

Review 1.  [Digital breast tomosynthesis : technical principles, current clinical relevance and future perspectives].

Authors:  K Hellerhoff
Journal:  Radiologe       Date:  2010-11       Impact factor: 0.635

2.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

3.  Detective quantum efficiency of photon-counting CdTe and Si detectors for computed tomography: a simulation study.

Authors:  Mats Persson; Adam Wang; Norbert J Pelc
Journal:  J Med Imaging (Bellingham)       Date:  2020-07-17

Review 4.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

5.  The effect of amorphous selenium detector thickness on dual-energy digital breast imaging.

Authors:  Yue-Houng Hu; Wei Zhao
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

6.  Energy weighting improves dose efficiency in clinical practice: implementation on a spectral photon-counting mammography system.

Authors:  Johan Berglund; Henrik Johansson; Mats Lundqvist; Björn Cederström; Erik Fredenberg
Journal:  J Med Imaging (Bellingham)       Date:  2014-08-28

7.  Physics considerations in MV-CBCT multi-layer imager design.

Authors:  Yue-Houng Hu; Rony Fueglistaller; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Shedlock; Daniel Morf; Paul Baturin; Pascal Huber; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-06-20       Impact factor: 3.609

8.  Breast density evaluation using spectral mammography, radiologist reader assessment, and segmentation techniques: a retrospective study based on left and right breast comparison.

Authors:  Sabee Molloi; Huanjun Ding; Stephen Feig
Journal:  Acad Radiol       Date:  2015-05-29       Impact factor: 3.173

9.  Quantitative contrast-enhanced spectral mammography based on photon-counting detectors: A feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

10.  Volumetric CT with sparse detector arrays (and application to Si-strip photon counters).

Authors:  A Sisniega; W Zbijewski; J W Stayman; J Xu; K Taguchi; E Fredenberg; Mats Lundqvist; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-11-27       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.