Detective quantum efficiency of photon-counting CdTe and Si detectors for computed tomography: a simulation study.

Mats Persson1,2, Adam Wang2, Norbert J Pelc1,2,3.   

Abstract

Purpose: Developing photon-counting CT detectors requires understanding the impact of parameters, such as converter material, thickness, and pixel size. We apply a linear-systems framework, incorporating spatial and energy resolution, to study realistic silicon (Si) and cadmium telluride (CdTe) detectors at a low count rate. Approach: We compared CdTe detector designs with 0.5 × 0.5    mm 2 and 0.225 × 0.225    mm 2 pixels and Si detector designs with 0.5 × 0.5    mm 2 pixels of 30 and 60 mm active thickness, with and without tungsten scatter blockers. Monte-Carlo simulations of photon transport were used together with Gaussian charge sharing models fitted to published data.
Results: For detection in a 300-mm-thick object at 120 kVp, the 0.5- and 0.225-mm pixel CdTe systems have 28% to 41% and 5% to 29% higher detective quantum efficiency (DQE), respectively, than the 60-mm Si system with tungsten, whereas the corresponding numbers for two-material decomposition are 2% lower to 11% higher DQE and 31% to 54% lower DQE compared to Si. We also show that combining these detectors with dual-spectrum acquisition is beneficial. Conclusions: In the low-count-rate regime, CdTe detector systems outperform the Si systems for detection tasks, whereas silicon outperforms one or both of the CdTe systems for material decomposition.
© 2020 Society of Photo-Optical Instrumentation Engineers (SPIE).

Entities:  

Keywords:  cadmium telluride detector; detective quantum efficiency; performance comparison; photon-counting; silicon detector; x-ray computed tomography

Year:  2020        PMID: 32715022      PMCID: PMC7365957          DOI: 10.1117/1.JMI.7.4.043501

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  42 in total

Review 1.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

2.  Dose-efficient ultrahigh-resolution scan mode using a photon counting detector computed tomography system.

Authors:  Shuai Leng; Zhicong Yu; Ahmed Halaweish; Steffen Kappler; Katharina Hahn; Andre Henning; Zhoubo Li; John Lane; David L Levin; Steven Jorgensen; Erik Ritman; Cynthia McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2016-12-22

3.  Spatio-energetic cross-talk in photon counting detectors: N × N binning and sub-pixel masking.

Authors:  Katsuyuki Taguchi; Karl Stierstorfer; Christoph Polster; Okkyun Lee; Steffen Kappler
Journal:  Med Phys       Date:  2018-09-27       Impact factor: 4.071

4.  A framework for performance characterization of energy-resolving photon-counting detectors.

Authors:  Mats Persson; Paurakh L Rajbhandary; Norbert J Pelc
Journal:  Med Phys       Date:  2018-10-12       Impact factor: 4.071

5.  Tungsten anode spectral model using interpolating cubic splines: unfiltered x-ray spectra from 20 kV to 640 kV.

Authors:  Andrew M Hernandez; John M Boone
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

6.  Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-Filter: Assessment of Image Quality, Radiation Dose, and Accuracy of Dual-Energy Applications in an In Vitro and In Vivo Study.

Authors:  André Euler; Anushri Parakh; Anna L Falkowski; Sebastian Manneck; David Dashti; Bernhard Krauss; Zsolt Szucs-Farkas; Sebastian T Schindera
Journal:  Invest Radiol       Date:  2016-08       Impact factor: 6.016

7.  Dual-source dual-energy CT with additional tin filtration: Dose and image quality evaluation in phantoms and in vivo.

Authors:  Andrew N Primak; Juan Carlos Ramirez Giraldo; Christian D Eusemann; Bernhard Schmidt; Birgit Kantor; Joel G Fletcher; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2010-11       Impact factor: 3.959

8.  Dual-source photon counting detector CT with a tin filter: a phantom study on iodine quantification performance.

Authors:  Ashley Tao; Richard Huang; Shengzhen Tao; Gregory J Michalak; Cynthia H McCollough; Shuai Leng
Journal:  Phys Med Biol       Date:  2019-05-31       Impact factor: 3.609

9.  Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array.

Authors:  Zhicong Yu; Shuai Leng; Steven M Jorgensen; Zhoubo Li; Ralf Gutjahr; Baiyu Chen; Ahmed F Halaweish; Steffen Kappler; Lifeng Yu; Erik L Ritman; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-02-02       Impact factor: 3.609

10.  Technical Note: spektr 3.0-A computational tool for x-ray spectrum modeling and analysis.

Authors:  J Punnoose; J Xu; A Sisniega; W Zbijewski; J H Siewerdsen
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

View more
  1 in total

1.  Spectral Photon Counting CT: Imaging Algorithms and Performance Assessment.

Authors:  Adam S Wang; Norbert J Pelc
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-07-07
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.