Marcony R Santhiago1, Marcelo V Netto, Jackson Barreto, Beatriz A F Gomes, Arthur Schaefer, Newton Kara-Junior.
Abstract
PURPOSE: To evaluate wavefront performance and modulation transfer function (MTF) in the human eye after the implantation of diffractive or refractive multifocal intraocular lenses (IOLs).Entities:
Mesh:
Year: 2010 PMID: 20195032 PMCID: PMC2854440 DOI: 10.4103/0301-4738.60075
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Wavefront analysis for a 5 mm pupil
| Mean (μm) ± SD | ||||
|---|---|---|---|---|
| Total RMS | HOA RMS | Coma RMS | Spherical aberration RMS | |
| Tecnis | 0.89 ± 0.74 | 0.58 ± 0.72 | 0.21 ± 0.22 | 0.06 ± 0.05 |
| ReZoom | 2.13 ± 2.16 | 1.27 ± 2.06 | 0.50 ± 0.72 | 0.27 ± 0.47 |
| ReSTOR | 1.30 ± 1.30 | 0.92 ± 1.14 | 0.35 ± 0.52 | 0.15 ± 0.18 |
| 0.001 | 0.094 | 0.060 | 0.007 | |
HOA - total higher order aberration; RMS - root mean square; SD - standard deviation.
Statistically significant difference between groups (ANOVA)
Figure 1Modulation transfer function curves according to IOL groups for the 5 mm pupil diameter at different spatial frequencies. The modulation transfer function curve shows the contrast transferred at different spatial frequency for each IOL. The aspheric multifocal IOL Tecnis performed similar to spheric multifocal IOL ReSTOR. And both performed better than multifocal IOL ReZoom in a 5 mm pupil (P < 0.05). The MTF does refer to distance vision
Figure 2Box plot showing the average height from the modulation transfer function curve according to IOL groups. The aberrometer device used in this study gives an average height for each MTF curve. This box plot shows the average height and the difference among three multifocal IOLs, considering the mean modulation transfer ratio at the spatial frequencies evaluated in the study. The contrast transferred was statistically similar in the Tecnis and ReSTOR group. For a 5 mm pupil diameter, the ReZoom group transferred less contrast (P < 0.05)