Literature DB >> 15130633

Improved functional vision with a modified prolate intraocular lens.

Mark Packer1, I Howard Fine, Richard S Hoffman, Patricia A Piers.   

Abstract

PURPOSE: To evaluate whether the Tecnis Z9000 intraocular lens (IOL) (Pfizer) with a modified prolate anterior surface provides better quality of vision than a conventional spherical IOL.
SETTING: Oregon Eye Institute, Eugene, Oregon, USA.
METHODS: Patients presenting for cataract surgery who were randomly assigned to receive a Tecnis Z9000 IOL (Pfizer) or a Sensar OptiEdge AR40e IOL (AMO) in 1 eye were followed for 3 months postoperatively. The patient could elect to have the same type of IOL implanted in the fellow eye. The results of sine-wave grating contrast sensitivity testing under mesopic and photopic conditions were compared interindividually.
RESULTS: Monocular comparison was made between the 2 IOL groups, which comprised 15 patients each. The Tecnis IOL provided significantly better contrast sensitivity at 6 cycles per degree (cpd) under photopic conditions and at 1.5 and 3 cpd under mesopic conditions. Seven patients with a Tecnis IOL and 9 patients with an AR40e IOL had subsequent implantation in the fellow eye. In all eyes, including fellow eyes, having IOL implantation, the Tecnis provided significantly better contrast sensitivity at 3 and 6 cpd under photopic conditions and at 1.5, 3, and 6 cpd under mesopic conditions. The mean contrast sensitivity in fellow eyes showed that the Tecnis IOL produced significantly better results at some spatial frequencies.
CONCLUSIONS: Results show the Tecnis IOL with a modified prolate anterior surface may produce better contrast sensitivity than a standard spherical IOL under mesopic and photopic conditions. Because contrast sensitivity testing correlates well with functional vision, a goal of future research should be to evaluate patient performance using functional tests such as driving simulation.

Entities:  

Mesh:

Year:  2004        PMID: 15130633     DOI: 10.1016/j.jcrs.2003.10.022

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  22 in total

1.  [Investigation of the theoretical image quality of aspheric intraocular lenses by decentration. Hoya AF-1 iMics1 und Zeiss ASPHINA(TM) (Invent ZO)].

Authors:  M Gillner; A Langenbucher; T Eppig
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

2.  Investigation and management of an epidemic of Hydroview intraocular lens opacification.

Authors:  R W Altaie; T Costigan; S Donegan; P O'brien; U Mahmood; A J Bogdan; S Beatty
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-06-11       Impact factor: 3.117

3.  Star testing: a novel evaluation of intraocular lens optical quality.

Authors:  L Mitchell; A C B Molteno; T H Bevin; G Sanderson
Journal:  Br J Ophthalmol       Date:  2006-05       Impact factor: 4.638

4.  Comparison of contrast sensitivity, depth of field and ocular wavefront aberrations in eyes with an IOL with zero versus positive spherical aberration.

Authors:  Jay S Pepose; Mujtaba A Qazi; Keith H Edwards; Jeff P Sanderson; Edwin J Sarver
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-03-11       Impact factor: 3.117

5.  [The aspheric blue light filter IOL AcrySof IQ compared to the AcrySof SA60AT : influence of IOL power, pupil diameter, and corneal asphericity on postoperative spherical aberration].

Authors:  U Mester; H Kaymak
Journal:  Ophthalmologe       Date:  2008-11       Impact factor: 1.059

6.  Comparative study of visual function and ocular aberrations of two different one-piece designed hydrophilic acrylic intraocular lens.

Authors:  Yuxin Zhao; Zhaoxia Wang; Xia Tian; Xuehong Wang; Xining Gao
Journal:  Int Ophthalmol       Date:  2017-05-30       Impact factor: 2.031

7.  Corneal spherical aberration in Saudi population.

Authors:  Tarfah M Al-Sayyari; Samah M Fawzy; Ahmed A Al-Saleh
Journal:  Saudi J Ophthalmol       Date:  2014-03-18

8.  Wavefront analysis and modulation transfer function of three multifocal intraocular lenses.

Authors:  Marcony R Santhiago; Marcelo V Netto; Jackson Barreto; Beatriz A F Gomes; Arthur Schaefer; Newton Kara-Junior
Journal:  Indian J Ophthalmol       Date:  2010 Mar-Apr       Impact factor: 1.848

9.  Visual Performance of Tecnis ZM900 Diffractive Multifocal IOL after 2500 Implants: A 3-Year Followup.

Authors:  Leonardo Akaishi; Rodrigo Vaz; Graziela Vilella; Rodrigo C Garcez; Patrick F Tzelikis
Journal:  J Ophthalmol       Date:  2010-10-05       Impact factor: 1.909

10.  A contralateral eye study comparing apodized diffractive and full diffractive lenses: wavefront analysis and distance and near uncorrected visual acuity.

Authors:  Marcony Rodrigues de Santhiago; Marcelo Vieira Netto; Jackson Barreto; Beatriz de Abreu Fiuza Gomes; Arthur Schaefer; Newton Kara-Junior
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365