Imaging quality of intraocular lenses.

Rainer Rawer1, Wilhelm Stork, Christoph W Spraul, Christian Lingenfelder.   

Abstract

PURPOSE: To compare the imaging quality of several commercially available intraocular lenses (IOLs) using the modulation transfer function (MTF) method according to the requirements of the International Standard European Industrial Norm/International Organization for Standardization 11979.
SETTING: Universitat für Technik der Informationsverarbeitung, Universität Karlsruhe, Karlsruhe and Universitäts Augenklinik, Ulm, Ulm, Germany.
METHODS: Two refractive powers (+13 diopter [D] and +26 D) of each IOL type were tested. The imaging quality of all tested IOLs was measured before the injection test and in 5-minute intervals after the injection for 40 minutes. The measurement data were analyzed according to 2 quality criteria, Strehl ratio and the ISO condition. For analysis of the type of optical design, SEM images of the central cross section were taken of each IOL. Basic terms in optics such as MTF, resolution, and optical aberration are introduced, and their relevance for ophthalmology is discussed in detail. This paper represents an extensive study comparing the imaging quality of various IOLs including the comparison of measurements before and after injection through a cartridge as used for state-of-the-art implantation techniques. All measurements were performed using an MTF measurement system using a 546 nm optimized laser source and an effective aperture of 3.0 mm on the IOL according to ISO standard. An ISO standard eye model using an artificial cornea (spherical design) and a watery solution to simulate in vivo conditions was used.
RESULTS: All IOLs apart from 1 +26 D lens complied with ISO standard requirements. However, the imaging quality varied up to 49% depending on the optical design and the manufacturing quality.
CONCLUSION: When the methods and materials recommended by the manufacturer for IOL injection through a cartridge were used, no long-term effect on the imaging quality of the IOLs was observed. However, the basic imaging quality of different designs and different manufacturers varied significantly (up to 50% at +26 D) IOLs and should be considered when choosing an IOL.

Mesh:

Year:  2005        PMID: 16129302     DOI: 10.1016/j.jcrs.2005.01.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  15 in total

1.  Star testing: a novel evaluation of intraocular lens optical quality.

Authors:  L Mitchell; A C B Molteno; T H Bevin; G Sanderson
Journal:  Br J Ophthalmol       Date:  2006-05       Impact factor: 4.638

2.  [Microincision intraocular lens with plate haptic design. Evaluation of rotational stability and centering of a microincision intraocular lens with plate haptic design in 12-19 months of follow-up].

Authors:  W Wehner
Journal:  Ophthalmologe       Date:  2007-05       Impact factor: 1.059

3.  Modulation transfer function of intraocular collamer lens with a central artificial hole.

Authors:  Hiroshi Uozato; Kimiya Shimizu; Takushi Kawamorita; Fumiko Ohmoto
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-01-13       Impact factor: 3.117

4.  Corneal spherical aberration and its impact on choosing an intraocular lens for cataract surgery.

Authors:  Tarfah M Al-Sayyari; Samah M Fawzy; Ahmed A Al-Saleh
Journal:  Saudi J Ophthalmol       Date:  2014-06-25

5.  Intra-ocular lens optical changes resulting from the loading of dexamethasone.

Authors:  José M Artigas; M Carmen García-Domene; Amparo Navea; Pablo Botella; Eduardo Fernández
Journal:  Biomed Opt Express       Date:  2017-09-21       Impact factor: 3.732

6.  Method for the generation and visualization of cross-sectional images of three-dimensional point spread functions for rotationally symmetric intraocular lenses.

Authors:  Jan Sievers; Ricardo Elsner; Sebastian Bohn; Melanie Schünemann; Heinrich Stolz; Rudolf F Guthoff; Oliver Stachs; Karsten Sperlich
Journal:  Biomed Opt Express       Date:  2022-02-01       Impact factor: 3.732

7.  Wavefront analysis and modulation transfer function of three multifocal intraocular lenses.

Authors:  Marcony R Santhiago; Marcelo V Netto; Jackson Barreto; Beatriz A F Gomes; Arthur Schaefer; Newton Kara-Junior
Journal:  Indian J Ophthalmol       Date:  2010 Mar-Apr       Impact factor: 1.848

8.  Visual function and higher order aberration after implantation of aspheric and spherical multifocal intraocular lenses: a meta-analysis.

Authors:  Jian-Ping Liu; Fan Zhang; Jiang-Yue Zhao; Li-Wei Ma; Jin-Song Zhang
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

9.  Deformable Surface Accommodating Intraocular Lens: Second Generation Prototype Design Methodology and Testing.

Authors:  Sean J McCafferty; Jim T Schwiegerling
Journal:  Transl Vis Sci Technol       Date:  2015-04-28       Impact factor: 3.283

10.  A contralateral eye study comparing apodized diffractive and full diffractive lenses: wavefront analysis and distance and near uncorrected visual acuity.

Authors:  Marcony Rodrigues de Santhiago; Marcelo Vieira Netto; Jackson Barreto; Beatriz de Abreu Fiuza Gomes; Arthur Schaefer; Newton Kara-Junior
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.