Literature DB >> 26949654

Comparison of two types of visual quality analyzer for the measurement of high order aberrations.

Jing Hao1, Lin Li2, Fang Tian3, Hong Zhang3.   

Abstract

AIM: To compare the difference and agreement of KR-1W and iTrace for measurement of high order aberrations.
METHODS: KR-1W and iTrace were respectively used in a group of healthy people (40 volunteers, 32 eyes) to measure the high order aberration (HOA) of corneal, internal and total ocular. The clinical difference and agreement of two instruments were respectively evaluated by paired-samples t-test and Bland-Altman analysis.
RESULTS: The paired-samples t-test showed that the corneal HOA measured by the two instruments had no statistical differences (P>0.05); but the internal and total ocular HOA had significant statistical differences (P<0.05), and the mean results measured by iTrace were higher than that of KR-1W. However, Bland-Altman analysis revealed that the HOA of corneal and internal were all in 95% limits of agreement; and just one point of total ocular HOA was beyond the 95% limits of agreement.
CONCLUSION: KR-1W and iTrace were consistent well in the measurement of corneal, internal and total ocular HOA, especially for the cornea.

Entities:  

Keywords:  KR-1W; agreement; difference; high order aberration; iTrace

Year:  2016        PMID: 26949654      PMCID: PMC4761746          DOI: 10.18240/ijo.2016.02.22

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  21 in total

1.  Evaluation of refractive error measurements of the Wavescan Wavefront system and the Tracey Wavefront aberrometer.

Authors:  Li Wang; Nan Wang; Douglas D Koch
Journal:  J Cataract Refract Surg       Date:  2003-05       Impact factor: 3.351

Review 2.  [Contemporary applications of wavefront aberrometry in ophthalmology practice].

Authors:  Lucyna Moćko; Marek Zajac; Anna M Roszkowska; Beata Weglarz; Michał Milka; Dariusz Dobrowolski; Edward Wylegała
Journal:  Klin Oczna       Date:  2012

3.  Intrasubject repeatability of internal aberrometry obtained with a new integrated aberrometer.

Authors:  David P Piñero; Javier Tomás Juan; Jorge L Alió
Journal:  J Refract Surg       Date:  2010-12-15       Impact factor: 3.573

4.  Corneal reshaping and wavefront aberrations during overnight orthokeratology.

Authors:  Yan Lian; Meixiao Shen; Shenghai Huang; Yimin Yuan; Yaozeng Wang; Dexi Zhu; Jun Jiang; Xinjie Mao; Jianhua Wang; Fan Lu
Journal:  Eye Contact Lens       Date:  2014-05       Impact factor: 2.018

5.  Comparison of wavefront aberrations in rabbit and human eyes.

Authors:  Li Chen; Ling C Huang; Brad Gray; Dimitri A Chernyak
Journal:  Clin Exp Optom       Date:  2014-07-29       Impact factor: 2.742

6.  Changes in functional visual acuity and ocular wavefront aberration after administration of eye ointment.

Authors:  Takahiro Hiraoka; Toshiya Yamamoto; Fumiki Okamoto; Tetsuro Oshika
Journal:  J Ocul Pharmacol Ther       Date:  2013-06-28       Impact factor: 2.671

7.  Comparison of monochromatic aberrations in young adults with different visual acuity and refractive errors.

Authors:  Seyhan Yazar; Alex W Hewitt; Hannah Forward; Charlotte M McKnight; Alex Tan; Jenny A Mountain; David A Mackey
Journal:  J Cataract Refract Surg       Date:  2014-01-11       Impact factor: 3.351

8.  Wavefront analysis and modulation transfer function of three multifocal intraocular lenses.

Authors:  Marcony R Santhiago; Marcelo V Netto; Jackson Barreto; Beatriz A F Gomes; Arthur Schaefer; Newton Kara-Junior
Journal:  Indian J Ophthalmol       Date:  2010 Mar-Apr       Impact factor: 1.848

9.  Comparison of internal and total optical aberrations for 2 aberrometers: iTrace and OPD scan.

Authors:  Jung Bin Won; Sun Woong Kim; Eung Kweon Kim; Byung Jin Ha; Tae-im Kim
Journal:  Korean J Ophthalmol       Date:  2008-12

10.  The relationship between higher-order aberrations and amblyopia treatment in hyperopic anisometropic amblyopia.

Authors:  Soo Han Lee; Ji Woong Chang
Journal:  Korean J Ophthalmol       Date:  2014-01-21
View more
  6 in total

1.  Bias of near-infrared light in evaluation of patients implanted with multifocal intraocular lenses.

Authors:  Francisco de Asís Bartol-Puyal; Galadriel Giménez; Silvia Méndez-Martínez; Irene Altemir; José Manuel Larrosa; Vicente Polo; Luis Pablo
Journal:  Int Ophthalmol       Date:  2021-05-25       Impact factor: 2.031

2.  Comparison of wavefront aberrations in the object and image spaces using wide-field individual eye models.

Authors:  Yongji Liu; Xiaolan Li; Lin Zhang; Xianglong Yi; Yuwei Xing; Kunqi Li; Yan Wang
Journal:  Biomed Opt Express       Date:  2022-08-25       Impact factor: 3.562

Review 3.  From Presbyopia to Cataracts: A Critical Review on Dysfunctional Lens Syndrome.

Authors:  Joaquín Fernández; Manuel Rodríguez-Vallejo; Javier Martínez; Ana Tauste; David P Piñero
Journal:  J Ophthalmol       Date:  2018-06-27       Impact factor: 1.909

4.  Distribution of corneal spherical aberration in a Tanzanian population.

Authors:  Hiroki Asano; Takahiro Hiraoka; Yusuke Seki; Teppei Shibata; Hiromi Osada; Takanori Saruta; Natsuko Hatsusaka; Fukumi Fujikake; Yoshiaki Tabata; Cellina Mhina; Anna Sanyiwa; Tetsuro Oshika; Hiroshi Sasaki
Journal:  PLoS One       Date:  2019-09-12       Impact factor: 3.240

5.  Optical aberrations in three different intraocular lens designs of a same platform.

Authors:  Francisco D A Bartol-Puyal; Galadriel Giménez; Irene Altemir; José M Larrosa; Vicente Polo; Luis Pablo
Journal:  Saudi J Ophthalmol       Date:  2022-02-18

6.  Comparison of the optical quality after SMILE and FS-LASIK for high myopia by OQAS and iTrace analyzer: a one-year retrospective study.

Authors:  Yewei Yin; Ying Lu; Aiqun Xiang; Yanyan Fu; Yang Zhao; Yuanjun Li; Tu Hu; Kaixuan Du; Shengfa Hu; Qiuman Fu; Xiaoying Wu; Dan Wen
Journal:  BMC Ophthalmol       Date:  2021-08-02       Impact factor: 2.209

  6 in total