Scan equalization digital radiography (SEDR) implemented with an amorphous selenium flat-panel detector: initial experience.

Xinming Liu1, Chao-Jen Lai, Lingyun Chen, Tao Han, Yuncheng Zhong, Youtao Shen, Tianpeng Wang, Chris C Shaw.   

Abstract

It is well recognized in projection radiography that low-contrast detectability suffers in heavily attenuating regions due to excessively low x-ray fluence to the image receptor and higher noise levels. Exposure equalization can improve image quality by increasing the x-ray exposure to heavily attenuating regions, resulting in a more uniform distribution of exposure to the detector. Image quality is also expected to be improved by using the slot-scan geometry to reject scattered radiation effectively without degrading primary x-rays. This paper describes the design of a prototype scan equalization digital radiography (SEDR) system implemented with an amorphous silicon (a-Si) thin-film transistor (TFT) array-based flat-panel detector. With this system, slot-scan geometry with alternate line erasure and readout (ALER) technique was used to achieve scatter rejection. A seven-segment beam height modulator assembly was mounted onto the fore collimator to regulate exposure regionally for chest radiography. The beam modulator assembly, consisting of micro linear motors, lead screw cartridge with lead (Pb) beam blockers attached, position feedback sensors and motor driver circuitry, has been tested and found to have an acceptable response for exposure equalization in chest radiography. An anthropomorphic chest phantom was imaged in the posterior-anterior (PA) view under clinical conditions. Scatter component, primary x-rays, scatter-to-primary ratios (SPRs) and primary signal-to-noise ratios (PSNRs) were measured in the SEDR images to evaluate the rejection and redistribution of scattered radiation, and compared with those for conventional full-field imaging with and without anti-scatter grid methods. SPR reduction ratios (SPRRRs, defined as the differences between the non-grid full-field SPRs and the reduced SPRs divided by the former) yielded approximately 59% for the full-field imaging with grid and 82% for the SEDR technique in the lungs, and 77% for the full-field imaging with grid and 95% for the SEDR technique in the subdiaphragm. The SEDR technique demonstrated a substantial improvement in PSNRs over the anti-scatter grid technique. The improvements of PSNRs varied with the regions and are more pronounced in heavily attenuating regions.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19887717      PMCID: PMC2855213          DOI: 10.1088/0031-9155/54/22/014

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  35 in total

1.  Area x-ray beam equalization for digital angiography.

Authors:  S Molloi; J Tang; T Mather; Y Zhou
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  X-ray beam equalization: feasibility and performance of an automated prototype system in a phantom and swine.

Authors:  S Molloi; A Van Drie; F Wang
Journal:  Radiology       Date:  2001-12       Impact factor: 11.105

3.  A slot-scanned photodiode-array/CCD hybrid detector for digital mammography.

Authors:  James G Mainprize; Nancy L Ford; Shi Yin; Türmay Tümer; Martin J Yaffe
Journal:  Med Phys       Date:  2002-02       Impact factor: 4.071

4.  Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: a chest phantom study.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Mustafa C Altunbas; Lingyun Chen; Tao Han; Tianpeng Wang
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

5.  A scanning system for chest radiography with regional exposure control: theoretical considerations.

Authors:  D B Plewes
Journal:  Med Phys       Date:  1983 Sep-Oct       Impact factor: 4.071

6.  A scanning system for chest radiography with regional exposure control: practical implementation.

Authors:  D B Plewes; E Vogelstein
Journal:  Med Phys       Date:  1983 Sep-Oct       Impact factor: 4.071

7.  A scanning equalization system for improved chest radiography.

Authors:  D B Plewes; J C Wandtke
Journal:  Radiology       Date:  1982-03       Impact factor: 11.105

8.  Diode array digital radiography: initial clinical experience.

Authors:  D Sashin; E J Sternglass; B S Slasky; K M Bron; J M Herron; W H Kennedy; L Shabason; J Boyer; A E Pollitt; R E Latchaw; B R Girdany; R W Simpson
Journal:  AJR Am J Roentgenol       Date:  1982-12       Impact factor: 3.959

9.  Improved contrast in special procedures using a rotating aperture wheel (RAW) device.

Authors:  S Rudin; D R Bednarek
Journal:  Radiology       Date:  1980-11       Impact factor: 11.105

10.  Digital selenium radiography: anti-scatter grid for chest radiography in a clinical study.

Authors:  T M Bernhardt; U Rapp-Bernhardt; T Hausmann; G Reichel; U W Krause; W Doehring
Journal:  Br J Radiol       Date:  2000-09       Impact factor: 3.039

View more
  2 in total

1.  Effects of exposure equalization on image signal-to-noise ratios in digital mammography: a simulation study with an anthropomorphic breast phantom.

Authors:  Xinming Liu; Chao-Jen Lai; Gary J Whitman; William R Geiser; Youtao Shen; Ying Yi; Chris C Shaw
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Tianpeng Wang
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.