Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: a chest phantom study.

Xinming Liu1, Chris C Shaw, Chao-Jen Lai, Mustafa C Altunbas, Lingyun Chen, Tao Han, Tianpeng Wang.   

Abstract

Anti-scatter grids have been widely used to reject scatter and increase the perceptibility of low-contrast object in chest radiography; however they also attenuate the primary x-rays, resulting in a substantial degradation of primary information. Compensation for this degradation requires the use of higher exposure technique hence higher dose to the patient. A more efficient approach to reject scatter is the slot-scan imaging technique which employs a narrow scanning x-ray fan beam in conjunction with a slit or slot shaped solid state detector or an area detector used with an aft-collimator. With this approach, scatter can be rejected effectively without the need to attenuate primary x-rays. This paper demonstrates an electronic aft-collimation method, referred to as the alternate line erasure and readout (ALER) technique, for implementing the slot-scan digital radiography with a modern flat-panel detector. With this technique, instead of first exposing the detector and then reading the image line by line, the image line on the leading edge of the scanning fan beam is reset to erase the scatter accumulated prior to the arrival of the fan beam x-rays, while the image line on the trailing edge of the scanning fan beam is read out to acquire the image signals following the fan-beam exposure. These reset and readout processes are alternated and repeated as the x-ray fan beam scans across the detector. An anthropomorphic chest phantom was imaged to evaluate the scatter rejection ability and the low-contrast performance for the ALER technique and compare them with those for the anti-scatter grid method in full-field chest imaging. With a projected beam width of 16 mm, the slot-scan/ALER technique resulted in an average reduction of the scatter-to-primary ratios by 81%, 84%, 82%, and 86% versus 65%, 73%, 74%, and 73% with the anti-scatter grid method in the lungs, mediastinum, retrocardium, and subdiaphragm, respectively. The average CNR for the slot-scan/ALER technique was found to improve by 135%, 133%, 176%, and 87% versus 15%, 15%, 38%, and -11% with the anti-scatter grid method in the mediastinum, retrocardium, subdiaphragm, and lungs, respectively. These results demonstrated that the slot-scan/ALER technique can be used to achieve equally effective scatter rejection but substantially higher low-contrast performance than the anti-scatter grid method.

Entities:  

Mesh:

Year:  2008        PMID: 18649472      PMCID: PMC2809722          DOI: 10.1118/1.2921132

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  39 in total

1.  An edge spread technique for measurement of the scatter-to-primary ratio in mammography.

Authors:  V N Cooper; J M Boone; J A Seibert; C J Pellot-Barakat
Journal:  Med Phys       Date:  2000-05       Impact factor: 4.071

2.  Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.

Authors:  S Vedantham; A Karellas; S Suryanarayanan; I Levis; M Sayag; R Kleehammer; R Heidsieck; C J D'Orsi
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

3.  A time-delay integration charge-coupled device camera for slot-scanned digital radiography.

Authors:  D W Holdsworth; R K Gerson; A Fenster
Journal:  Med Phys       Date:  1990 Sep-Oct       Impact factor: 4.071

Review 4.  Contrast and scatter in x-ray imaging.

Authors:  G T Barnes
Journal:  Radiographics       Date:  1991-03       Impact factor: 5.333

5.  Fundamental imaging characteristics of a slot-scan digital chest radiographic system.

Authors:  Ehsan Samei; Robert S Saunders; Joseph Y Lo; James T Dobbins; Jonathan L Jesneck; Carey E Floyd; Carl E Ravin
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

6.  Grids or air gaps for scatter reduction in digital radiography: a model calculation.

Authors:  U Neitzel
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

7.  Examination of aperture signals in digital radiography.

Authors:  K P Maher; J F Malone
Journal:  Phys Med Biol       Date:  1998-03       Impact factor: 3.609

8.  Digital radiographic imaging system with multiple-slit scanning x-ray beam: preliminary report.

Authors:  K Doi; H Fujita; K Ohara; K Ono; H Matsui; M L Giger; H P Chan
Journal:  Radiology       Date:  1986-11       Impact factor: 11.105

9.  Scatter and veiling glare estimation based on sampled primary intensity.

Authors:  Y Zhou; T Mathur; S Molloi
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

10.  Photostimulable phosphor digital radiography of the extremities: diagnostic accuracy compared with conventional radiography.

Authors:  A J Wilson; F A Mann; W A Murphy; B S Monsees; M R Linn
Journal:  AJR Am J Roentgenol       Date:  1991-09       Impact factor: 3.959

View more
  4 in total

1.  Effects of exposure equalization on image signal-to-noise ratios in digital mammography: a simulation study with an anthropomorphic breast phantom.

Authors:  Xinming Liu; Chao-Jen Lai; Gary J Whitman; William R Geiser; Youtao Shen; Ying Yi; Chris C Shaw
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Tianpeng Wang
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

3.  Scan equalization digital radiography (SEDR) implemented with an amorphous selenium flat-panel detector: initial experience.

Authors:  Xinming Liu; Chao-Jen Lai; Lingyun Chen; Tao Han; Yuncheng Zhong; Youtao Shen; Tianpeng Wang; Chris C Shaw
Journal:  Phys Med Biol       Date:  2009-11-04       Impact factor: 3.609

4.  A static multi-slit collimator system for scatter reduction in cone-beam CT.

Authors:  Jina Chang; Siyong Kim; Doh-Yun Jang; Tae-Suk Suh
Journal:  J Appl Clin Med Phys       Date:  2010-10-07       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.