T M Bernhardt1, U Rapp-Bernhardt, T Hausmann, G Reichel, U W Krause, W Doehring.
Abstract
The purpose of this study was to evaluate the diagnostic performance of an additional stationary anti-scatter grid in digital selenium radiography (DSR) compared with images acquired with only an air gap. Chest radiographs were obtained with DSR in 100 patients with and without an anti-scatter grid. Four observers scored 12 anatomical landmarks, catheters and wire cerclages for their visualization in both subsets of images. Statistical analysis was performed using a paired t-test. Anatomical landmarks, catheters and wire cerclages were statistically better visualized in regions of high attenuation when the images were performed with an anti-scatter grid. No statistically significant difference was noted for peripheral regions, nor for sex and weight of the patient between the two modalities. Therefore, an anti-scatter grid is not recommended for chest radiography as it increases the radiation exposure of patients without having a significant impact on visualization for all regions of the chest.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11064649 DOI: 10.1259/bjr.73.873.11064649
Source DB: PubMed Journal: Br J Radiol ISSN: 0007-1285 Impact factor: 3.039