Literature DB >> 16637838

Clinical evaluation of rebound tonometer.

Leon N Davies1, Hannah Bartlett, Edward A H Mallen, James S Wolffsohn.   

Abstract

PURPOSE: To evaluate the reliability and repeatability of intraocular pressure (IOP) measurements using a new rebound tonometer.
METHODS: Intraocular pressure was measured in 42 healthy human eyes of subjects aged 18-30 years (mean +/- standard deviation [SD] 21.5 +/- 3.2 years) using the ICare Rebound and Goldmann tonometers in two separate sessions.
RESULTS: Intraocular pressure measurements were found to read slightly, but not significantly, higher with the ICare tonometer compared with the Goldmann instrument in both sessions (first session: mean bias +/- SD + 0.50 +/- 2.33 mmHg; second session: mean bias +/- SD + 0.52 +/- 1.92 mmHg). Limits of agreement between repeated readings were +/- 5.11 mmHg for measurements taken with the ICare tonometer, compared with +/- 3.15 mmHg for measurements taken with the Goldmann method.
CONCLUSION: Measurement of IOP in normal, healthy subjects using the ICare rebound tonometer produced a small, statistically insignificant, positive bias when compared with the Goldmann tonometer. Intersessional repeatability of IOP taken with the ICare is poorer than that of IOP taken with the Goldmann tonometer, but is comparable with that of other non-Goldman-type tonometers currently available.

Entities:  

Mesh:

Year:  2006        PMID: 16637838     DOI: 10.1111/j.1600-0420.2005.00610.x

Source DB:  PubMed          Journal:  Acta Ophthalmol Scand        ISSN: 1395-3907


  28 in total

1.  Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects.

Authors:  Ioannis Halkiadakis; Aimilianos Stratos; George Stergiopoulos; Eleni Patsea; Sotiris Skouriotis; Panagiotis Mitropoulos; Dimitrios Papaconstantinou; Gerasimos Georgopoulos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08       Impact factor: 3.117

2.  Effects of trabeculectomy on posture-induced intraocular pressure changes over time.

Authors:  Akira Sawada; Tetsuya Yamamoto
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-02-10       Impact factor: 3.117

3.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

Review 4.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

5.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

6.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

7.  Intraocular pressure increases after complex simulated surgical procedures in residents: an experimental study.

Authors:  Jesús Vera; Carolina Diaz-Piedra; Raimundo Jiménez; Jose M Sanchez-Carrion; Leandro L Di Stasi
Journal:  Surg Endosc       Date:  2018-07-02       Impact factor: 4.584

8.  [Comparison of the iCare tonometer with the Goldmann tonometer in Malawi].

Authors:  J Hohmann; M Schulze-Schwering; T Chirambo Nyaka; V Moyo; P C Kayange; D Doycheva; N H Batumba; M S Spitzer
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

9.  Effectiveness of the ICare rebound tonometer in patients with overestimated intraocular pressure due to tight orbit syndrome.

Authors:  You Kyung Lee; Y K Lee; Ji Young Lee; J Y Lee; Jung Il Moon; J I Moon; Myoung Hee Park; M H Park
Journal:  Jpn J Ophthalmol       Date:  2014-08-29       Impact factor: 2.447

10.  Immediate changes in intraocular pressure after clear corneal micro-incision versus small-incision cataract surgery.

Authors:  Ken Hayashi; Motoaki Yoshida; Koichi Yoshimura
Journal:  Jpn J Ophthalmol       Date:  2014-06-28       Impact factor: 2.447