Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic.

Antti Kontiola1, Päivi Puska.   

Abstract

PURPOSE: To evaluate the utility of the new Rebound tonometer for measuring intraocular pressure (IOP) in an unanesthetized eye; to test patient tolerance, measurement time, and accuracy compared with the Pulsair 3000 tonometer.
METHODS: IOP was measured with the Rebound tonometer and the Pulsair 3000 tonometer without an anesthetic in 131 residents of two Finnish nursing homes. The measurement time and possible pain or discomfort experienced by the inhabitants was recorded.
RESULTS: The mean differences in IOP readings between the two tonometers were 0.31 mmHg, SD 2.45 mmHg for the right eyes and 0.36 mmHg, SD 2.17 mmHg for the left eyes (P=0.28, multivariate analysis). The correlation constants between the tonometers were 0.84 (right eyes) and 0.80 (left eyes). The Pulsair 3000 caused more discomfort than the Rebound tonometer (36% vs 15%, P=0.01). With the Pulsair, 85%, and with the Rebound tonometer, 95% of the patients felt no pain ( P= 0.14). Measurement of both eyes with the Rebound tonometer took less time (55+/-22 s vs 138+/-55 s, P<0.001). The mean difference was 82 s and the 95% confidence interval of the difference was 66-98 s.
CONCLUSION: Measurement of IOP with the Rebound tonometer without an anesthetic is a rapid and well-tolerated procedure. IOP readings of the two tonometers were within +/-1 mmHg in 52.5% of the measurements and within +/-2 mmHg in 71.7% of the measurements.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14634828     DOI: 10.1007/s00417-003-0671-3

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  15 in total

1.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

2.  Measurement of intraocular pressure with a flat anterior chamber.

Authors:  M M Wright; A L Grajewski
Journal:  Ophthalmology       Date:  1991-12       Impact factor: 12.079

3.  Measurement of intraocular pressure with a modern noncontact tonometer.

Authors:  I Koçak; S Orgül; A Saruhan; I Haefliger; P Hendrickson; J Flammer
Journal:  Ophthalmologica       Date:  1998       Impact factor: 3.250

4.  The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat.

Authors:  A I Kontiola; D Goldblum; T Mittag; J Danias
Journal:  Exp Eye Res       Date:  2001-12       Impact factor: 3.467

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  [A new site-independent tonometer (ProTon) in comparison with the Goldmann applanation tonometer].

Authors:  M Gräf; H Wassill; B Dick
Journal:  Klin Monbl Augenheilkd       Date:  1995-04       Impact factor: 0.700

Review 7.  Sources of error with use of Goldmann-type tonometers.

Authors:  M M Whitacre; R Stein
Journal:  Surv Ophthalmol       Date:  1993 Jul-Aug       Impact factor: 6.048

Review 8.  The epidemiology of open-angle glaucoma: a review.

Authors:  M C Leske
Journal:  Am J Epidemiol       Date:  1983-08       Impact factor: 4.897

9.  Statistical analysis of multi-eye data in ophthalmic research.

Authors:  W A Ray; D M O'Day
Journal:  Invest Ophthalmol Vis Sci       Date:  1985-08       Impact factor: 4.799

10.  Dynamic tonometry.

Authors:  H M Dekking; H D Coster
Journal:  Ophthalmologica       Date:  1967       Impact factor: 3.250

View more
  22 in total

1.  Biocompatible Multifunctional Black-Silicon for Implantable Intraocular Sensor.

Authors:  Jeong Oen Lee; Vinayak Narasimhan; Juan Du; Blaise Ndjamen; David Sretavan; Hyuck Choo
Journal:  Adv Healthc Mater       Date:  2017-01-12       Impact factor: 9.933

Review 2.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

3.  Evaluation of the impedance tonometers TGDc-01 and iCare according to the international ocular tonometer standards ISO 8612.

Authors:  Peter Christian Ruokonen; Thomas Schwenteck; Jörg Draeger
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-02-21       Impact factor: 3.117

4.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

5.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

6.  Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients.

Authors:  Milena Pahlitzsch; Jeanette Brünner; Johannes Gonnermann; Anna-Karina B Maier; Necip Torun; Eckart Bertelmann; Matthias Kj Klamann
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

7.  Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness.

Authors:  M E Iliev; D Goldblum; K Katsoulis; C Amstutz; B Frueh
Journal:  Br J Ophthalmol       Date:  2006-05-03       Impact factor: 4.638

8.  Self-tonometry in glaucoma.

Authors:  Ahti Tarkkanen; Kristian Ulfves; Tuulikki Ulfves
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-06-08       Impact factor: 3.117

9.  Reliability and sensitivity of the TonoLab rebound tonometer in awake Brown Norway rats.

Authors:  John C Morrison; Lijun Jia; William Cepurna; Ying Guo; Elaine Johnson
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-03-25       Impact factor: 4.799

10.  Twenty-four-hour efficacy of preservative-free tafluprost for open-angle glaucoma patients, assessed by home intraocular pressure (Icare-ONE) and blood-pressure monitoring.

Authors:  Soon Young Cho; Yong Yeon Kim; Chungkwon Yoo; Tae-Eun Lee
Journal:  Jpn J Ophthalmol       Date:  2015-09-28       Impact factor: 2.447

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.