Literature DB >> 24489462

The utility of rebound tonometer in late elderly subjects.

Rei Sakata1, Jiro Numaga2.   

Abstract

BACKGROUND: To compare intraocular pressure (IOP) readings using the Icare rebound tonometer (RBT) versus the Goldmann applanation tonometer (GAT) in late elderly (aged 75 years or older) subjects with or without glaucoma, and to evaluate the influence of central corneal thickness (CCT) on IOP readings.
METHODS: The IOP measurements were obtained using the RBT and GAT, and CCT was measured using a specular-type microscope. Bland-Altman analysis was used to assess the clinical agreement between the two instruments. The influence of CCT adjusted for age on IOP readings was analyzed by multiple linear regression analysis.
RESULTS: This study included 58 eyes of 29 normal subjects and 54 eyes of 28 glaucoma patients. The mean age was 80.7±4.3 years (normal subjects) and 83.1±5.1 years (glaucoma patients). The mean IOP readings were 13.6±3.5 mmHg and 13.2±2.8 mmHg (using the RBT and GAT, respectively) for normal subjects, and 13.6±3.3 mmHg and 13.5±2.9 mmHg for glaucoma patients. The 95% confidence interval of the differences between the two instruments was -3.3 to 4.0 mmHg for normal subjects and -2.9 to 1.6 mmHg for glaucoma patients. The IOP readings by two instruments were significantly correlated with CCT in eyes with glaucoma (for the RBT, β=0.036 and P=0.002, and for the GAT, β=0.021 and P=0.033) but not in normal eyes.
CONCLUSION: IOP readings measured using the RBT and GAT showed within the allowable range in the late elderly subjects with or without glaucoma. Eyes with glaucoma were correlated closely with CCT using each instrument.

Entities:  

Keywords:  Goldmann applanation tonometer; central corneal thickness; late elderly; rebound tonometer

Mesh:

Year:  2014        PMID: 24489462      PMCID: PMC3904777          DOI: 10.2147/CIA.S57644

Source DB:  PubMed          Journal:  Clin Interv Aging        ISSN: 1176-9092            Impact factor:   4.458


Introduction

Previous randomized clinical trials have established the efficacy of reducing intraocular pressure (IOP) in patients with glaucoma,1–5 confirming that IOP management plays a central role in preventing further progression of glaucomatous optic neuropathy.6–8 Although the Goldmann applanation tonometer (GAT) (Haag Streit, Koeniz, Switzerland) has been recognized as the gold standard tonometer up to the present date,9 some limitations associated with the instrument have been reported, including the influence of central corneal thickness (CCT) on IOP readings,10–12 the necessity of topical anesthesia,13 the requirement of a certain amount of experience in measurement, the nonutility for babies or bedridden patients, and the risk of secondary infection via exposure to the cornea.14 The Icare® rebound tonometer (RBT) (Icare TA01i, Tiolat, Finland) is a compact, handheld contact tonometer that does not require topical anesthesia.15 Consequently, because of its simplicity and noninvasive nature, the RBT is used widely in children.16–18 However, the agreement between RBT and GAT readings in children is debated, and more data are necessary to assess the differences, including data at various IOPs.19,20 One reason is that the RBT might be affected by IOP.21,22 Moreover, it is necessary to investigate the effect of CCT on the RBT in infants and young children. In recent years, various RBT models have been marketed. It is now possible for patients to measure their own IOP;23 alternatively, it can be measured in the clinic in positions other than horizontal (ie, spine or decubitus positions).24 Studies comparing IOP measured in normal eyes and in those with glaucoma, using the RBT (Icare) and GAT, reported that the IOP readings tended to be higher using the RBT than the GAT.21,25–41 However, one study reported higher IOP readings using the GAT.42 Like the GAT, the RBT has also been reported to be influenced by corneal properties.33,36,43–45 Both RBT and GAT overestimate IOP in eyes with a thicker cornea, and vice versa. To date, no studies have projected to compare IOP readings measured by both instruments in late elderly (aged 75 years or older) subjects. The mean age of subjects in previous studies ranged not more than 72 years. As the average life expectancy in society continues to increase, greater numbers of geriatric patients with difficult-to-measure IOP by GAT are expected because of their ocular or physical problems; therefore, occasions for using handheld tonometers such as RBT will increase much more in upcoming medical practice. The aim of this study was to evaluate the utility of the RBT by comparing it with the GAT in late elderly subjects with or without glaucoma, and to investigate a correlation between IOP readings and CCT.

Materials and methods

Late elderly Japanese normal subjects and glaucoma patients, including those undergoing medical treatment, were recruited at the Tokyo Metropolitan Geriatric Hospital, Tokyo, Japan. Glaucoma was defined according to the presence of glaucomatous optic nerve head damage with corresponding visual field damage. Eyes were excluded in patients with ocular surface disorders (eg pterygium), which prevent IOP measurements. IOP measurements were taken by one well-trained ophthalmologist at a routine outpatient clinic. To minimize the possible reduction in IOP caused by eye oppression from contact with the tonometer,46 IOP measurements were taken by the RBT (Icare) first. Two sequential IOP readings (the right eye first) were measured. For the RBT, the mean of six consecutive measurements after removing the highest and lowest measurements were displayed with a quality indicator (P). IOP readings were accepted when the horizontal line (standard error; standard deviation) showed in the lowermost part in the display. Next, after applying one drop of topical oxybuprocaine hydrochloride anesthesia (0.4% Benoxil®, Santen, Tokyo, Japan), we used the calibrated GAT connected to the slit-lamp microscopy. Also, two sequential IOP readings (the right eye first) were measured. The average IOPs were used for analysis. CCT was measured using a specular-type microscope (Noncon Robo, Konan Medical, Nishinomiya, Japan) on the same visiting day. We are planning a study of a continuous response variable from matched pairs of study subjects. Preliminary data indicate that the difference in the response of matched pairs is normally distributed, with a standard deviation of 2.5 mmHg. If the true difference in the mean response of matched pairs is 1 mmHg, we will need to study 58 pairs of subjects to be able to reject the null hypothesis that this response difference is zero with a probability (power) of 0.85. The type I error probability associated with this test of this null hypothesis is 0.05. The Bland–Altman plot47 was used to evaluate the differences in IOP readings. The difference between IOP measurements was plotted against the mean. The correlation between the IOP obtained using the two instruments and the CCT was analyzed by multiple linear regression analysis adjusted for age. In normal subjects or glaucoma patients, we used the following equation: where β0 is intercept and β1 and β2 are regression coefficients for CCT and age, respectively. Regarding IOP, we analyzed RBT and GAT separately. Two-tailed values of P<0.05 indicated statistical significance. All analyses were performed using JMP version 9.0.2 statistical software (SAS Institute Inc., Cary, NC, USA). This study was conducted in accordance with the Declaration of Helsinki, and the study protocol was approved by the Ethics Committee of the Tokyo Metropolitan Geriatric Hospital. Informed consent was obtained from participants.

Results

A total of 29 normal subjects (58 eyes) and 28 glaucoma patients (54 eyes) completed this study with no ocular/systemic side effects. The results for 54 glaucoma eyes were as follows: 43 eyes were normal tension glaucoma (34 eyes with medications, nine eyes without medications), eight eyes were primary open-angle glaucoma (eight eyes with medications), one eye was secondary open-angle glaucoma (with medications), and two eyes were without glaucomatous optic neuropathy.

IOP readings and consistency of the two instruments

The mean IOP readings used by the RBT and GAT were 13.6±3.5 mmHg and 13.2±2.8 mmHg (mean difference 0.34 mmHg, P=0.22) in normal subjects, and 13.6±3.3 mmHg and 13.5±2.9 mmHg (mean difference 0.14 mmHg, P=0.53) in glaucoma patients. The 95% limits of agreement between the RBT and the GAT ranged from −3.3 mmHg to 4.0 mmHg in normal subjects and from −2.0 mmHg to 1.6 mmHg in glaucoma patients (Figures 1 and 2). Eyes with differences within ±1 mmHg measured by both instruments were 23/58 (40%) for normal subjects and 24/54 (44%) for glaucoma patients.
Figure 1

Bland–Altmann plot of the rebound tonometer (RBT) readings minus the Goldmann applanation tonometer (GAT) readings (mmHg) versus the mean of both in normal subjects.

Figure 2

Bland–Altmann plot of the rebound tonometer (RBT) readings minus the Goldmann applanation tonometer (GAT) readings (mmHg) versus the mean of both in glaucoma patients.

CCT influence on IOP readings

Mean CCT was 561±38 μm (range 465–632 μm; median 567 μm) for normal subjects and 549±44 μm (range; 439–624 μm; median 546 μm) for glaucoma patients. In multiple linear regression analyses adjusted for age, IOP reading for glaucoma eyes was correlated positively with CCT (for the RBT, β=0.036 and P=0.002 and for the GAT, β=0.021 and P=0.033), whereas no correlation was found in normal eyes (for the RBT, β=0.005 and P=0.76 and for the GAT, β=0.015 and P=0.27) (Table 1).
Table 1

Multivariable analysis of factors associated with RBT-IOP and GAT-IOP in normal subjects and glaucoma patientsa

Normal subjects
Glaucoma patients
RBT-IOPGAT-IOPRBT-IOPGAT-IOP
Age, y0.14 (0.43)0.076 (0.57)0.016 (0.87)0.011 (0.90)
CCT, μm0.005 (0.76)0.015 (0.27)0.036 (0.002)0.021 (0.034)

Note:

Data are shown as regression coefficient (P-value).

Abbreviations: CCT, central corneal thickness; GAT-IOP, Goldmann applanation tonometer intraocular pressure; RBT-IOP, rebound tonometer intraocular pressure; y, years.

Discussion

The RBT is a handheld tonometer that does not require topical anesthesia, which has been proved to be useful in animal/human studies.42,48 Although many previous comparative studies have investigated its usability,21,25–41 our study included patients with the greatest mean age (81.9±4.8 years, overall). Previous studies demonstrated that the RBT instrument resulted in a slight overestimation (mean 0.9 mmHg) of IOP measurement in eyes in the presence or absence of glaucoma compared with the GAT.40 In contrast, Kontiola et al42 demonstrated that IOP readings using the GAT were higher than those using the RBT (estimated equation RBT =0.9 × GAT −4.8 mmHg). However, it should be considered that the RBT used in that study was a prototype; moreover, the background of the enrolled patients was unclear. For this reason, it may be too hard to make an assertion about the homologous comparison between the RBT and the GAT. In our study, the difference in IOP readings using the two instruments was 0.34 mmHg for normal subjects and 0.14 mmHg for glaucoma patients. These results would demonstrate that IOP measurements were not consistently overestimated or underestimated using the RBT compared with the GAT, suggesting that the RBT is useful and practicable compared with the GAT, even in late elderly subjects. The width of 95% limit of agreement was widely distributed (−4.3 mmHg to 6.1 mmHg) in a systematic review.40 Although a direct comparison cannot be made, our study showed a 95% limit of agreement within −3.3 mmHg to 4.0 mmHg (overall subjects), which was in the allowance range among the previous studies. CCT is known to influence IOP measurements when using the contact applanation tonometer.10–12,33,36,43–45 The IOP change could be somewhat overestimated in eyes with a thicker cornea. One study that investigated the influence of corneal properties (CCT, corneal hysteresis [CH], and corneal resistance factor [CRF]) on the RBT revealed that RBT measurements were affected by CH and CRF but not by CCT in multiple regression analysis.44 However, it can be difficult to measure the CH or CRF in the outpatient clinic, and the CCT measurement is a classic and convenient method for measuring corneal properties. Indeed, the IOP was positively correlated with the CCT in univariate analysis in the same study (r=0.5, P<0.01).44 Wong et al49 had reported that CCT did not correlate with IOP after being adjusted for age in elderly subjects aged over 60 years, although the demographic data of age and population are different from ours. Our result showed a positive correlation between CCT and the IOP readings in glaucoma eyes but not in normal eyes. The fact that correlation with CCT was observed in glaucoma eyes by each instrument reconfirms the importance of CCT measurements in glaucoma management. Although IOP shows a trend toward a subtle inverted U-shape relationship with increasing age,49 the prevalence of glaucoma increases in proportion to increasing age.50 The main advantage of the handheld tonometer is its utility in assessing some geriatric patients who cannot transfer to an examination table due to a physical disability or a wheelchair, or those with ocular disability such as blepharospasm. Although other portable devices apart from RBT may also be serviceable: eg, the Tono-Pen® (Tono-Pen XL, Reichert, Inc., Depew, NY, USA), Perkins tonometer (Haag-Streit USA, OH, USA), or pneumatonometer, (PTG, Reichert, Inc., Depew, NY, USA), they require calibration adjustment and topical anesthesia at the time of measurement. Use of these instruments should be considered in light of the advantages and disadvantages of each. Because most participants in this study were normal subjects or patients with normal tension glaucoma including treatment, the mean IOP difference could not help but be small. The difference between the RBT and the GAT is reported to become more obvious in cases with high IOP values.21 Future studies should increase the number and include subjects with a varied range of IOP. Furthermore, because other factors related to IOP (such as blood pressure, refraction, and body mass index) are reported,51–53 we need to consider the result from various orientations in the future.

Conclusion

We evaluated the utility of the RBT in terms of comparing with the GAT in late elderly subjects. A slight difference in IOP readings appeared to be within the allowable range. Careful attention is required when interpreting IOP readings in glaucoma eyes, because both instruments were positively correlated with the CCT. Our findings might suggest the utility of this handheld tonometer in the late elderly, especially for those in whom it may be difficult to perform measurements using the GAT.
  51 in total

1.  Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers.

Authors:  Makoto Nakamura; Urtogtah Darhad; Yasuko Tatsumi; Miyuki Fujioka; Azusa Kusuhara; Hidetaka Maeda; Akira Negi
Journal:  Am J Ophthalmol       Date:  2006-08       Impact factor: 5.258

2.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

3.  The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness.

Authors:  Ali Poostchi; Robert Mitchell; Simon Nicholas; Gordon Purdie; Anthony Wells
Journal:  Clin Exp Ophthalmol       Date:  2009-09       Impact factor: 4.207

4.  Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice.

Authors:  Jose M Martinez-de-la-Casa; Maria Jimenez-Santos; Federico Saenz-Frances; Maria Matilla-Rodero; Carmen Mendez-Hernandez; Rocio Herrero-Vanrell; Julian Garcia-Feijoo
Journal:  Acta Ophthalmol       Date:  2009-11-09       Impact factor: 3.761

Review 5.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

6.  Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients.

Authors:  Paolo Brusini; Maria Letizia Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

Review 7.  Sources of error with use of Goldmann-type tonometers.

Authors:  M M Whitacre; R Stein
Journal:  Surv Ophthalmol       Date:  1993 Jul-Aug       Impact factor: 6.048

8.  Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures. Collaborative Normal-Tension Glaucoma Study Group.

Authors: 
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

9.  Comparison of the rebound tonometer with the Goldmann applanation tonometer in glaucoma patients.

Authors:  Afsun Sahin; Leyla Niyaz; Nilgun Yildirim
Journal:  Clin Exp Ophthalmol       Date:  2007 May-Jun       Impact factor: 4.207

10.  The influence of central corneal thickness and corneal curvature on intraocular pressure measured by tono-pen and rebound tonometer in children.

Authors:  Afsun Sahin; Hikmet Basmak; Nilgun Yildirim
Journal:  J Glaucoma       Date:  2008 Jan-Feb       Impact factor: 2.503

View more
  1 in total

1.  Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons.

Authors:  Saulius Galgauskas; Rasa Strupaite; Ernesta Strelkauskaite; Rimvydas Asoklis
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.