| Literature DB >> 19638227 |
Pierre Squara1, Dominique Rotcajg, Dominique Denjean, Philippe Estagnasie, Alain Brusset.
Abstract
INTRODUCTION: This study was designed to test the hypothesis of equivalence in cardiac output (CO) and stroke volume (SV) monitoring capabilities of two devices: non invasive transthoracic bioreactance (NICOM), and a pulse contour analysis (PICCO PC) coupled to transpulmonary thermodilution (PICCO TD).Entities:
Mesh:
Year: 2009 PMID: 19638227 PMCID: PMC2750178 DOI: 10.1186/cc7981
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Figure 1Typical original recordings of PICCO PC, PICCO TD and NICOM during the 30-minute study protocol. Recordings included the baseline period (minutes 0 to 10), during positive end expiratory pressure (PEEP; minutes 10 to 20), and following PEEP removal (minutes 20 to 30). CO = cardiac output.
Inter technology agreements based on 20 values for each technology
| NICOM vs. PICCO TD | PICCO PC vs. PICCO TD | NICOM vs. PICCO PC | |
|---|---|---|---|
| 0,26 | 0,04 | 0,22 | |
| 0,76 | 0,77 | 0,55 | |
| 0,12 | 0,11 | 0,18 | |
| 1,52 | 1,77 | 2,41 | |
| 0,31 | 0,38 | 0,49 | |
| 0,93 | 0,93 | 0,80 |
Values of cardiac output obtained during positive end expiratory pressure were excluded from this analysis to minimize the impact of differences in time responsiveness between the devices.
R = coefficient of determination for inter-technologies comparisons; RE = averaged value of relative error (absolute bias/mean); LOA = limits of agreements of the Bland and Altman representation (two × standard deviations of the difference) in L/min; CV = coefficient of variation (two × standard deviation/mean); C&C = percentage of the differences inside the 30% limit of acceptability as suggested by Critchley and Critchley [16].
Figure 2Comparison of cardiac output measured by NICOM and PICCO PC in comparison of PICCO TD. The three different colours represent baseline, positive end expiratory pressure (PEEP) application, and return to baseline. The regression lines did not differ significantly from the line of identity (PICCO TD vs. NICOM: y = 1.2 + 0.77x, r = 0.77; PICCO TD vs. PICCO PC: y = -0.5 + 0.9x, r = 0.79).
Figure 3Bland-Altman plots comparing NICOM and PICCO PC to PICCO TD. The three different colours represent baseline, positive end expiratory pressure application, and return to baseline. Mean bias and limits of agreements are given in Table 1. Mean bias and limits of agreements are 0.22 ± 1.67 L/min/m for NICOM and 0.01 ± 1.86 L/min/m for PICCO PC.