First demonstration of combined kV/MV image-guided real-time dynamic multileaf-collimator target tracking.

Byungchul Cho1, Per R Poulsen, Alex Sloutsky, Amit Sawant, Paul J Keall.   

Abstract

PURPOSE: For intrafraction motion management, a real-time tracking system was developed by combining fiducial marker-based tracking via simultaneous kilovoltage (kV) and megavoltage (MV) imaging and a dynamic multileaf collimator (DMLC) beam-tracking system. METHODS AND MATERIALS: The integrated tracking system employed a Varian Trilogy system equipped with kV/MV imaging systems and a Millennium 120-leaf MLC. A gold marker in elliptical motion (2-cm superior-inferior, 1-cm left-right, 10 cycles/min) was simultaneously imaged by the kV and MV imagers at 6.7 Hz and segmented in real time. With these two-dimensional projections, the tracking software triangulated the three-dimensional marker position and repositioned the MLC leaves to follow the motion. Phantom studies were performed to evaluate time delay from image acquisition to MLC adjustment, tracking error, and dosimetric impact of target motion with and without tracking.
RESULTS: The time delay of the integrated tracking system was approximately 450 ms. The tracking error using a prediction algorithm was 0.9 +/- 0.5 mm for the elliptical motion. The dose distribution with tracking showed better target coverage and less dose to surrounding region over no tracking. The failure rate of the gamma test (3%/3-mm criteria) was 22.5% without tracking but was reduced to 0.2% with tracking.
CONCLUSION: For the first time, a complete tracking system combining kV/MV image-guided target tracking and DMLC beam tracking was demonstrated. The average geometric error was less than 1 mm, and the dosimetric error was negligible. This system is a promising method for intrafraction motion management.

Entities:  

Mesh:

Year:  2009        PMID: 19480969      PMCID: PMC2720135          DOI: 10.1016/j.ijrobp.2009.02.012

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  24 in total

1.  Accuracy of a wireless localization system for radiotherapy.

Authors:  James M Balter; J Nelson Wright; Laurence J Newell; Barry Friemel; Steven Dimmer; Yuki Cheng; John Wong; Edward Vertatschitsch; Timothy P Mate
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-03-01       Impact factor: 7.038

2.  Evaluation of internal lung motion for respiratory-gated radiotherapy using MRI: Part I--correlating internal lung motion with skin fiducial motion.

Authors:  Nicholas Koch; H Helen Liu; George Starkschall; Marc Jacobson; Kenneth Forster; Zhongxing Liao; Ritsuko Komaki; Craig W Stevens
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-12-01       Impact factor: 7.038

3.  Determination of maximum leaf velocity and acceleration of a dynamic multileaf collimator: implications for 4D radiotherapy.

Authors:  K Wijesooriya; C Bartee; J V Siebers; S S Vedam; P J Keall
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

4.  The stability of mechanical calibration for a kV cone beam computed tomography system integrated with linear accelerator.

Authors:  Michael B Sharpe; Douglas J Moseley; Thomas G Purdie; Mohammad Islam; Jeffrey H Siewerdsen; David A Jaffray
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

5.  The correlation between internal and external markers for abdominal tumors: implications for respiratory gating.

Authors:  David P Gierga; Johanna Brewer; Gregory C Sharp; Margrit Betke; Christopher G Willett; George T Y Chen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-04-01       Impact factor: 7.038

6.  Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor.

Authors:  H Shirato; S Shimizu; K Kitamura; T Nishioka; K Kagei; S Hashimoto; H Aoyama; T Kunieda; N Shinohara; H Dosaka-Akita; K Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-09-01       Impact factor: 7.038

7.  Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors.

Authors:  Ross I Berbeco; Steve B Jiang; Gregory C Sharp; George T Chen; Hassan Mostafavi; Hiroki Shirato
Journal:  Phys Med Biol       Date:  2004-01-21       Impact factor: 3.609

8.  A fiducial detection algorithm for real-time image guided IMRT based on simultaneous MV and kV imaging.

Authors:  Weihua Mao; Nadeem Riaz; Louis Lee; Rodney Wiersma; Lei Xing
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

9.  Potential for reduced toxicity and dose escalation in the treatment of inoperable non-small-cell lung cancer: a comparison of intensity-modulated radiation therapy (IMRT), 3D conformal radiation, and elective nodal irradiation.

Authors:  Inga S Grills; Di Yan; Alvaro A Martinez; Frank A Vicini; John W Wong; Larry L Kestin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-11-01       Impact factor: 7.038

10.  Respiration tracking in radiosurgery.

Authors:  Achim Schweikard; Hiroya Shiomi; John Adler
Journal:  Med Phys       Date:  2004-10       Impact factor: 4.071

View more
  38 in total

1.  Feasibility of low-dose single-view 3D fiducial tracking concurrent with external beam delivery.

Authors:  Michael A Speidel; Brian P Wilfley; Annie Hsu; Dimitre Hristov
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

2.  The dosimetric impact of inversely optimized arc radiotherapy plan modulation for real-time dynamic MLC tracking delivery.

Authors:  Marianne Falk; Tobias Larsson; Paul Keall; Byung Chul Cho; Marianne Aznar; Stine Korreman; Per Poulsen; Per Munck Af Rosenschold
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

3.  Detailed analysis of latencies in image-based dynamic MLC tracking.

Authors:  Per Rugaard Poulsen; Byungchul Cho; Amit Sawant; Dan Ruan; Paul J Keall
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

4.  Use of dMLC for implementation of dynamic respiratory-gated radiation therapy.

Authors:  Eric W Pepin; Huanmei Wu; Hiroki Shirato
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

5.  Failure mode and effect analysis-based quality assurance for dynamic MLC tracking systems.

Authors:  Amit Sawant; Sonja Dieterich; Michelle Svatos; Paul Keall
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

6.  Feasibility study on inverse four-dimensional dose reconstruction using the continuous dose-image of EPID.

Authors:  Inhwan Jason Yeo; Jae Won Jung; Byong Yong Yi; Jong Oh Kim
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

7.  The potential of positron emission tomography for intratreatment dynamic lung tumor tracking: a phantom study.

Authors:  Jaewon Yang; Tokihiro Yamamoto; Samuel R Mazin; Edward E Graves; Paul J Keall
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

8.  Optimizing fiducial visibility on periodically acquired megavoltage and kilovoltage image pairs during prostate volumetric modulated arc therapy.

Authors:  Pengpeng Zhang; Laura Happersett; Bosky Ravindranath; Michael Zelefsky; Gig Mageras; Margie Hunt
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

9.  Experimental investigation of a general real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring.

Authors:  Byungchul Cho; Per Poulsen; Dan Ruan; Amit Sawant; Paul J Keall
Journal:  Phys Med Biol       Date:  2012-10-24       Impact factor: 3.609

10.  Couch and multileaf collimator tracking: A clinical feasibility study for pancreas and liver treatment.

Authors:  Lei Zhang; Thomas LoSasso; Pengpeng Zhang; Margie Hunt; Gig Mageras; Grace Tang
Journal:  Med Phys       Date:  2020-09-11       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.