Couch and multileaf collimator tracking: A clinical feasibility study for pancreas and liver treatment.

Lei Zhang1, Thomas LoSasso1, Pengpeng Zhang1, Margie Hunt1, Gig Mageras1, Grace Tang1.   

Abstract

PURPOSE: Real-time tumor tracking through active correction by the multileaf collimator or treatment couch offers a promising strategy to mitigate delivery uncertainty due to intrafractional tumor motion. This study evaluated the performance of MLC and couch tracking using the prototype iTools Tracking system in TrueBeam Developer Mode and the application for abdominal cancer treatments.
METHODS: Experiments were carried out using a phantom with embedded Calypso transponders and a motion simulation platform. Geometric evaluations were performed using a circular conformal field with sinusoidal traces and pancreatic tumor motion traces. Geometric tracking accuracy was retrospectively calculated by comparing the compensational MLC or couch motion extracted from machine log files to the target motion reconstructed from real-time MV and kV images. Dosimetric tracking accuracy was measured with radiochromic films using clinical abdominal VMAT plans and pancreatic tumor traces.
RESULTS: Geometrically, the root-mean-square errors for MLC tracking were 0.5 and 1.8 mm parallel and perpendicular to leaf travel direction, respectively. Couch tracking, in contrast, showed an average of 0.8 mm or less geometric error in all directions. Dosimetrically, both MLC and couch tracking reduced motion-induced local dose errors compared to no tracking. Evaluated with five pancreatic tumor motion traces, the average 2%/2 mm global gamma pass rate of eight clinical abdominal VMAT plans was 67.4% (range: 26.4%-92.7%) without tracking, which was improved to 86.0% (range: 67.9%-95.6%) with MLC tracking, and 98.1% (range: 94.9%-100.0%) with couch tracking. In 16 out of 40 deliveries with different plans and motion traces, MLC tracking did not achieve clinically acceptable dosimetric accuracy with 3%/3mm gamma pass rate below 95%.
CONCLUSIONS: This study demonstrated the capability of MLC and couch tracking to reduce motion-induced dose errors in abdominal cases using a prototype tracking system. Clinically significant dose errors were observed with MLC tracking for certain plans which could be attributed to the inferior MLC tracking accuracy in the direction perpendicular to leaf travel, as well as the interplay between motion tracking and plan delivery for highly modulated plans. Couch tracking outperformed MLC tracking with consistently high dosimetric accuracy in all plans evaluated, indicating its clinical potential in the treatment of abdominal cancers.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  MLC tracking; abdominal cancers; couch tracking; motion management; real-time tumor tracking

Mesh:

Year:  2020        PMID: 32757298      PMCID: PMC8330968          DOI: 10.1002/mp.14438

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  60 in total

1.  Breathing-synchronized radiotherapy program at the University of California Davis Cancer Center.

Authors:  H D Kubo; P M Len; S Minohara; H Mostafavi
Journal:  Med Phys       Date:  2000-02       Impact factor: 4.071

2.  Image-based dynamic multileaf collimator tracking of moving targets during intensity-modulated arc therapy.

Authors:  Per Rugaard Poulsen; Walther Fledelius; Byungchul Cho; Paul Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-03-06       Impact factor: 7.038

Review 3.  The CyberKnife Robotic Radiosurgery System in 2010.

Authors:  W Kilby; J R Dooley; G Kuduvalli; S Sayeh; C R Maurer
Journal:  Technol Cancer Res Treat       Date:  2010-10

4.  DMLC motion tracking of moving targets for intensity modulated arc therapy treatment: a feasibility study.

Authors:  Jens Zimmerman; Stine Korreman; Gitte Persson; Herb Cattell; Michelle Svatos; Amit Sawant; Raghu Venkat; David Carlson; Paul Keall
Journal:  Acta Oncol       Date:  2009       Impact factor: 4.089

5.  Geometric and dosimetric comparison of four intrafraction motion adaptation strategies for stereotactic liver radiotherapy.

Authors:  Saber Nankali; Esben S Worm; Rune Hansen; Britta Weber; Morten Høyer; Alireza Zirak; Per Rugaard Poulsen
Journal:  Phys Med Biol       Date:  2018-07-16       Impact factor: 3.609

6.  Geometric and dosimetric accuracy of dynamic tumor tracking during volumetric-modulated arc therapy using a gimbal mounted linac.

Authors:  Tomohiro Ono; Yuki Miyabe; Kunio Takahashi; Mami Akimoto; Nobutaka Mukumoto; Yoshitomo Ishihara; Mitsuhiro Nakamura; Takashi Mizowaki; Masahiro Hiraoka
Journal:  Radiother Oncol       Date:  2017-11-11       Impact factor: 6.280

7.  Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer.

Authors:  D Mah; J Hanley; K E Rosenzweig; E Yorke; L Braban; C C Ling; S A Leibel; G Mageras
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-11-01       Impact factor: 7.038

8.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

9.  Compensating for tumor motion by a 6-degree-of-freedom treatment couch: is patient tolerance an issue?

Authors:  Reinhart A Sweeney; Winfried Arnold; Eva Steixner; Meinhard Nevinny-Stickel; Peter Lukas
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-01-23       Impact factor: 7.038

10.  Investigation of motion sickness and inertial stability on a moving couch for intra-fraction motion compensation.

Authors:  Warren D D'Souza; Kathleen T Malinowski; Seth Van Liew; Gypsyamber D'Souza; Kristen Asbury; Thomas J McAvoy; Mohan Suntharalingam; William F Regine
Journal:  Acta Oncol       Date:  2009       Impact factor: 4.089

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.