Experimental investigation of a general real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring.

Byungchul Cho1, Per Poulsen, Dan Ruan, Amit Sawant, Paul J Keall.   

Abstract

The goal of this work was to experimentally quantify the geometric accuracy of a novel real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring for clinically realistic arc and static field treatment delivery and target motion conditions. A general method for real-time target localization using kV imaging and respiratory monitoring was developed. Each dimension of internal target motion T(x, y, z; t) was estimated from the external respiratory signal R(t) through the correlation between R(t(i)) and the projected marker positions p(x(p), y(p); t(i)) on kV images by a state-augmented linear model: T(x, y, z; t) = aR(t) + bR(t - τ) + c. The model parameters, a, b, c, were determined by minimizing the squared fitting error ∑‖p(x(p), y(p); t(i)) - P(θ(i)) · (aR(t(i)) + bR(t(i) - τ) + c)‖(2) with the projection operator P(θ(i)). The model parameters were first initialized based on acquired kV arc images prior to MV beam delivery. This method was implemented on a trilogy linear accelerator consisting of an OBI x-ray imager (operating at 1 Hz) and real-time position monitoring (RPM) system (30 Hz). Arc and static field plans were delivered to a moving phantom programmed with measured lung tumour motion from ten patients. During delivery, the localization method determined the target position and the beam was adjusted in real time via dynamic multileaf collimator (DMLC) adaptation. The beam-target alignment error was quantified by segmenting the beam aperture and a phantom-embedded fiducial marker on MV images and analysing their relative position. With the localization method, the root-mean-squared errors of the ten lung tumour traces ranged from 0.7-1.3 mm and 0.8-1.4 mm during the single arc and five-field static beam delivery, respectively. Without the localization method, these errors ranged from 3.1-7.3 mm. In summary, a general method for real-time target localization using kV imaging and respiratory monitoring has been experimentally investigated for arc and static field delivery. The average beam-target error was 1 mm.

Entities:  

Mesh:

Year:  2012        PMID: 23093356      PMCID: PMC3701090          DOI: 10.1088/0031-9155/57/22/7395

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  24 in total

1.  Tracking moving organs in real time.

Authors:  Martin J Murphy
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

2.  Development of a four-dimensional image-guided radiotherapy system with a gimbaled X-ray head.

Authors:  Yuichiro Kamino; Kenji Takayama; Masaki Kokubo; Yuichiro Narita; Etsuro Hirai; Noriyuki Kawawda; Takashi Mizowaki; Yasushi Nagata; Takehiro Nishidai; Masahiro Hiraoka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-07-03       Impact factor: 7.038

3.  Geometric accuracy of a real-time target tracking system with dynamic multileaf collimator tracking system.

Authors:  Paul J Keall; Herbert Cattell; Damodar Pokhrel; Sonja Dieterich; Kenneth H Wong; Martin J Murphy; S Sastry Vedam; Krishni Wijesooriya; Radhe Mohan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-08-01       Impact factor: 7.038

4.  Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers.

Authors:  Ross I Berbeco; Hassan Mostafavi; Gregory C Sharp; Steve B Jiang
Journal:  Phys Med Biol       Date:  2005-09-13       Impact factor: 3.609

5.  Real-time target position estimation using stereoscopic kilovoltage/megavoltage imaging and external respiratory monitoring for dynamic multileaf collimator tracking.

Authors:  Byungchul Cho; Per Rugaard Poulsen; Amit Sawant; Dan Ruan; Paul J Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-07-07       Impact factor: 7.038

6.  Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor.

Authors:  H Shirato; S Shimizu; K Kitamura; T Nishioka; K Kagei; S Hashimoto; H Aoyama; T Kunieda; N Shinohara; H Dosaka-Akita; K Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-09-01       Impact factor: 7.038

7.  Reduction of observer variation using matched CT-PET for lung cancer delineation: a three-dimensional analysis.

Authors:  Roel J H M Steenbakkers; Joop C Duppen; Isabelle Fitton; Kirsten E I Deurloo; Lambert J Zijp; Emile F I Comans; Apollonia L J Uitterhoeve; Patrick T R Rodrigus; Gijsbert W P Kramer; Johan Bussink; Katrien De Jaeger; José S A Belderbos; Peter J C M Nowak; Marcel van Herk; Coen R N Rasch
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-09-28       Impact factor: 7.038

8.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy.

Authors:  H Shirato; S Shimizu; T Kunieda; K Kitamura; M van Herk; K Kagei; T Nishioka; S Hashimoto; K Fujita; H Aoyama; K Tsuchiya; K Kudo; K Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-11-01       Impact factor: 7.038

9.  Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors.

Authors:  Ross I Berbeco; Steve B Jiang; Gregory C Sharp; George T Chen; Hassan Mostafavi; Hiroki Shirato
Journal:  Phys Med Biol       Date:  2004-01-21       Impact factor: 3.609

10.  On using an adaptive neural network to predict lung tumor motion during respiration for radiotherapy applications.

Authors:  Marcus Isaksson; Joakim Jalden; Martin J Murphy
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

View more
  5 in total

Review 1.  Review of Real-Time 3-Dimensional Image Guided Radiation Therapy on Standard-Equipped Cancer Radiation Therapy Systems: Are We at the Tipping Point for the Era of Real-Time Radiation Therapy?

Authors:  Paul J Keall; Doan Trang Nguyen; Ricky O'Brien; Pengpeng Zhang; Laura Happersett; Jenny Bertholet; Per R Poulsen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-04-14       Impact factor: 7.038

2.  Technical Note: 3D localization of lung tumors on cone beam CT projections via a convolutional recurrent neural network.

Authors:  Chuang Wang; Margie Hunt; Lei Zhang; Andreas Rimner; Ellen Yorke; Michael Lovelock; Xiang Li; Tianfang Li; Gig Mageras; Pengpeng Zhang
Journal:  Med Phys       Date:  2020-01-28       Impact factor: 4.071

3.  Design and validation of a MV/kV imaging-based markerless tracking system for assessing real-time lung tumor motion.

Authors:  Pengpeng Zhang; Margie Hunt; Arina B Telles; Hai Pham; Michael Lovelock; Ellen Yorke; Guang Li; Laura Happersett; Andreas Rimner; Gig Mageras
Journal:  Med Phys       Date:  2018-11-13       Impact factor: 4.071

4.  Optimization of training periods for the estimation model of three-dimensional target positions using an external respiratory surrogate.

Authors:  Hiraku Iramina; Mitsuhiro Nakamura; Yusuke Iizuka; Takamasa Mitsuyoshi; Yukinori Matsuo; Takashi Mizowaki; Ikuo Kanno
Journal:  Radiat Oncol       Date:  2018-04-19       Impact factor: 3.481

5.  Automatic patient positioning and gating window settings in respiratory-gated stereotactic body radiation therapy for pancreatic cancer using fluoroscopic imaging.

Authors:  Niclas Pettersson; Daniel Simpson; Todd Atwood; Jona Hattangadi-Gluth; James Murphy; Laura Cerviño
Journal:  J Appl Clin Med Phys       Date:  2018-01-27       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.