Optimizing fiducial visibility on periodically acquired megavoltage and kilovoltage image pairs during prostate volumetric modulated arc therapy.

Pengpeng Zhang1, Laura Happersett1, Bosky Ravindranath1, Michael Zelefsky1, Gig Mageras1, Margie Hunt1.   

Abstract

PURPOSE: Robust detection of implanted fiducials is essential for monitoring intrafractional motion during hypofractionated treatment. The authors developed a plan optimization strategy to ensure clear visibility of implanted fiducials and facilitate 3D localization during volumetric modulated arc therapy (VMAT).
METHODS: Periodic kilovoltage (kV) images were acquired at 20° gantry intervals and paired with simultaneously acquired 4.4° short arc megavoltage digital tomosynthesis (MV-DTS) to localize three fiducials during VMAT delivery for hypofractionated prostate cancer treatment. Beginning with the original optimized plan, control point segments where fiducials were consistently blocked by multileaf collimator (MLC) within each 4.4° MV-DTS interval were first identified. For each segment, MLC apertures were edited to expose the fiducial that led to the least increase in the cost function. Subsequently, MLC apertures of all control points not involved with fiducial visualization were reoptimized to compensate for plan quality losses and match the original dose-volume histogram. MV dose for each MV-DTS was also kept above 0.4 MU to ensure acceptable image quality. Different imaging (gantry) intervals and visibility margins around fiducials were also evaluated.
RESULTS: Fiducials were consistently blocked by the MLC for, on average, 36% of the imaging control points for five hypofractionated prostate VMAT plans but properly exposed after reoptimization. Reoptimization resulted in negligible dosimetric differences compared with original plans and outperformed simple aperture editing: on average, PTV D98 recovered from 87% to 94% of prescription, and PTV dose homogeneity improved from 9% to 7%. Without violating plan objectives and compromising delivery efficiency, the highest imaging frequency and largest margin that can be achieved are a 10° gantry interval, and 15 mm, respectively.
CONCLUSIONS: VMAT plans can be made to accommodate MV-kV imaging of fiducials. Fiducial visualization rate and workflow efficiency are significantly improved with an automatic modification and reoptimization approach.

Entities:  

Mesh:

Year:  2016        PMID: 27147314      PMCID: PMC4826385          DOI: 10.1118/1.4944737

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Optimization of collimator trajectory in volumetric modulated arc therapy: development and evaluation for paraspinal SBRT.

Authors:  Pengpeng Zhang; Laura Happersett; Yingli Yang; Yoshiya Yamada; Gig Mageras; Margie Hunt
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-02-19       Impact factor: 7.038

2.  Real-time 3D internal marker tracking during arc radiotherapy by the use of combined MV-kV imaging.

Authors:  W Liu; R D Wiersma; W Mao; G Luxton; L Xing
Journal:  Phys Med Biol       Date:  2008-11-28       Impact factor: 3.609

3.  Simulated real time image guided intrafraction tracking-delivery for hypofractionated prostate IMRT.

Authors:  Sabbir Hossain; Ping Xia; Cynthia Chuang; Lynn Verhey; Alexander R Gottschalk; Guanwei Mu; Lijun Ma
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

4.  Combined kV and MV imaging for real-time tracking of implanted fiducial markers.

Authors:  R D Wiersma; Weihua Mao; L Xing
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

5.  Image-guided radiotherapy in near real time with intensity-modulated radiotherapy megavoltage treatment beam imaging.

Authors:  Weihua Mao; Annie Hsu; Nadeem Riaz; Louis Lee; Rodney Wiersma; Gary Luxton; Christopher King; Lei Xing; Timothy Solberg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-10-01       Impact factor: 7.038

6.  Considering marker visibility during leaf sequencing for segmental intensity-modulated radiation therapy.

Authors:  Bo Zhao; Jianrong Dai; C Clifton Ling
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

7.  Four-dimensional inverse treatment planning with inclusion of implanted fiducials in IMRT segmented fields.

Authors:  Yunzhi Ma; Louis Lee; O Keshet; Paul Keall; Lei Xing
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

8.  Intrafractional 3D localization using kilovoltage digital tomosynthesis for sliding-window intensity modulated radiation therapy.

Authors:  Pengpeng Zhang; Margie Hunt; Hai Pham; Grace Tang; Gig Mageras
Journal:  Phys Med Biol       Date:  2015-08-25       Impact factor: 3.609

9.  First demonstration of combined kV/MV image-guided real-time dynamic multileaf-collimator target tracking.

Authors:  Byungchul Cho; Per R Poulsen; Alex Sloutsky; Amit Sawant; Paul J Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-07-01       Impact factor: 7.038

10.  A fiducial detection algorithm for real-time image guided IMRT based on simultaneous MV and kV imaging.

Authors:  Weihua Mao; Nadeem Riaz; Louis Lee; Rodney Wiersma; Lei Xing
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

View more
  2 in total

1.  Simultaneous MV-kV imaging for intrafractional motion management during volumetric-modulated arc therapy delivery.

Authors:  Margie A Hunt; Mark Sonnick; Hai Pham; Rajesh Regmi; Jian-ping Xiong; Daniel Morf; Gig S Mageras; Michael Zelefsky; Pengpeng Zhang
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

2.  Retrospective analysis of MV-kV imaging-based fiducial tracking in prostate SBRT treatment.

Authors:  Kevin Crotteau; Wei Lu; Sean Berry; Laura Happersett; Sarah Burleson; Weixing Cai
Journal:  J Appl Clin Med Phys       Date:  2022-03-26       Impact factor: 2.243

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.