Abstract

PURPOSE: This study aims to evaluate the potential and feasibility of positron emission tomography for dynamic lung tumor tracking during radiation treatment. The authors propose a center of mass (CoM) tumor tracking algorithm using gated-PET images combined with a respiratory monitor and investigate the geometric accuracy of the proposed algorithm.
METHODS: The proposed PET dynamic lung tumor tracking algorithm estimated the target position information through the CoM of the segmented target volume on gated PET images reconstructed from accumulated coincidence events. The information was continuously updated throughout a scan based on the assumption that real-time processing was supported (actual processing time at each frame ≈ 10 s). External respiratory motion and list-mode PET data were acquired from a phantom programmed to move with measured respiratory traces (external respiratory motion and internal target motion) from human subjects, for which the ground truth target position was known as a function of time. The phantom was cylindrical with six hollow sphere targets (10, 13, 17, 22, 28, and 37 mm in diameter). The measured respiratory traces consisted of two sets: (1) 1D-measured motion from ten healthy volunteers and (2) 3D-measured motion from four lung cancer patients. The authors evaluated the geometric accuracy of the proposed algorithm by quantifying estimation errors (Euclidean distance) between the actual motion of targets (1D-motion and 3D-motion traces) and CoM trajectories estimated by the proposed algorithm as a function of time.
RESULTS: The time-averaged error of 1D-motion traces over all trajectories of all targets was 1.6 mm. The error trajectories decreased with time as coincidence events were accumulated. The overall error trajectory of 1D-motion traces converged to within 2 mm in approximately 90 s. As expected, more accurate results were obtained for larger targets. For example, for the 37 mm target, the average error over all 1D-motion traces was 1.1 mm; and for the 10 mm target, the average error over all 1D-motion traces was 2.8 mm. The overall time-averaged error of 3D-motion traces was 1.6 mm, which was comparable to that of the 1D-motion traces. There were small variations in the errors between the 3D-motion traces, although the motion trajectories were very different. The accuracy of the estimates was consistent for all targets except for the smallest.
CONCLUSIONS: The authors developed an algorithm for dynamic lung tumor tracking using list-mode PET data and a respiratory motion signal, and demonstrated proof-of-principle for PET-guided lung tumor tracking. The overall tracking error in phantom studies is less than 2 mm.

Entities:  

Mesh:

Year:  2014        PMID: 24506609      PMCID: PMC3977800          DOI: 10.1118/1.4861816

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  69 in total

1.  Dynamic MLC tracking of moving targets with a single kV imager for 3D conformal and IMRT treatments.

Authors:  Per R Poulsen; Byungchul Cho; Amit Sawant; Dan Ruan; Paul J Keall
Journal:  Acta Oncol       Date:  2010-10       Impact factor: 4.089

2.  Design and performance of a respiratory amplitude gating device for PET/CT imaging.

Authors:  Guoping Chang; Tingting Chang; John W Clark; Osama R Mawlawi
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

3.  First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system.

Authors:  B G Fallone; B Murray; S Rathee; T Stanescu; S Steciw; S Vidakovic; E Blosser; D Tymofichuk
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

4.  Geometric accuracy of a novel gimbals based radiation therapy tumor tracking system.

Authors:  Tom Depuydt; Dirk Verellen; Olivier Haas; Thierry Gevaert; Nadine Linthout; Michael Duchateau; Koen Tournel; Truus Reynders; Katrien Leysen; Mischa Hoogeman; Guy Storme; Mark De Ridder
Journal:  Radiother Oncol       Date:  2011-03       Impact factor: 6.280

Review 5.  GPU computing in medical physics: a review.

Authors:  Guillem Pratx; Lei Xing
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

6.  MRI-guided tumor tracking in lung cancer radiotherapy.

Authors:  Laura I Cerviño; Jiang Du; Steve B Jiang
Journal:  Phys Med Biol       Date:  2011-05-31       Impact factor: 3.609

7.  Design study of an in situ PET scanner for use in proton beam therapy.

Authors:  S Surti; W Zou; M E Daube-Witherspoon; J McDonough; J S Karp
Journal:  Phys Med Biol       Date:  2011-04-05       Impact factor: 3.609

8.  Dynamic multileaf collimator tracking of respiratory target motion based on a single kilovoltage imager during arc radiotherapy.

Authors:  Per Rugaard Poulsen; Byungchul Cho; Dan Ruan; Amit Sawant; Paul J Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-02-03       Impact factor: 7.038

9.  Four-dimensional IMRT treatment planning using a DMLC motion-tracking algorithm.

Authors:  Yelin Suh; Amit Sawant; Raghu Venkat; Paul J Keall
Journal:  Phys Med Biol       Date:  2009-05-28       Impact factor: 3.609

10.  Extension of a data-driven gating technique to 3D, whole body PET studies.

Authors:  Paul J Schleyer; Michael J O'Doherty; Paul K Marsden
Journal:  Phys Med Biol       Date:  2011-06-10       Impact factor: 3.609

View more
  8 in total

1.  Quantitative analysis of multiple biokinetic models using a dynamic water phantom: A feasibility study.

Authors:  Fu-Tsai Chiang; Pei-Jung Li; Shih-Ping Chung; Lung-Fa Pan; Lung-Kwang Pan
Journal:  Bioengineered       Date:  2016-06-10       Impact factor: 3.269

2.  Real-time marker-less tumor tracking with TOF PET:in silicofeasibility study.

Authors:  Xinyi Cheng; Dongxu Yang; Yuncheng Zhong; Yiping Shao
Journal:  Phys Med Biol       Date:  2022-05-26       Impact factor: 4.174

3.  Feasibility of biology-guided radiotherapy using PSMA-PET to boost to dominant intraprostatic tumour.

Authors:  Mathieu Gaudreault; David Chang; Nicholas Hardcastle; Price Jackson; Tomas Kron; Michael S Hofman; Shankar Siva
Journal:  Clin Transl Radiat Oncol       Date:  2022-05-17

4.  Fundamental study on quality assurance (QA) procedures for a real-time tumor tracking radiotherapy (RTRT) system from the viewpoint of imaging devices.

Authors:  Suguru Kimura; Naoki Miyamoto; Kenneth L Sutherland; Ryusuke Suzuki; Hiroki Shirato; Masayori Ishikawa
Journal:  J Appl Clin Med Phys       Date:  2021-06-02       Impact factor: 2.102

Review 5.  An overview of current practice in external beam radiation oncology with consideration to potential benefits and challenges for nanotechnology.

Authors:  Raymond B King; Stephen J McMahon; Wendy B Hyland; Suneil Jain; Karl T Butterworth; Kevin M Prise; Alan R Hounsell; Conor K McGarry
Journal:  Cancer Nanotechnol       Date:  2017-02-03

6.  Utility of Biology-Guided Radiotherapy to De Novo Metastases Diagnosed During Staging of High-Risk Biopsy-Proven Prostate Cancer.

Authors:  Mathieu Gaudreault; David Chang; Nicholas Hardcastle; Price Jackson; Tomas Kron; Gerard G Hanna; Michael S Hofman; Shankar Siva
Journal:  Front Oncol       Date:  2022-04-12       Impact factor: 5.738

7.  The technical design and concept of a PET/CT linac for biology-guided radiotherapy.

Authors:  Oluwaseyi M Oderinde; Shervin M Shirvani; Peter D Olcott; Gopinath Kuduvalli; Samuel Mazin; David Larkin
Journal:  Clin Transl Radiat Oncol       Date:  2021-04-17

8.  Online Prostate-Specific Membrane Antigen and Positron Emission Tomography-Guided Radiation Therapy for Oligometastatic Prostate Cancer.

Authors:  William T Hrinivich; Ryan Phillips; Angela J Da Silva; Noura Radwan; Michael A Gorin; Steven P Rowe; Kenneth J Pienta; Martin G Pomper; John Wong; Phuoc T Tran; Ken Kang-Hsin Wang
Journal:  Adv Radiat Oncol       Date:  2019-11-06
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.