The stability of mechanical calibration for a kV cone beam computed tomography system integrated with linear accelerator.

Michael B Sharpe1, Douglas J Moseley, Thomas G Purdie, Mohammad Islam, Jeffrey H Siewerdsen, David A Jaffray.   

Abstract

The geometric accuracy and precision of an image-guided treatment system were assessed. Image guidance is performed using an x-ray volume imaging (XVI) system integrated with a linear accelerator and treatment planning system. Using an amorphous silicon detector and x-ray tube, volumetric computed tomography images are reconstructed from kilovoltage radiographs by filtered backprojection. Image fusion and assessment of geometric targeting are supported by the treatment planning system. To assess the limiting accuracy and precision of image-guided treatment delivery, a rigid spherical target embedded in an opaque phantom was subjected to 21 treatment sessions over a three-month period. For each session, a volumetric data set was acquired and loaded directly into an active treatment planning session. Image fusion was used to ascertain the couch correction required to position the target at the prescribed iso-center. Corrections were validated independently using megavoltage electronic portal imaging to record the target position with respect to symmetric treatment beam apertures. An initial calibration cycle followed by repeated image-guidance sessions demonstrated the XVI system could be used to relocate an unambiguous object to within less than 1 mm of the prescribed location. Treatment could then proceed within the mechanical accuracy and precision of the delivery system. The calibration procedure maintained excellent spatial resolution and delivery precision over the duration of this study, while the linear accelerator was in routine clinical use. Based on these results, the mechanical accuracy and precision of the system are ideal for supporting high-precision localization and treatment of soft-tissue targets.

Entities:  

Mesh:

Year:  2006        PMID: 16485420     DOI: 10.1118/1.2143141

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  37 in total

1.  Quantifying the gantry sag on linear accelerators and introducing an MLC-based compensation strategy.

Authors:  Weiliang Du; Song Gao; Xiaochun Wang; Rajat J Kudchadker
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

Review 2.  Flat-detector computed tomography (FD-CT).

Authors:  Willi A Kalender; Yiannis Kyriakou
Journal:  Eur Radiol       Date:  2007-06-23       Impact factor: 5.315

3.  Precision radiotherapy for small animal research.

Authors:  Mohammad Matinfar; Iulian Iordachita; Eric Ford; John Wong; Peter Kazanzides
Journal:  Med Image Comput Comput Assist Interv       Date:  2008

Review 4.  Anniversary paper. Development of x-ray computed tomography: the role of medical physics and AAPM from the 1970s to present.

Authors:  Xiaochuan Pan; Jeffrey Siewerdsen; Patrick J La Riviere; Willi A Kalender
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

5.  Geometric calibration of a mobile C-arm for intraoperative cone-beam CT.

Authors:  M J Daly; J H Siewerdsen; Y B Cho; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

6.  Development of a QA phantom and automated analysis tool for geometric quality assurance of on-board MV and kV x-ray imaging systems.

Authors:  Weihua Mao; Louis Lee; Lei Xing
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

7.  Estimation of CT cone-beam geometry using a novel method insensitive to phantom fabrication inaccuracy: implications for isocenter localization accuracy.

Authors:  J Chetley Ford; Dandan Zheng; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

8.  Bow-tie wobble artifact: effect of source assembly motion on cone-beam CT.

Authors:  Dandan Zheng; John C Ford; Jun Lu; Dimitrios Lazos; Geoffrey D Hugo; Damodar Pokhrel; Lisha Zhang; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

9.  Transmission characteristics of a two dimensional antiscatter grid prototype for CBCT.

Authors:  Cem Altunbas; Brian Kavanagh; Timur Alexeev; Moyed Miften
Journal:  Med Phys       Date:  2017-06-16       Impact factor: 4.071

10.  Image-guided small animal radiation research platform: calibration of treatment beam alignment.

Authors:  Mohammad Matinfar; Eric Ford; Iulian Iordachita; John Wong; Peter Kazanzides
Journal:  Phys Med Biol       Date:  2009-01-14       Impact factor: 3.609