Crystal identification in positron emission tomography using nonrigid registration to a Fourier-based template.

Abhijit J Chaudhari1, Anand A Joshi, Spencer L Bowen, Richard M Leahy, Simon R Cherry, Ramsey D Badawi.   

Abstract

Modern positron emission tomography (PET) detectors are typically made from 2D modular arrays of scintillation crystals. Their characteristic flood field response (or flood histogram) must be segmented in order to correctly determine the crystal of annihilation photon interaction in the system. Crystal identification information thus generated is also needed for accurate system modeling as well as for detailed detector characterization and performance studies. In this paper, we present a semi-automatic general purpose template-guided scheme for the segmentation of flood histograms. We first generate a template image that exploits the spatial frequency information in the given flood histogram using Fourier-space analysis. This template image is a lower order approximation of the flood histogram, and can be segmented with horizontal and vertical lines drawn midway between adjacent peaks in the histogram. The template is then registered to the given flood histogram by a diffeomorphic polynomial-based warping scheme that is capable of iteratively minimizing intensity differences. The displacement field thus calculated is applied to the segmentation of the template resulting in a segmentation of the given flood histogram. We evaluate our segmentation scheme for a photomultiplier tube based PET detector, a detector with readout by a position-sensitive avalanche photodiode (PSAPD) and a detector consisting of a stack of photomultiplier tubes and scintillator arrays. Further, we quantitatively compare the performance of the proposed method to that of a manual segmentation scheme using reconstructed images of a line-source phantom. We also present an adaptive method for distortion reduction in flood histograms obtained for PET detectors that use PSAPDs.

Entities:  

Mesh:

Year:  2008        PMID: 18723924      PMCID: PMC2748910          DOI: 10.1088/0031-9155/53/18/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  12 in total

Review 1.  From PET detectors to PET scanners.

Authors:  John L Humm; Anatoly Rosenfeld; Alberto Del Guerra
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-10-02       Impact factor: 9.236

2.  SCINTILLATION CAMERA WITH MULTICHANNEL COLLIMATORS.

Authors:  H O ANGER
Journal:  J Nucl Med       Date:  1964-07       Impact factor: 10.057

3.  The performance of the single-slice rebinning technique for imaging the human striatum as evaluated by phantom studies.

Authors:  V Sossi; M W Stazyk; P E Kinahan; T J Ruth
Journal:  Phys Med Biol       Date:  1994-03       Impact factor: 3.609

Review 4.  Positron emission tomography.

Authors:  Gerd Muehllehner; Joel S Karp
Journal:  Phys Med Biol       Date:  2006-06-20       Impact factor: 3.609

5.  Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes.

Authors:  Yongfeng Yang; Purushottam A Dokhale; Robert W Silverman; Kanai S Shah; Mickel A McClish; Richard Farrell; Gerald Entine; Simon R Cherry
Journal:  Phys Med Biol       Date:  2006-04-10       Impact factor: 3.609

6.  Use of digital front-end electronics for optimization of a modular PET detector.

Authors:  P D Cutler; E J Hoffman
Journal:  IEEE Trans Med Imaging       Date:  1994       Impact factor: 10.048

7.  A New Positioning Algorithm for Position-Sensitive Avalanche Photodiodes.

Authors:  Jin Zhang; Peter D Olcott; Craig S Levin
Journal:  IEEE Trans Nucl Sci       Date:  2007-06-01       Impact factor: 1.679

8.  Treatment of axial data in three-dimensional PET.

Authors:  M E Daube-Witherspoon; G Muehllehner
Journal:  J Nucl Med       Date:  1987-11       Impact factor: 10.057

9.  Investigation of Depth of Interaction Encoding for a Pixelated LSO Array with a Single Multi-Channel PMT.

Authors:  Yongfeng Yang; Yibao Wu; Simon R Cherry
Journal:  IEEE Trans Nucl Sci       Date:  2009-10-07       Impact factor: 1.679

10.  A prototype PET scanner with DOI-encoding detectors.

Authors:  Yongfeng Yang; Yibao Wu; Jinyi Qi; Sara St James; Huini Du; Purushottam A Dokhale; Kanai S Shah; Richard Farrell; Simon R Cherry
Journal:  J Nucl Med       Date:  2008-06-13       Impact factor: 10.057

View more
  8 in total

1.  PET Performance Evaluation of an MR-Compatible PET Insert.

Authors:  Yibao Wu; Ciprian Catana; Richard Farrell; Purushottam A Dokhale; Kanai S Shah; Jinyi Qi; Simon R Cherry
Journal:  IEEE Trans Nucl Sci       Date:  2009-06-16       Impact factor: 1.679

2.  A fast method for optical simulation of flood maps of light-sharing detector modules.

Authors:  Han Shi; Dong Du; JianFeng Xu; William W Moses; Qiyu Peng
Journal:  Nucl Instrum Methods Phys Res A       Date:  2015-09-03       Impact factor: 1.455

3.  An Advanced 100-Channel Readout System for Nuclear Imaging.

Authors:  Zhixiang Zhao; Siwei Xie; Xi Zhang; Jingwu Yang; Qiu Huang; Jianfeng Xu; Qiyu Peng
Journal:  IEEE Trans Instrum Meas       Date:  2018-11-15       Impact factor: 4.016

4.  A Study of Position-Sensitive Solid-State Photomultiplier Signal Properties.

Authors:  Jeffrey P Schmall; Junwei Du; Martin S Judenhofer; Purushottam Dokhale; James Christian; Mickel McClish; Kanai S Shah; Simon R Cherry
Journal:  IEEE Trans Nucl Sci       Date:  2014-06-12       Impact factor: 1.679

5.  PET characteristics of a dedicated breast PET/CT scanner prototype.

Authors:  Yibao Wu; Spencer L Bowen; Kai Yang; Nathan Packard; Lin Fu; George Burkett; Jinyi Qi; John M Boone; Simon R Cherry; Ramsey D Badawi
Journal:  Phys Med Biol       Date:  2009-06-17       Impact factor: 3.609

6.  A study of the timing properties of position-sensitive avalanche photodiodes.

Authors:  Yibao Wu; Thomas S C Ng; Yongfeng Yang; Kanai Shah; Richard Farrell; Simon R Cherry
Journal:  Phys Med Biol       Date:  2009-08-11       Impact factor: 3.609

7.  Characterization of a high-resolution hybrid DOI detector for a dedicated breast PET/CT scanner.

Authors:  Felipe Godinez; Abhijit J Chaudhari; Yongfeng Yang; Richard Farrell; Ramsey D Badawi
Journal:  Phys Med Biol       Date:  2012-05-11       Impact factor: 3.609

8.  Spatial distortion correction and crystal identification for MRI-compatible position-sensitive avalanche photodiode-based PET scanners.

Authors:  Abhijit J Chaudhari; Anand A Joshi; Yibao Wu; Richard M Leahy; Simon R Cherry; Ramsey D Badawi
Journal:  IEEE Trans Nucl Sci       Date:  2009-06-01       Impact factor: 1.679

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.