[Evaluation of dynamic contour tonometry in keratoconus].

N Schädle1, J D Unterlauft, T Klink, G Geerling.   

Abstract

PURPOSE: The gold standard for measuring intraocular pressure (IOP) until now has been Goldmann applanation tonometry (GAT), which depends on the central corneal thickness (CCT) and curvature. In patients with keratoconus who have an abnormal corneal geometry and thickness, measurement of pressure with GAT is often difficult and not very reproducible. We compared the impact of the central corneal thickness (CCT) on the IOP measured with dynamic contour tonometry (DCT), a digital method which is adapted to the corneal geometry, and GAT in patients with keratoconus.
METHODS: IOP was measured in 54 patients (38 men and 16 women, mean age of 36+/-9.9 years) with GAT and DCT in randomized order. All patients had a keratoconus which was assured by topography. In addition central corneal thickness (CCT) was measured with the Pentacam. For statistical analysis the Pearson correlation was calculated and a Bland-Altman diagram plotted.
RESULTS: Mean corneal thickness was 486.2+/-45.5 microm. DCT measured the IOP at a mean value of 14.9+/-2.6 mmHg and GAT at 13.3+/-2.9 mmHg. With a mean difference of 1.6+/-2.4 mmHg DCT measured significantly higher than GAT (p<or=0.05) Neither IOP measurements with GAT (r=-0.03; p>0.05) nor those with DCT (r=0.08; p>0.05) showed a significant correlation to central corneal thickness.
CONCLUSION: The example of keratoconus confirms that IOP measured by GAT is lower than if measured by DCT. Because both methods are independent of the CCT they are equally acceptable for IOP follow-up in eyes with keratoconus, which may result in progressive corneal thinning in the long term.

Entities:  

Mesh:

Year:  2009        PMID: 18670779     DOI: 10.1007/s00347-008-1812-1

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  21 in total

1.  Correlation between intraocular pressure, central corneal thickness, stage of glaucoma, and demographic patient data: prospective analysis of biophysical parameters in tertiary glaucoma practice populations.

Authors:  Christoph Kniestedt; Shan Lin; Joyce Choe; Michelle Nee; Alan Bostrom; Jörg Stürmer; Robert L Stamper
Journal:  J Glaucoma       Date:  2006-04       Impact factor: 2.503

2.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

3.  Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry.

Authors:  Markus Kohlhaas; Andreas G Boehm; Eberhard Spoerl; Antje Pürsten; Hans J Grein; Lutz E Pillunat
Journal:  Arch Ophthalmol       Date:  2006-04

4.  Intraocular pressure measurement-comparison of dynamic contour tonometry and goldmann applanation tonometry.

Authors:  Evelin Schneider; Franz Grehn
Journal:  J Glaucoma       Date:  2006-02       Impact factor: 2.503

5.  [Measuring intraocular pressure in keratoconus. Effect of the changed biomechanics].

Authors:  A Böhm; M Kohlhaas; R C Lerche; B Bischoff; G Richard
Journal:  Ophthalmologe       Date:  1997-11       Impact factor: 1.059

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Dynamic contour tonometry: a new way to assess intraocular pressure in ectatic corneas.

Authors:  Zeynep Ozbek; Elisabeth J Cohen; Kristin M Hammersmith; Christopher J Rapuano
Journal:  Cornea       Date:  2006-09       Impact factor: 2.651

8.  Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis.

Authors:  Claude Kaufmann; Lucas M Bachmann; Michael A Thiel
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-09       Impact factor: 4.799

9.  Central corneal thickness as a risk factor for advanced glaucoma damage.

Authors:  Leon W Herndon; Jennifer S Weizer; Sandra S Stinnett
Journal:  Arch Ophthalmol       Date:  2004-01

10.  Dynamic contour tonometry and goldman applanation tonometry in eyes with keratoconus.

Authors:  Jackson Barreto; Mirko Babic; Roberto Murad Vessani; Remo Susanna
Journal:  Clinics (Sao Paulo)       Date:  2006-12       Impact factor: 2.365

View more
  2 in total

1.  [Glaucoma and corneal transplantation].

Authors:  G Geerling; M Müller; M Zierhut; T Klink
Journal:  Ophthalmologe       Date:  2010-05       Impact factor: 1.059

2.  Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus.

Authors:  Fatih Özcura; Nilgün Yıldırım; Emre Tambova; Afsun Şahin
Journal:  J Optom       Date:  2016-07-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.