Literature DB >> 18389666

Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients.

Jacoline Steinfelder-Visscher1, Steven Teerenstra, Jacqueline M T Klein Gunnewiek, Patrick W Weerwind.   

Abstract

Point-of-care analyzers may benefit therapeutic decision making by reducing turn-around-time for samples. This is especially true when biochemical parameters exceed the clinical reference range, in which acute and effective treatment is essential. We therefore evaluated the analytical performance of the i-STAT point-of-care analyzer in two critically ill adult patient populations. During a 3-month period, 48 blood samples from patients undergoing cardiac surgery with cardiopulmonary bypass (CPB) and 42 blood samples from non-cardiac patients who needed intensive care treatment were analyzed on both the i-STAT analyzer (CPB and non-CPB mode, respectively) and our laboratory analyzers (RapidLab 865/Sysmex XE-2100 instrument). The agreement analysis for quantitative data was used to compare i-STAT to RapidLab for blood gas/electrolytes and for hematocrit with the Sysmex instrument. Point-of-care electrolytes and blood gases had constant deviation, except for pH, pO2, and hematocrit. A clear linear trend in deviation of i-STAT from RapidLab was noticed for pH during CPB (r = 0.32, p = .03) and for pO2 > 10 kPa during CPB (r = -0.59, p < .0001 when 10 <pO2 < 30 kPa) and in the intensive care unit (r = -0.61,p < .001 when 10 < pO2 < 30 kPa). In the normal pO2 range (10.6 < pO2 < 13.3 kPa), the performance of the i-STAT was comparable to the RapidLab. In contrast to hematocrit measured during CPB, hematocrit using the non-CPB mode in the non-cardiac intensive care population showed an underestimation up to 2.2% (p < .0001) in the hematocrit range below 25% (n = 11) using the i-STAT. The i-STAT analyzer is suitable for point-of-care testing of electrolytes and blood gases in critically ill patients, except for high pO2. However, the discrepancy in hematocrit bias shows that accuracy established in one patient population cannot be automatically extrapolated to other patient populations, thus stressing the need for separate evaluation.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18389666      PMCID: PMC4680657     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  16 in total

1.  The portable laboratory: an evaluation of the accuracy and reproducibility of i-STAT.

Authors:  D Bingham; J Kendall; M Clancy
Journal:  Ann Clin Biochem       Date:  1999-01       Impact factor: 2.057

Review 2.  Point of care technologies.

Authors:  Hugh C Gilbert; Joseph W Szokol
Journal:  Int Anesthesiol Clin       Date:  2004

3.  Impact of point-of-care testing on patients' length of stay in a large emergency department.

Authors:  C A Parvin; S F Lo; S M Deuser; L G Weaver; L M Lewis; M G Scott
Journal:  Clin Chem       Date:  1996-05       Impact factor: 8.327

4.  The use of the iSTAT portable analyzer in patients undergoing cardiopulmonary bypass.

Authors:  N R Connelly; M Magee; B Kiessling
Journal:  J Clin Monit       Date:  1996-07

5.  Accuracy and precision of the i-STAT portable clinical analyzer: an analytical point of view.

Authors:  P Pidetcha; S Ornvichian; S Chalachiva
Journal:  J Med Assoc Thai       Date:  2000-04

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Point of care laboratory testing in the emergency department.

Authors:  P Jatlow
Journal:  Am J Clin Pathol       Date:  1993-12       Impact factor: 2.493

8.  Analytical evaluation of i-STAT Portable Clinical Analyzer and use by nonlaboratory health-care professionals.

Authors:  E Jacobs; E Vadasdi; L Sarkozi; N Colman
Journal:  Clin Chem       Date:  1993-06       Impact factor: 8.327

9.  Accuracy of the i-STAT bedside blood gas analyser.

Authors:  S Sediame; F Zerah-Lancner; M P d'Ortho; S Adnot; A Harf
Journal:  Eur Respir J       Date:  1999-07       Impact factor: 16.671

10.  Evaluation of the i-STAT Portable Clinical Analyzer for point-of-care blood testing in the intensive care units of a university children's hospital.

Authors:  Christine Papadea; Joyce Foster; Sharon Grant; Sandra A Ballard; John C Cate; W Michael Southgate; Dilip M Purohit
Journal:  Ann Clin Lab Sci       Date:  2002       Impact factor: 1.256

View more
  6 in total

1.  Evaluation of point-of-care analyzers' ability to reduce bias in conductivity-based hematocrit measurement during cardiopulmonary bypass.

Authors:  Steven Teerenstra; Jacoline Steinfelder-Visscher; Jacqueline Klein Gunnewiek; Patrick W Weerwind
Journal:  J Clin Monit Comput       Date:  2013-08-30       Impact factor: 2.502

2.  Design and Utility of a Point-of-Care Microfluidic Platform to Assess Hematocrit and Blood Coagulation.

Authors:  Jevgenia Zilberman-Rudenko; Rachel M White; Dmitriy A Zilberman; Hari H S Lakshmanan; Rachel A Rigg; Joseph J Shatzel; Jeevan Maddala; Owen J T McCarty
Journal:  Cell Mol Bioeng       Date:  2018-07-19       Impact factor: 2.321

3.  Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: a cross-sectional method comparison study.

Authors:  Arpa Chutipongtanate; Churairat Yasaeng; Tanit Virankabutra; Somchai Chutipongtanate
Journal:  BMC Anesthesiol       Date:  2020-04-22       Impact factor: 2.217

4.  Transcutaneous Flexible Sensor for In Vivo Photonic Detection of pH and Lactate.

Authors:  Dat Nguyen; Micah M Lawrence; Haley Berg; Monika Aya Lyons; Samir Shreim; Mark T Keating; John Weidling; Elliot L Botvinick
Journal:  ACS Sens       Date:  2022-02-17       Impact factor: 7.711

Review 5.  Significance of Cardiac Troponins as an Identification Tool in COVID-19 Patients Using Biosensors: An Update.

Authors:  Yousef Rasmi; Osama F Mosa; Shahriar Alipour; Nadia Heidari; Farzaneh Javanmard; Ali Golchin; Shiva Gholizadeh-Ghaleh Aziz
Journal:  Front Mol Biosci       Date:  2022-02-24

6.  Errors in potassium measurement: a laboratory perspective for the clinician.

Authors:  Jaya R Asirvatham; Viju Moses; Loring Bjornson
Journal:  N Am J Med Sci       Date:  2013-04
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.