Literature DB >> 35175733

Transcutaneous Flexible Sensor for In Vivo Photonic Detection of pH and Lactate.

Dat Nguyen1,2,3, Micah M Lawrence1,2,3, Haley Berg3, Monika Aya Lyons1,2,3, Samir Shreim2, Mark T Keating2, John Weidling2, Elliot L Botvinick1,2,3,4.   

Abstract

Clinical research shows that frequent measurements of both pH and lactate can help guide therapy and improve patient outcome. However, current methods of sampling blood pH and lactate make it impractical to take readings frequently (due to the heightened risk of blood infection and anemia). As a solution, we have engineered a subcutaneous pH and lactate sensor (PALS) that can provide continuous, physiologically relevant measurements. To measure pH, a sheet containing a pH-sensitive fluorescent dye is placed over 400 and 465 nm light-emitting diodes (LEDs) and a filter-coated photodetector. The filter-coated photodetector collects an emitted signal from the dye for each LED excitation, and the ratio of the emitted signals is used to monitor pH. To measure lactate, two sensing sheets comprising an oxygen-sensitive phosphorescent dye are each mounted to a 625 nm LED. One sheet additionally comprises the enzyme lactate oxidase. The LEDs are sequentially modulated to excite the sensing sheets, and their phase shift at the LED drive frequency is used to monitor lactate. In vitro results indicate that PALS successfully records pH changes from 6.92 to 7.70, allowing for discrimination between acidosis and alkalosis, and can track lactate levels up to 9 mM. Both sensing strategies exhibit fast rise times (< 5 min) and stable measurements. Multianalyte in vitro models of physiological disorders show that the sensor measurements consistently quantify the expected pathophysiological trends without cross talk; in vivo rabbit testing further indicates usefulness in the clinical setting.

Entities:  

Keywords:  continuous monitoring; critical care monitoring; lactate sensor; multianalyte; pH sensor; photonic

Mesh:

Substances:

Year:  2022        PMID: 35175733      PMCID: PMC8886565          DOI: 10.1021/acssensors.1c01720

Source DB:  PubMed          Journal:  ACS Sens        ISSN: 2379-3694            Impact factor:   7.711


pH and lactate are biomarkers that can be monitored during sepsis and liver and lung disease to improve patient outcome.[1−4] Blood pH is naturally buffered to within a range of 7.35 to 7.45 and deviations from this range can be indicative of serious disease.[4,5] Examples of these deviations are in cases of metabolic acidosis and alkalosis where blood pH can change when the body has too high or low concentration of hydrogen ions (H+) and/or bicarbonate.[4,5] A form of metabolic acidosis is lactic acidosis, which can occur when lactate is generated in excess (hyperlactatemia, >2 mM) and the amount of circulating H+ exceeds the capacity of the blood’s buffering system.[6,7] Elevated lactate levels are often a sign of cellular hypoxia as lactate is excessively produced during anaerobic glycolysis.[8] An increase in lactate can result in a decrease in pH; however, this is not always the case. For this reason, it is important to recognize that measuring only lactate or pH is not a surrogate for measuring the other. For example, in respiratory acidosis, pH may be directly influenced by an accumulation of carbon dioxide in the blood, while lactate remains unchanged.[9] In this case, frequent pH measurements can guide therapy. Importantly, there are a variety of circumstances when both pH and lactate should be measured simultaneously. Frequently referenced examples are during sepsis or septic shock as Lee et al. found that both the lactate levels and pH are essential when predicting patient mortality.[1] Current clinical standards for measuring blood pH and lactate require intermittent blood draws and analysis using benchtop instruments (Yellow Springs Instrument, YSI) or hand-held monitors (Abbott Laboratories i-STAT).[10−12] The frequency of measurements is limited ultimately by the frequency of blood draws which increase the likelihood of blood infections and anemia.[10,13] Meanwhile, there is substantial evidence suggesting that attentive monitoring and treatment of pH and lactate can improve patient prognosis and outcome.[14,15] In a multisite study with 348 patients having initial lactate ≥3.0 mEq/L, goal-oriented treatment targeting a 10% decrease in lactate per hour increased patient survivability by 10% as compared to the control (standard treatment, p = 0.067).[16] In addition to reduced mortality, patients were discharged from the intensive care unit (ICU) earlier than those in the control group. Moreover, in a prospective study with 75 patients experiencing sepsis and metabolic acidosis, both pH and lactate endpoint values were concluded as vital biomarkers toward patient outcome.[17] In this study, the 11 nonsurvivors had a lower mean pH and a higher mean lactate level than survivors after 5 days in the ICU. Changes in pH and lactate were not statistically significant (p > 0.714) as compared to initial values for the nonsurvivors, whereas significant changes were observed for the 64 survivors (P < 0.002).[17] Monitoring pH and lactate clearly has major implications toward patient prognosis, underscoring the value of a continuous pH and lactate sensor. There are well-established pH sensing modalities including electrochemical-based electrodes and luminescent dyes. Electrochemical-based glass and metal oxide electrodes measure pH through a difference in H+ concentration between a sensitive and reference electrode.[18,19] Although glass electrodes provide the highest level of specificity, they cannot be easily miniaturized and require frequent calibration.[19] Metal oxide electrodes can be easily miniaturized but exhibit low resolution, large drift, and hysteresis.[18,20] Meanwhile, pH-sensitive luminescent probes vary in their linear range of sensitivity (pKa), quantum yield, and mode of operation (intensity, emission ratio, etc.).[21−24] To continuously monitor pH on a transcutaneous sensor, we have selected luminescent dye 8-hydroxypyrene-1,3,6-trisulfonic acid trisodium salt (HPTS). HPTS has a reported pKa of 7.3 and exhibits minimal toxicity, a quantum yield of 0.82 in water, and a linear response ranging from 6.7 to 8.7, which spans the pathophysiological range of acidosis and alkalosis.[25,26] HPTS peak emission intensity at 520 nm is a function of its spectral absorption efficiency. HPTS absorbs light more efficiently at 450 nm at high pH as compared to low pH. Conversely, HTPS absorbs 405 nm light more efficiently at lower pH values.[27] It has been shown that with serial light excitation at 450 and 405 nm, HPTS can reliably measure pH through a ratiometric analysis at its peak emission wavelength.[27,28] In an approach similar to our own, Wencel et al. immobilized HPTS in a sol–gel placed at the tip of a multicore fiber and calibrated pH to the ratio of 520 nm emission with dual LED excitation.[29] Numerous strategies exist to measure lactate, including colorimetric assays, high-performance liquid chromatography, and enzyme-linked immunosorbent assays.[30−32] One common mode of enzyme-based lactate sensing (as used in the YSI 2300 STAT Plus Glucose and Lactate Analyzer) employs lactate oxidase (LOX).[33] The LOX reaction consumes both oxygen and lactate while producing hydrogen peroxide and pyruvate. This allows lactate to be indirectly measured through monitoring the consumption of oxygen or generation of hydrogen peroxide.[30,31] The generation of hydrogen peroxide is commonly used for electrochemical-based LOX sensing of lactate.[33−35] Common challenges in such electrochemical sensing include high operation voltages, dependencies on electron-transfer mediators (which can be toxic and exhibit poor solubility), and acetaminophen interference.[34,36,37] As an alternative to electrochemical sensing, our group and others have published the use of a phosphorescent oxygen-sensitive dye for continuous measurements of lactate and other analytes.[38,39] In our clinical studies, oxygen was detected by the metalloporphyrin dye platinum(II) meso-tetraphenyl tetrabenzoporphine (PtTPTBP).[39] PtTPTBP phosphorescence is quenched by oxygen, reducing its luminescence lifetime. This lifetime can be assessed by determining the phase shift between excitation and emission waveforms.[40] An indirect measurement of lactate may therefore be reported by calculating the phase shift of PtTPTBP emission in the presence of LOX, lactate, and oxygen. This sensing scheme is adapted here to continuously monitor lactate within a multianalyte sensor. Herein, we introduce a transcutaneous, continuous multianalyte sensor referred to as the PALS (Figure A). PALS employs two unique sensing modalities at the tip of a transcutaneous multianalyte sensing flexible (MSF) Sensor to monitor pH and lactate: (1) a dual excitation, single-band detection scheme that collects pH-sensitive light emissions and (2) a luminescence lifetime detection scheme that captures oxygen and lactate-sensitive light emissions (Figure B). PALS in vitro results show rapid rise time for both pH and lactate sensing, measurement reversibility, and sensitivity across their respective pathophysiological ranges. Furthermore, when measuring both analytes simultaneously, measurements were shown to be free of cross talk. An implant study in a rabbit model of hypoxemia provides further evidence that PALS signals are reversible and in agreement with reference values from a handheld blood gas analyzer. The ability of PALS to optically measure pH and lactate using a transcutaneous sensor contributes to the novelty of our technology and provides an advantage over electrochemical sensors which are prone to chemical species interference.
Figure 1

Schematic overview of PALS. (A) PALS comprises a wearable unit and a transcutaneous multianalyte sensing flexible (MSF) Sensor for pH and lactate monitoring. (B) Exploded view of the MSF tip. LEDs and the detector (DET) are soldered onto conductive pads of the flexible circuit. Sensing sheets are adhered to their corresponding optoelectronic components. The green bar above DET indicates the color filter. (C) (left) Oxygen is monitored by analyzing the PtTPTBP light emission phase shift using a photodetector (PD) housed in the wearable unit. (middle) Lactate oxidase consumes both oxygen and lactate; oxygen consumption can be used as an indirect measurement of lactate. (right) pH is measured by ratiometric analysis of HPTS light emission.

Schematic overview of PALS. (A) PALS comprises a wearable unit and a transcutaneous multianalyte sensing flexible (MSF) Sensor for pH and lactate monitoring. (B) Exploded view of the MSF tip. LEDs and the detector (DET) are soldered onto conductive pads of the flexible circuit. Sensing sheets are adhered to their corresponding optoelectronic components. The green bar above DET indicates the color filter. (C) (left) Oxygen is monitored by analyzing the PtTPTBP light emission phase shift using a photodetector (PD) housed in the wearable unit. (middle) Lactate oxidase consumes both oxygen and lactate; oxygen consumption can be used as an indirect measurement of lactate. (right) pH is measured by ratiometric analysis of HPTS light emission.

Methods

Filter-Coated Photodetector

For the pH sensor, silicon photodiodes (PDs) were coated with an optical filter as follows. A plastic green bandpass filter (Primary Green Filter, Lee filters, USA) was first cut into a 2 cm × 2 cm square. The square filter was then placed onto the surface of a 2 mm × 1.25 mm silicon PD (SFH2716, OSRAM Opto Semiconductors, Germany) that had been mounted on a microscope glass slide. The edges of the filter were held in place by two additional glass slides. A Varitemp VT-750C heat gun (Master Appliance, USA) at a temperature setting of 250°C was then used to melt the plastic green bandpass filter onto the PD (Figure S1).

pH Sensor Sheet

First, a 12.7 mM HPTS stock solution was formulated by dissolving 99 mg of HPTS (MilliporeSigma, USA) in 15 mL of Milli-Q water (MilliporeSigma, 18.2 MΩ·cm at 25 °C). A total of 10 g of 45–150 μm diameter Dowex 1X8 resin beads (MilliporeSigma, USA) was then suspended in the stock solution within a 20 mL disposable scintillation vial. This process yields the resin bead suspension (Figure A) and allows the negatively charged sulfonate groups on HPTS to ionically bind to the positively charged Dowex Resin.[25] A total of 500 μL of the resin bead suspension was then added to a 1.5 mL amber glass vial along with 50 mg of poly(ethylene glycol) dimethacrylate 8000 (PEGDMA8000, Polysciences, USA) and 12 mg of 2-hydroxy-1-(4-(2-hydroxyethoxy)phenyl)-2-methylpropan-1-one (Irgacure 2959, Sigma-Aldrich, USA), creating the pH sensor suspension (Figure B). PEGDMA is considered a biocompatible material with applications including drug delivery and cell encapsulation.[41−43]
Figure 2

HSS Fabrication. (A) Fluorescence confocal micrograph (Olympus Fluoview 1200) of the pH-sensitive resin bead suspension (405 nm laser excitation, emission bandpass filter: 505–540 nm). Scale bar = 100 μm. (B) Suspension is pipetted onto a PTFE sheet. (C) Suspension is sandwiched between two glass slides. (D) Suspension is polymerized, yielding the HSS. The HSS is hydrated to allow for hydrogel swelling.

HSS Fabrication. (A) Fluorescence confocal micrograph (Olympus Fluoview 1200) of the pH-sensitive resin bead suspension (405 nm laser excitation, emission bandpass filter: 505–540 nm). Scale bar = 100 μm. (B) Suspension is pipetted onto a PTFE sheet. (C) Suspension is sandwiched between two glass slides. (D) Suspension is polymerized, yielding the HSS. The HSS is hydrated to allow for hydrogel swelling. To fabricate the pH sensor sheet (HSS), a thin circular sheet of hydrophilic polytetrafluoroethylene (PTFE, H050A047A, 35 μm thick, 0.50 μm pores, 47 mm diameter, Sterlitech, USA) was cut into a 1.2 cm × 1 cm rectangle and placed onto a microscope glass slide. PTFE was chosen as the sensing film substrate because PTFE is biocompatible and is commonly used in dwelling commercial insulin infusion cannulas.[44−48] A total of 40 μL of the pH sensor suspension was then pipetted onto the cut PTFE sheet. The sheet was then sandwiched between two glass slides (Figure C) until the pH sensor suspension uniformly coated one side of the sheet. The pores of the PTFE sheet are 2 orders of magnitude smaller than the resin bead diameter, preventing resin penetration into the sheet. The coated sheet was then polymerized for 15 min using 365 nm wavelength light emitted from an 8 W dual-ultraviolet (UV) transilluminator (VWR, USA) to produce the HSS (Figure D). Upon UV polymerization, the resin bead suspension is encapsulated within a PEGDMA network that also polymerizes throughout the porous network of the PTFE. Thus, all materials in contact with the tissue are known to be biocompatible, nontoxic, and nonfouling.[49] To test for HPTS leaching out of the HSS, HSSs were incubated in 200 μL of PBS for 8 days. Absorption of 405 nm light in the incubation solution was measured in triplicate using a spectrophotometer (NanoDrop OneC, Thermo Scientific). Standards were formulated at concentrations 12.7 × [100 101 10–2 10–3 10–4 10–5] mM HPTS. All but the 12.7 × 10–5 mM HPTS solution had detectable absorption. Neither the HSS sample nor a negative control HSS lacking HPTS had detectable absorption, indicating negligible leaching over the 8 days. The HSS was then retrieved with tweezers and left to swell for at least 2 h in Milli-Q water within a 20 mL scintillation vial.

LOX and Oxygen Sensor Sheets

LOX sheets were fabricated from PTFE sheets coated with the oxygen-sensitive dye, the protein mixture containing enzymes LOX and catalase, and pretreatment solution, all of which are detailed in the Supporting Information. First, a 1 cm × 1 cm square of the dye-coated PTFE sheet was excised using a razor blade. A total of 4.5 μL of pretreatment solution was then pipetted onto each of the two microscope glass slides (Figure A). Next, the dye-coated PTFE sheet was sandwiched between the two glass slides to force the pretreatment solution into the pores of the sheet. The purpose of the pretreatment solution is to establish a porous network within the dye-coated PTFE sheet that limits the diffusion of lactate and oxygen. Next, two parallel double-layered strips of 25 μm-thick Kapton tape (Tapes Master, USA) were applied to a glass slide to function as spacers. A total of 4.5 μL of protein mixture was then pipetted in between the spacers and onto a second slide. The pretreated dye-coated PTFE sheet was then placed between the spacers and sandwiched between the glass slides. While being sandwiched, the sensor sheet was polymerized with the 8 W dual-UV transilluminator for 5 min to yield the LOX sheet (Figure B). The oxygen sheet was fabricated in a similar fashion except for the exclusion of LOX and catalase in its protein mixture.
Figure 3

LOX sheet fabrication. (A) Pretreatment solution wets the dye-coated PTFE sheet. (B) Pretreated dye-coated PTFE sheet is sandwiched between glass slides containing the protein mixture. Strips of Kapton tape serve as spacers to control the membrane thickness. UV polymerization cures the LOX sheet.

LOX sheet fabrication. (A) Pretreatment solution wets the dye-coated PTFE sheet. (B) Pretreated dye-coated PTFE sheet is sandwiched between glass slides containing the protein mixture. Strips of Kapton tape serve as spacers to control the membrane thickness. UV polymerization cures the LOX sheet.

PALS Circuitry

PALS consists of custom-made printed circuit boards (fabricated by OSH Park, USA) designed using Eagle (Autodesk, USA). Essential components of the circuits are shown in Figure S2. These custom boards were manually assembled and make up the three main subunits of PALS. The first subunit is the MSF (Figure A). The second subunit is the wearable unit comprising three circuit boards: (1) the integrator board for pH sensing that utilizes an IVC102 transimpedance amplifier (Texas Instruments, USA, Figure C), (2) the photodetector board for lactate and oxygen sensing (Figure D), and (3) the connector board to connect the photodetector and integrator boards to the MSF (Figure E). The last subunit is the backend controller unit (shown in Figure S2) comprising a microcontroller Teensy 3.2 (PJRC, USA) and two circuit boards: (1) the controller board for charlieplexing and tuning the drive current of the MSF LEDs (Figure S2A) and (2) the IVC power supply board. Custom software was written for Teensy 3.2 to charlieplex the LEDs and acquire data, using a combination of Arduino (Arduino, USA) with Teensyduino library (PJRC, USA) and LabView (National Instruments, USA).
Figure 4

PALS sensing components. (A) MSF containing five on-strip optoelectronic components required for sensing. (B) Fully assembled MSF comprising sensing sheets and waterproof coating. (C) Integrator board for amplifying pH signals. (D) Photodetector board for lactate and oxygen sensing. (E) PALS wearable unit connected to the MSF.

PALS sensing components. (A) MSF containing five on-strip optoelectronic components required for sensing. (B) Fully assembled MSF comprising sensing sheets and waterproof coating. (C) Integrator board for amplifying pH signals. (D) Photodetector board for lactate and oxygen sensing. (E) PALS wearable unit connected to the MSF. Custom-designed housings were printed using a stereolithography printer (Prusa SL1, Prusa Research, Czech Republic) to protect and house the wearable unit, as illustrated in Figure . The housing unit serves to also pressure-connect the prongs of the 5-spring battery connector (009155005852006, AVX Corporation, USA) on the connector board to the gold pads of the MSF, as depicted in Figure E.

MSF Fabrication

The MSF was printed and designed with OSH Park and Eagle. The MSF comprises polyimide, which exhibits insignificant cytotoxicity and protein adsorption when compared to PTFE and polydimethylsiloxane.[50,51] The MSF contains optoelectronic components for both pH and lactate sensing (Figure A). The optoelectronic components for pH sensing are a 400 nm LED (SM0603UV-400, Bivar, USA), a 465 nm LED (APT1608QBC/G, Kingbright, USA), and a filter-coated photodetector. Two 625 nm LEDs (APHHS1005LSECK/J3-PF, Kingbright, USA) are used for lactate and oxygen sensing. The flex circuit is 0.135 mm thick, the detector is 0.8 mm tall, and the pH sensing sheet is 0.035 μm thick. Collectively, the thickest portion of MSF is 1.15 mm. Each LED spectrum is shown in Figure S3. Additionally, Video S1 shows sequential activation of the MSF LEDs. The MSF was coated with Loctite EA E-60NC (1:1 resin to hardener mix ratio, Henkel, Germany) for waterproofing. The waterproof coating was left to cure overnight. The sensing sheets were applied as described in the Supporting Information (Figure B).

Measurement Parameters

For pH sensing, LED currents were set to 6 mA and HSS-emitted light was sampled at 10Khz. Detection integration times (<1 s) were tuned at the start of each experiment. The tuning occurs in the first test solution of each in vitro experiment or following insertion for the in vivo experiment. Specifically, integration time was tuned such that the IVC hold voltage ranged between 1 and 2 V, corresponding to 30–61% of the total integration capacity. For lactate sensing, each of the 625 nm LEDs was illuminated one at a time. The LEDs were driven by a square waveform comprising 21 cycles having a peak current of 9 mA at a frequency of 5 kHz and 25% duty cycle. PtTPTBP emission was sampled simultaneously at 500 kHz. A unique MSF was used for each of the in vitro and in vivo experiments.

Statistical Analysis

Measurements are reported as the mean and standard error of the mean. Statistical analyses were conducted using a one-way analysis of variance (ANOVA) with Tukey post hoc comparison or coefficient of variance (CV). p < 0.05 denotes statistical significance. Prism 8 (GraphPad, USA) was used for statistical analysis.

In Vivo Rabbit Study

The PALS implant study was conducted with the approval of the Institutional Animal Care and Use Committee at the University of California Irvine. A New Zealand white rabbit was sedated with a 2:1 ratio of ketamine hydrochloride (100 mg/mL, Ketaject, Phoenix Pharmaceutical Inc., USA): xylazine (20 mg/mL, Anased, Lloyd Laboratories) at a dose of 37.5 mg/kg of ketamine and 5 mg/kg of xylazine IM using a 25 gauge 5/8 inch needle. The mixture of ketamine and xylazine was infused into the animal’s right marginal ear vein. The animal was intubated and placed on mechanical ventilation with a tidal volume of 50 mL per breath, respiratory rate of 20 breaths/min, and 100% oxygen. An arterial catheter was placed within the right femoral artery for systemic blood pressure measurements and arterial blood gas sampling. The MSF was implanted into the subcutaneous space of the inner left thigh of the rabbit. First, an incision (length = 0.8 cm) was created with a scalpel. Next, Metzenbaum dissecting scissors (Cole-Parmer, USA) were inserted inside the incision to create a pocket by separating the skin from underlying muscle. This pocket runs parallel to the thin skin of the rabbit and accommodates approximately 2 cm of the MSF tip. After MSF-tip insertion, Loctite 4981 adhesive (Henkel, Germany) was applied at the incision site to adhere the MSF onto the skin and seal the incision. The PALS wearable unit was then placed on the skin and aligned to the MSF tip. Hypafix adhesive (USA) was placed over the PALS wearable unit to limit sensor movement. After baseline measurements, the PALS was disconnected, and the animal was placed inside a sealed chamber which was then moved into a fume hood. The animal received 800 ppm chlorine gas (Airgas, USA) for 6 min followed by a 5 min rest period. A total of 1 mL of 100 mM trihistidyl cobinamide was then administered through the right marginal ear vein. The animal was taken back to the surgical room, and the PALS was reconnected to resume monitoring. At the conclusion of the study, the animal was euthanized as per standard procedures (1 mL of Euthasol, Virbac, USA).

Analysis of In Vivo Data

During the study, blood was drawn from the right femoral artery at eight time points. Four blood draws were retrieved before and after chlorine gas and cobinamide infusion. Blood pH and lactate concentration was immediately assessed with an i-STAT (Abbott Laboratories, USA) using Abbott CG4+cartridges (Abbott Laboratories, USA). PALS pH and lactate measurements were retrospectively calibrated to the blood pH and lactate concentrations obtained from the i-STAT with a linear regression model.

Results and Discussion

pH Sensing by Dual LED Excitation and Single-Band Detection

The PALS uses a dual LED excitation, single-band photodetection scheme for measuring pH. To enable such sensing, the tip of the MSF comprises two surface-mount LEDs to excite an HSS and a coated photodetector to collect pH-sensitive emissions. To first determine if this is a viable scheme, a benchtop optical system was constructed (Figure A; fabrication methods in the Supporting Information). This system includes surface-mount 400 nm and 465 nm dominant-wavelength LEDs soldered onto a protoboard and a microscope objective lens that couples emitted light to a spectrometer (Figure A). pH-sensitive spectra from an HSS in a pH 7.6 solution were collected with serial illumination from the two LEDs (Figure B). The HSS emission spectrum ranges from 470 to 660 nm and has a peak value at 520 nm. While the 400 nm LED light does not overlap the HSS emission spectrum, the 465 nm LED does, necessitating an optical filter that mitigates the effects of this spectral cross talk. Ideally, such a filter should transmit the HSS emission spectrum (Figure B) and be easily applied onto the surface of a PD. The primary green filter was selected which can be applied onto a PD as described in the Methods. To simulate the filtering behavior of the primary green filter, its transmission spectrum was multiplied with HSS emission spectra (one per LED). The resulting filtered spectra show that the 465 nm LED light has a negligible contribution to the fluorescence signal (Figure C). To test for pH sensitivity and reversibility, HSS spectra in solutions having pH 7.2 or 7.6 were collected and then multiplied by the primary green filter transmission curve. The ratio of area under the curve (AUC Ratio: Ex465/Ex400) was determined at these two pH values. Results indicate that the dual LED, single-band detection scheme is stable and reversible and can be utilized as a pH optode at the tip of the MSF (Figure D).
Figure 5

Dual LED excitation of the HSS. (A) (left) 465 nm and 400 nm LEDs soldered onto a protoboard and covered by an HSS; (right) HSS emission signals are collected using a benchtop optical system for analysis. An optical fiber routes the collected light to a spectrometer. (B) Dual emission spectra at pH 7.6. The green spectrum shows the transmission curve of the primary green filter. The black dotted line indicates the HPTS peak emission wavelength. (C) HPTS emission spectra following multiplication with the primary green filter transmission spectrum. (D) AUC ratios for solutions of pH 7.2 and 7.6. Mean ratios were analyzed for statistical significance. *p < 0.05.

Dual LED excitation of the HSS. (A) (left) 465 nm and 400 nm LEDs soldered onto a protoboard and covered by an HSS; (right) HSS emission signals are collected using a benchtop optical system for analysis. An optical fiber routes the collected light to a spectrometer. (B) Dual emission spectra at pH 7.6. The green spectrum shows the transmission curve of the primary green filter. The black dotted line indicates the HPTS peak emission wavelength. (C) HPTS emission spectra following multiplication with the primary green filter transmission spectrum. (D) AUC ratios for solutions of pH 7.2 and 7.6. Mean ratios were analyzed for statistical significance. *p < 0.05.

pH Optode

The pH optode comprises three of the optoelectronic elements at the tip of the MSF (Figure A). The optode is fabricated by placing an HSS at the tip of the MSF such that it overlaps the two LEDs and filter-coated photodetector (Figure B) as described in the Supporting Information. Photocurrents from the photodetector are amplified and converted to a voltage using a transimpedance amplifier. pH is related to the ratio of the voltages acquired with 465 and 400 nm LED illumination (REx465/Ex400) after background (in the absense of LED emission) subtraction.

Lactate Optode

The lactate optode was adapted from our previous work. In brief, two oxygen-sensitive PtTPTBP dye sheets are mounted on two 625 nm dominant-wavelength LEDs. One of these sheets also contains the enzymes LOX and catalase. The working optode has the enzyme-containing sheet, whereas the reference optode has the enzyme-free sheet. The LEDs are illuminated in sequence, and oxygen-sensitive dye emission is detected using a photodetector within the PALS wearable unit (Figures and 4E).[52] Lactate is related to the phase shift difference between the reference and working optodes. A phase shift (ϕ) is defined as the phase difference (assessed by Fourier transform) between the LED drive-current waveform and the corresponding PD signal at the drive frequency. The Stern–Vollmer equation relates oxygen partial pressure to phosphorescence lifetime under measurement conditions (τ) and in the absence of oxygen (τ0).[53] ϕ is related to τ by the function tan (ϕ) = 2πfτ, where f is the modulation frequency of LED emission.[54] This gives the equationswhere ϕw and ϕR correspond to the working and reference optode, respectively, ϕ0 corresponds to the absence of oxygen, Ksv is the Stern Vollmer constant, and pO2,w and pO2,R are the oxygen partial pressures at the working and reference optode, respectively. Subtracting eq from eq yieldswhich can be rewritten to define the sensor signal β Here, is a proportionality constant determined through linear calibration in a series of lactate solutions at room air.

pH Sensing In Vitro

To determine baseline pH sensor stability, measurements were obtained every 30 min for 8 h in pH 7.45 solution. Unless otherwise stated, each reported steady-state measurement is an average of 20 single-point measurements acquired in succession. No signal drift was detected (Figure S4; CV = 0.002). The stability is due in part to the quality of our custom LED current control circuitry (provided by the TLC driver, Figure S2A) that produces consistent LED output power. To check the stability of both LEDs, their emission spectra were acquired every 3 min for 30 min. Spectra show no significant differences in AUC (CV = 0.001) or peak emission intensity (CV ≤ 0.002; Figure S5). pH sensing rise time was assessed by first incubating a sensor in a solution with pH 7.45 and then exchanging for a solution with pH 7.01. REx400/Ex465 (the ratio of detector signals under 400nm and 465 nm excitation) was measured every 4.8 s following media exchange. Response kinetics are described using a rising exponential plateau model (R2 = 0.99, standard error of the estimate (Sy.x) = 0.006) with a t90 rise time of 2.01 min (Figure A). This rise time is short enough to capture critical physiological events such as subdermal scalp acidosis following fetal tachycardia in high-risk births, which occur on the time scale of minutes.[55]
Figure 6

Ratiometric pH Sensing. (A) pH sensor rise time from pH 7.45 to 7.01. The inset shows the hold voltages for each LED. (B) pH sensor steady-state measurements in a series of pH solutions. Arrows indicate the sequence of test solutions. (C) Calibration curve. The red shaded region indicates a normal physiological range of pH in blood, 7.35–7.45, while the orange and gray shaded regions indicate acidosis and alkalosis, respectively.

Ratiometric pH Sensing. (A) pH sensor rise time from pH 7.45 to 7.01. The inset shows the hold voltages for each LED. (B) pH sensor steady-state measurements in a series of pH solutions. Arrows indicate the sequence of test solutions. (C) Calibration curve. The red shaded region indicates a normal physiological range of pH in blood, 7.35–7.45, while the orange and gray shaded regions indicate acidosis and alkalosis, respectively. pH sensing reversibility was assessed by sequential measurements of solutions having a pH of 7.05, 7.22, 7.40, 7.22, and 7.05, in that order (Figure B). To obtain a steady measurement at each pH, the sensor was first washed ten times with that pH solution. One-way ANOVA indicates a significant effect across groups (p ≪ 0.01). Tukey post hoc comparison with adjusted p-values shows significant differences between the three unique pH solutions (p ≪ 0.01 for each comparison) but no significant differences between repeated pH solutions (p > 0.25 for each comparison). pH sensing sensitivity and range were assessed by exposure to 13 solutions, with pH ranging from 6.92 to 7.59. Data follow a Boltzmann sigmoidal model (R2 > 0.99 and Sy.x = 0.002) with a pKa of 7.16 (Figure C), which is similar to the pKa of the HPTS dye alone.[25] Importantly, wellness-of-fit indicates that PALS will have clinical relevance across the pathophysiological range including acidosis and alkalosis.[56]

Lactate Sensing In Vitro

To determine lactate sensor stability, measurements were obtained every 30 min for 4 h in 4 mM lactate solution. Unless otherwise stated, each reported steady-state value is the mean of 10 measurements. No significant signal drift was detected (Figure S6; CV = 0.007). The stability is due in part to the inclusion of the enzyme catalase that scavenges hydrogen peroxide, known to degrade proteins.[57,58] In this reaction, iron in catalase extracts one oxygen molecule from hydrogen peroxide, producing a single water molecule. Then, another oxygen molecule from a second hydrogen peroxide molecule binds to the oxygen-bound catalase, producing a second water molecule. Molecular oxygen is then released from the catalase.[59] Lactate sensing rise time was then assessed by changing the test media from 0 mM (1× phosphate-buffered saline solution, PBS) to 6 mM lactate. Phase shifts were recorded and β was calculated every 12 s following media exchange. Response kinetics follow an exponential plateau model (R2 = 0.95 and Sy.x = 0.01) with a t90 rise time of 3.65 min (Figure A). This rise time is comparable to commercial continuous glucose monitors such as the Dexcom G6 and Medtronic Guardian, which are reported to have an average rise time in vivo of 9.5 min and are effective in guiding insulin therapy.[60]
Figure 7

In vitro lactate sensing. (A) Rise time measured after media exchange from 0 to 6 mM lactate. The inset represents the raw lactate and oxygen phase shift measurements. (B) β for a series of test solutions. (C) Lactate sensor calibration and (D) β acquired during the calibration experiment. Green lines in A and C represent model fit to the data (equation is listed next to each curve).

In vitro lactate sensing. (A) Rise time measured after media exchange from 0 to 6 mM lactate. The inset represents the raw lactate and oxygen phase shift measurements. (B) β for a series of test solutions. (C) Lactate sensor calibration and (D) β acquired during the calibration experiment. Green lines in A and C represent model fit to the data (equation is listed next to each curve). Lactate sensing reversibility was evaluated by sequential testing of lactate solutions having concentrations of 0, 2.60, 4.64, 6.53, 4.64, 2.60, and 0 mM (Figure B). To obtain a steady measurement at each lactate concentration, the sensor was first washed ten times with that lactate solution, after which measurements were obtained. One-way ANOVA detected differences between groups (p ≪ 0.01). Tukey post hoc comparison with adjusted p-values shows no significant differences between the pairing of identical lactate solutions (p > 0.87), while significant differences were found between solutions of different lactate concentrations (p ≪ 0.01). Lactate sensitivity and range were evaluated by incubation in seven lactate solutions ranging in concentration from 0 to 14 mM. Data follow an exponential plateau model (R2 = 0.99 and Sy.x = 0.03), showing sensitivity to lactate from 0 to 9 mM (Figure C). Discrete β measurements are shown in Figure D. These results show that PALS lactate sensing can distinguish between concentrations within the pathophysiological range.

Multianalyte Sensing In Vitro

Combined pH and lactate sensing was tested using in vitro models of five pathophysiological conditions (Figure ). The conditions were designed to capture sensitivity and trends of both pH and lactate in such pathologies and to detect any interference or cross talk. A series of test solutions with prescribed pH and lactate concentration were formulated and analyzed using calibrated instruments (a Mettler Toledo FiveEasy pH probe and a YSI 2300 STAT Plus Glucose and Lactate Analyzer). Solutions were then probed using PALS in an order corresponding to each model. In a model of hyperlactatemia, pH was maintained at 7.43 ± 0.02 while lactate was increased (Figure A). PALS signals report the increase in lactate, while the pH signals did not significantly change (CV = 0.018). In a model of lactic acidosis, as experienced in sepsis, solutions were formulated to have a decrease in pH and an increase in lactate (Figure B), and PALS successfully reports these changes. In a model of panic-disorder patients who exhibit both respiratory alkalosis and hyperlactatemia, solutions were formulated to have an increase in both pH and lactate concentrations.[3] PALS signals successfully reflect these increases (Figure C). In a model of extreme metabolic alkalosis, as can occur with chronic vomiting and diarrhea, pH was increased while lactate was held at 3.50 mM ± 0.02 (Figure D).[61] PALS signals report the increase in pH and constant lactate concentration (CV = 0.037). In a model of respiratory acidosis (as seen in chronic obstructive pulmonary disease), pH was formulated to decrease while lactate levels were held constant at 3.52 ± 0.02 (Figure E).[62] PALS signals report the decrease in pH and constant lactate concentrations (CV = 0.016).
Figure 8

pH and lactate multianalyte sensing. A series of solutions were formulated to model pathological conditions. Solutions were formulated then measured with a Mettler Toledo FiveEasy pH probe and YSI 2300 STAT Plus Glucose and Lactate Analyzer; values are shown on the horizontal axes. Solutions were tested by PALS in order from left to right. Solutions model: (A) hyperlactatemia, (B) lactic acidosis, (C) respiratory alkalosis and hyperlactatemia, (D) extreme metabolic alkalosis, and (E) respiratory acidosis.

pH and lactate multianalyte sensing. A series of solutions were formulated to model pathological conditions. Solutions were formulated then measured with a Mettler Toledo FiveEasy pH probe and YSI 2300 STAT Plus Glucose and Lactate Analyzer; values are shown on the horizontal axes. Solutions were tested by PALS in order from left to right. Solutions model: (A) hyperlactatemia, (B) lactic acidosis, (C) respiratory alkalosis and hyperlactatemia, (D) extreme metabolic alkalosis, and (E) respiratory acidosis. To test for and quantify signal interference, correlation between lactate concentration and pH (as measured by analytical instruments) in test solutions was first determined by aggregating values across all experiments (Figure A–E). The Pearson’s product-moment coefficient was low (−0.01) with a nonsignificant p-value (0.95), confirming independence within test solutions. Next, to test for correlations between the two sensing modalities, we aggregated raw signals across all experiments. The Pearson’s product-moment coefficient was low (−0.1) with a nonsignificant p-value (0.68). This insignificant correlation demonstrates the lack of interference between PALS sensing modalities. Collectively, Figure demonstrates that PALS does not exhibit sensing modality cross talk and can report physiologically relevant pH and lactate changes, which may aid clinicians in their practice of analyte-guided treatment toward improving patient outcome.

Multianalyte Sensing In Vivo

PALS multianalyte sensing was tested in an in vivo rabbit model of chlorine gas poisoning and cobinamide treatment. When chlorine gas reacts with water in the lungs, hydrochloric and hypochlorous acid are produced. The production of these acids damages the respiratory mucus membrane resulting in pulmonary edema and hypoxemia.[61,63] PALS should therefore detect decreasing pH and increasing lactate values following chlorine gas administration. At the start of the experiment, the MSF tip was implanted and connected to the wearable unit (Figure A) for data collection (LED serial illumination shown in Video S2). PALS signals were converted to analyte values using retrospective linear regression (Figure S7). Both blood pH and lactate concentration decreased prior to drug infusion as measured by PALS and intermittent i-STAT blood assays (Figure B,C). These decreases may be a function of the sedative chemistry and the ventilation control. After poisoning and cobinamide treatment, PALS detected the expected increase in lactate concentration and decrease in blood pH, which were also in agreement with the intermittent i-STAT blood assays (Figure B,C). The area under the concentration-time curve ratio between PALS and i-STAT measurements was calculated and assessed as described in Dror et al., following FDA guidelines.[39,64] The area under the concentration-time curve ratio for pH (0.979) and lactate (1.084) indicates PALS bioequivalence to the i-STAT blood analyzer in this study. There is a difference in pH measurements between i-STAT and PALS beginning at 110 min. PALS reports a higher pH value, with a difference of approximately 0.025. In previous work, Bland-Altman analysis comparing i-STAT to i-Smart and pHOx ultra-pH meters shows that i-Stat underestimates pH (relatively) by as much as 0.5 for values of pH < 7.3, indicating that the i-STAT may be a source of error for the last two time points of blood comparisons.[65] Results demonstrate that PALS reports clinically relevant pH and lactate changes and at a higher frequency than can be offered by blood analysis alone.
Figure 9

In vivo multianalyte sensing in a rabbit exposed to chlorine gas. The purple region indicates when PALS measurements were paused for chlorine gas and cobinamide administration. (A) Sensor implanted in the subcutaneous space of the inner left thigh. Lactate (B) and pH (C) sensing signals with intermittent blood reference values (i-STAT).

In vivo multianalyte sensing in a rabbit exposed to chlorine gas. The purple region indicates when PALS measurements were paused for chlorine gas and cobinamide administration. (A) Sensor implanted in the subcutaneous space of the inner left thigh. Lactate (B) and pH (C) sensing signals with intermittent blood reference values (i-STAT).

Conclusions

We have shown that two different photonic detection schemes on an implanted flexible sensor can yield reliable signals for monitoring both pH and lactate across clinically relevant ranges. Our results suggest that PALS can be used as a one-day device, which in many cases is sufficient in the ICU considering the ease of replacement and the average ICU stay. As pH and lactate values are strong predictors of mortality, the ability to provide stable and reliable measurements of pH and lactate, on the time scale of minutes to hours, can be utilized to guide therapy and improve patient outcome.[15,66] However, the technology should be improved before it can be used in the clinic over extended periods. Because the process of PALS fabrication is not controlled at the level expected for a commercial product, we do observe differences in calibration parameters between units. Currently, methods are being developed by our group to reduce variances in sheet fabrication and to improve the process by which the plastic optical filter is applied onto the photodetector. Additional engineering development will include miniaturization of the backend electronics to include a battery as well as wireless data control and acquisition, which are achievable by the existing manufacturing methods. In regard to miniaturizing the MSF, our research group has successfully reduced the lactate optode to a width of 300 μm.[39] For the pH optode, the dimensions of the photodetector are the limiting factor for miniaturization. In future work, we will incorporate photodetectors having dimensions similar to the MSF LEDs, yielding a 1/3 mm wide device. PALS can have a major impact on patient outcomes under conditions such as sepsis and organ failure where attentive monitoring of pH and lactate has been reported to improve patient outcome.[1−4] Importantly, the core elements of PALS can be replicated and modified to sense additional analytes on the MSF. For example, similar to the operating principle of HPTS, fluorescent dyes Fura Red AM and SBFI-AM exhibit a change in their absorbance behavior based on the concentration of calcium and sodium, respectively.[67,68] Consequently, our dual-excitation, single-band detection scheme could be readily employed to continuously monitor these analytes. Moreover, our lactate sensing scheme is generalizable to additional oxidases and their corresponding analytes including glucose and alcohol.[69,70] The addition of these analytes can broaden the applicability of the MSF to other medical conditions. For example, continuous monitoring of sodium, pH, lactate, oxygen, and glucose in individuals experiencing diabetic ketoacidosis can diagnose dehydration (loss of sodium), ketoacidosis (high glucose and low pH), and ischemia (low oxygen and high lactate), directing healthcare professionals toward a specific mode of intervention.[71−73] An expanded MSF could also be applicable beyond the medical field, such as in bioreactors for protein expression, agriculture, water management, and food industry, where sensing pH, glucose, and sodium would provide vital information to each respective field.[74−77]
  55 in total

1.  OXYGEN CONSUMPTION, OXYGEN DEBT AND LACTIC ACID IN CIRCULATORY FAILURE.

Authors:  J Meakins; C N Long
Journal:  J Clin Invest       Date:  1927-06       Impact factor: 14.808

Review 2.  Acute inhalation injury.

Authors:  Metin Gorguner; Metin Akgun
Journal:  Eurasian J Med       Date:  2010-04

3.  Imaging of mitochondrial pH using SNARF-1.

Authors:  Venkat K Ramshesh; John J Lemasters
Journal:  Methods Mol Biol       Date:  2012

Review 4.  Etiology and therapeutic approach to elevated lactate levels.

Authors:  Lars W Andersen; Julie Mackenhauer; Jonathan C Roberts; Katherine M Berg; Michael N Cocchi; Michael W Donnino
Journal:  Mayo Clin Proc       Date:  2013-10       Impact factor: 7.616

5.  Photophysics of the fluorescent pH indicator BCECF.

Authors:  Noël Boens; Wenwu Qin; Nikola Basarić; Angel Orte; Eva M Talavera; Jose M Alvarez-Pez
Journal:  J Phys Chem A       Date:  2006-08-03       Impact factor: 2.781

6.  Characterization of Lactate Sensors Based on Lactate Oxidase and Palladium Benzoporphyrin Immobilized in Hydrogels.

Authors:  Liam P Andrus; Rachel Unruh; Natalie A Wisniewski; Michael J McShane
Journal:  Biosensors (Basel)       Date:  2015-07-07

7.  Clinical evaluation of a novel subcutaneous lactate monitor.

Authors:  Nitsan Dror; John Weidling; Sean White; Francesca Ortenzio; Samir Shreim; Mark T Keating; Hoang Pham; Shlomit Radom-Aizik; Elliot Botvinick
Journal:  J Clin Monit Comput       Date:  2021-04-10       Impact factor: 1.977

Review 8.  Biomaterials in cardiovascular research: applications and clinical implications.

Authors:  Saravana Kumar Jaganathan; Eko Supriyanto; Selvakumar Murugesan; Arunpandian Balaji; Manjeesh Kumar Asokan
Journal:  Biomed Res Int       Date:  2014-05-08       Impact factor: 3.411

9.  Influence of pH control in the formation of inclusion bodies during production of recombinant sphingomyelinase-D in Escherichia coli.

Authors:  Andrea Castellanos-Mendoza; Ricardo M Castro-Acosta; Alejandro Olvera; Guadalupe Zavala; Miguel Mendoza-Vera; Enrique García-Hernández; Alejandro Alagón; Mauricio A Trujillo-Roldán; Norma A Valdez-Cruz
Journal:  Microb Cell Fact       Date:  2014-09-12       Impact factor: 5.328

10.  Printed, Flexible Lactate Sensors: Design Considerations Before Performing On-Body Measurements.

Authors:  Margaret E Payne; Alla Zamarayeva; Veronika I Pister; Natasha A D Yamamoto; Ana Claudia Arias
Journal:  Sci Rep       Date:  2019-09-23       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.