The use of the iSTAT portable analyzer in patients undergoing cardiopulmonary bypass.

N R Connelly1, M Magee, B Kiessling.   

Abstract

OBJECTIVE: To evaluate the utility of the iSTAT blood analyzer, a bedside device for hematocrit, sodium, potassium, and glucose measurement during cardiopulmonary bypass (CPB).
METHODS: Forty patients scheduled for elective CPB were evaluated prospectively. In addition to using the iSTAT analyzer, blood samples were analyzed at four time points: following induction of anesthetic, 10 min. after initiation of CPB, 60 min. after initiation of CPB, and following heparin neutralization by protamine. Blood glucose concentration was measured by the hospital laboratory using a Kodak Analyzer and by a glucose meter, electrolytes were evaluated by the Kodak Analyzer and BGE (a device which is commonly used for "satellite laboratory" determinations of electrolyte and blood gas results), and hematocrit samples were measured by the hospital laboratory using an NE 8,000 and a centrifuge. The means and standard deviations of the differences between the methods were calculated.
RESULTS: The hematocrit values determined by the iSTAT machine, when adjusted for the level of total protein (according to manufacturer's directions), differed from the laboratory values by 0.53 +/- 1.46 percentage points. An alternative to measuring total protein and making the adjustment is simply adding 1% to the hematocrit in the pre-CPB period and 3% on-CPB or post-CPB, which we found to yield values that differed from the laboratory by 0.52 +/- 1.42 percentage points. For all four tests (hematocrit, sodium, potassium, and glucose) the iSTAT had a similar relationship to the laboratory values as did the other commonly used means (centrifuge, BGE, and glucose meter) of clinical evaluation.
CONCLUSION: In summary, we found that in patients undergoing CPB, the iSTAT values agreed sufficiently well with standard laboratory values and that the iSTAT instrument can be relied upon for bedside measurements.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8863111     DOI: 10.1007/bf02221752

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  6 in total

1.  A clinical evaluation of the Hemocue haemoglobinometer using capillary, venous and arterial samples.

Authors:  P P Chen; T G Short; D H Leung; T E Oh
Journal:  Anaesth Intensive Care       Date:  1992-11       Impact factor: 1.669

2.  A comparison of conductivity-based hematocrit determinations with conventional laboratory methods in autologous blood transfusions.

Authors:  D J McMahon; R L Carpenter
Journal:  Anesth Analg       Date:  1990-11       Impact factor: 5.108

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  Evaluation of STAT-CRIT hematocrit determination in comparison to Coulter and centrifuge: the effects of isotonic hemodilution and albumin administration.

Authors:  S E McNulty; S J Sharkey; B Asam; J H Lee
Journal:  Anesth Analg       Date:  1993-04       Impact factor: 5.108

5.  Analytical evaluation of i-STAT Portable Clinical Analyzer and use by nonlaboratory health-care professionals.

Authors:  E Jacobs; E Vadasdi; L Sarkozi; N Colman
Journal:  Clin Chem       Date:  1993-06       Impact factor: 8.327

6.  Bedside hemoglobin measurements: sensitivity to changes in serum protein and electrolytes.

Authors:  S E McNulty; S J Sharkey; H Schieren
Journal:  J Clin Monit       Date:  1994-11
  6 in total
  5 in total

1.  Conductivity-based hematocrit measurement during cardiopulmonary bypass.

Authors:  Jacoline Steinfelder-Visscher; Patrick W Weerwind; Steven Teerenstra; Gheorghe A M Pop; René M H J Brouwer
Journal:  J Clin Monit Comput       Date:  2006-11-04       Impact factor: 2.502

2.  Evaluation of point-of-care analyzers' ability to reduce bias in conductivity-based hematocrit measurement during cardiopulmonary bypass.

Authors:  Steven Teerenstra; Jacoline Steinfelder-Visscher; Jacqueline Klein Gunnewiek; Patrick W Weerwind
Journal:  J Clin Monit Comput       Date:  2013-08-30       Impact factor: 2.502

3.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

4.  Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients.

Authors:  Jacoline Steinfelder-Visscher; Steven Teerenstra; Jacqueline M T Klein Gunnewiek; Patrick W Weerwind
Journal:  J Extra Corpor Technol       Date:  2008-03

5.  Improving access to diagnostics: an evaluation of a satellite laboratory service in the emergency department.

Authors:  P Leman; D Guthrie; R Simpson; F Little
Journal:  Emerg Med J       Date:  2004-07       Impact factor: 2.740

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.