The portable laboratory: an evaluation of the accuracy and reproducibility of i-STAT.

D Bingham1, J Kendall, M Clancy.   

Abstract

The i-STAT point-of-care testing analyser, which utilizes single-use multisensor cartridges, was evaluated. Two cartridges were assessed: the 6+ and the G3+, which provide results for urea, glucose, sodium, potassium, chloride, haematocrit, pH, PCO2, PO2 and various calculated parameters. The results for all analytes agreed well with the analysers in routine use in the laboratory. The reproducibility was comparable even when analysis was carried out by a nurse with only 5 min training. The system was found to be reliable, easy to use and required no maintenance (only a 2-min daily check of the electronics). These features, together with portability and the storage capacity for results, make the i-STAT suitable for point-of-care use, particularly in critical care units.

Entities:  

Mesh:

Year:  1999        PMID: 10370763     DOI: 10.1177/000456329903600109

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  7 in total

1.  Point of care testing: randomised controlled trial of clinical outcome.

Authors:  J Kendall; B Reeves; M Clancy
Journal:  BMJ       Date:  1998-04-04

2.  Comparison of Blood Glucose Monitoring Systems for Use in Insulin Clamp Studies During Either Intralipid or Glycerol Co-infusions.

Authors:  Jonathan Q Purnell; Julie Martin; Melanie Gillingham
Journal:  J Diabetes Sci Technol       Date:  2017-05-16

3.  Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients.

Authors:  Jacoline Steinfelder-Visscher; Steven Teerenstra; Jacqueline M T Klein Gunnewiek; Patrick W Weerwind
Journal:  J Extra Corpor Technol       Date:  2008-03

4.  Effect of nutritional status and sampling intensity on recovery after dorsal aorta cannulation in free-swimming Atlantic salmon (Salmo salar L.).

Authors:  B Djordjevic; T Kristensen; Ø Øverli; B O Rosseland; A Kiessling
Journal:  Fish Physiol Biochem       Date:  2009-10-25       Impact factor: 2.794

5.  Evaluation of the i-STAT (portable clinical analyser) for measuring haematological parameters in Atlantic cod (Gadus morhua) at different CO2 and temperature conditions.

Authors:  Radoslav S Borissov; Sølvi Espeland; Martin H Iversen
Journal:  Fish Physiol Biochem       Date:  2019-10-03       Impact factor: 2.794

Review 6.  Multianalyte Physiological Microanalytical Devices.

Authors:  Anna Nix Davis; Adam R Travis; Dusty R Miller; David E Cliffel
Journal:  Annu Rev Anal Chem (Palo Alto Calif)       Date:  2017-06-12       Impact factor: 12.400

7.  Optimal Analytical Performance for POCT.

Authors:  Callum G Fraser
Journal:  EJIFCC       Date:  2001-04-05
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.