Analytical evaluation of i-STAT Portable Clinical Analyzer and use by nonlaboratory health-care professionals.

E Jacobs1, E Vadasdi, L Sarkozi, N Colman.   

Abstract

We evaluated the performance of the i-STAT Portable Clinical Analyzer, a hand-held instrument that, with its current cartridge, analyzes for electrolytes, urea nitrogen, glucose, and hematocrit in approximately 60 microL of whole blood in approximately 90 s. Accuracy, imprecision, and linearity studies were performed with aqueous controls and standards and by split-sample analysis. Intrarun imprecision (CV) ranged from 0.34% to 3.97%. Total imprecision over a 2-month period ranged from 0.42% to 4.83%, with urea nitrogen and glucose analyses generating the higher values. Patients' results from the Portable Clinical Analyzer correlated well with those obtained for whole blood or plasma by the Nova Stat Profile 5, the Beckman Synchron CX3, or the Technicon H1 Hematology Analyzer, with Sylx values < 0.2 mmol/L for potassium; < 1.5 mmol/L for sodium, glucose, and urea nitrogen; < 2.4 mmol/L for chloride, and < 2.4% for hematocrit. We also ascertained imprecision and accuracy of the system placed in a cardiothoracic intensive-care unit and operated by nurses. There were no significant differences in either the imprecision or accuracy of the system in this setting. We conclude that operator technique is not a factor in the analytical performance of the system and that it can be used by nonlaboratorians with a high degree of confidence that reliable results will be obtained.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8504539

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  8 in total

1.  Point of care testing: randomised controlled trial of clinical outcome.

Authors:  J Kendall; B Reeves; M Clancy
Journal:  BMJ       Date:  1998-04-04

2.  The use of the iSTAT portable analyzer in patients undergoing cardiopulmonary bypass.

Authors:  N R Connelly; M Magee; B Kiessling
Journal:  J Clin Monit       Date:  1996-07

3.  Lack of arterial PO2 downregulation in Atlantic salmon (Salmo salar L.) during long-term normoxia and hyperoxia.

Authors:  T Kristensen; B O Rosseland; A Kiessling; B Djordevic; J C Massabau
Journal:  Fish Physiol Biochem       Date:  2010-03-11       Impact factor: 2.794

4.  Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients.

Authors:  Jacoline Steinfelder-Visscher; Steven Teerenstra; Jacqueline M T Klein Gunnewiek; Patrick W Weerwind
Journal:  J Extra Corpor Technol       Date:  2008-03

5.  Effect of nutritional status and sampling intensity on recovery after dorsal aorta cannulation in free-swimming Atlantic salmon (Salmo salar L.).

Authors:  B Djordjevic; T Kristensen; Ø Øverli; B O Rosseland; A Kiessling
Journal:  Fish Physiol Biochem       Date:  2009-10-25       Impact factor: 2.794

6.  Evaluation of the i-STAT (portable clinical analyser) for measuring haematological parameters in Atlantic cod (Gadus morhua) at different CO2 and temperature conditions.

Authors:  Radoslav S Borissov; Sølvi Espeland; Martin H Iversen
Journal:  Fish Physiol Biochem       Date:  2019-10-03       Impact factor: 2.794

Review 7.  Electrochemical Sensors for Clinic Analysis.

Authors:  You Wang; Hui Xu; Jianming Zhang; Guang Li
Journal:  Sensors (Basel)       Date:  2008-03-27       Impact factor: 3.576

8.  Developing an animal model to detect drug-drug interactions impacting drug-induced respiratory depression.

Authors:  Lin Xu; Ashok Chockalingam; Sharron Stewart; Katherine Shea; Murali K Matta; Suresh Narayanasamy; Nageswara R Pilli; Donna A Volpe; James Weaver; Hao Zhu; Michael C Davis; Rodney Rouse
Journal:  Toxicol Rep       Date:  2020-01-25
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.