Dose attenuation effect of hip prostheses in a 9-MV photon beam: commercial treatment planning system versus Monte Carlo calculations.

Asghar Mesbahi1, Farshad Seyed Nejad.   

Abstract

PURPOSE: The purpose of this study was to investigate the dosimetric effect of various hip prostheses on pelvis lateral fields treated by a 9-MV photon beam using Monte Carlo (MC) and effective path-length (EPL) methods.
MATERIAL AND METHODS: The head of the Neptun 10 pc linac was simulated using the MCNP4C MC code. The accuracy of the MC model was evaluated using measured dosimetric features including depth dose values and dose profiles in a water phantom. The Alfard treatment planning system (TPS) was used for EPL calculations. A virtual water phantom with dimensions of 30 x 30 x 30 cm(3) and a cube with dimensions of 4 x 4 x 4 cm(3) made of various metals centered in 12 cm depth was used for MC and EPL calculations. Various materials including titanium, Co-Cr-Mo, and steel alloys were used as hip prostheses.
RESULTS: Our results showed significant attenuation in absorbed dose for points after and inside the prostheses. Attenuations of 32%, 54% and 55% were seen for titanium, Co-Cr-Mo, and steel alloys, respectively, at a distance of 5 cm from the prosthesis. Considerable dose increase (up to 18%) was found at the water-prosthesis interface due to back-scattered electrons using the MC method. The results of EPL calculations for the titanium implant were comparable to the MC calculations. This method, however, was not able to predict the interface effect or calculate accurately the absorbed dose in the presence of the Co-Cr-Mo and steel prostheses.
CONCLUSION: The dose perturbation effect of hip prostheses is significant and cannot be predicted accurately by the EPL method for Co-Cr-Mo or steel prostheses. The use of MC-based TPS is recommended for treatments requiring fields passing through hip prostheses.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18085404     DOI: 10.1007/s11604-007-0181-z

Source DB:  PubMed          Journal:  Radiat Med        ISSN: 0288-2043


  21 in total

1.  Tolerances for the accuracy of photon beam dose calculations of treatment planning systems.

Authors:  J Venselaar; H Welleweerd; B Mijnheer
Journal:  Radiother Oncol       Date:  2001-08       Impact factor: 6.280

2.  How accurate is a CT-based dose calculation on a pencil beam TPS for a patient with a metallic prosthesis?

Authors:  R Roberts
Journal:  Phys Med Biol       Date:  2001-09       Impact factor: 3.609

3.  The effect of a metal hip prosthesis on the radiation dose in therapeutic photon beam irradiations.

Authors:  Sung-Yen Lin; Tieh-Chi Chu; Jao-Perng Lin; Mu-Tai Liu
Journal:  Appl Radiat Isot       Date:  2002-07       Impact factor: 1.513

4.  Monte Carlo dose calculations in the treatment of a pelvis with implant and comparison with pencil-beam calculations.

Authors:  Wolfram U Laub; Fridtjof Nüsslin
Journal:  Med Dosim       Date:  2003       Impact factor: 1.482

5.  Monte Carlo calculation of Varian 2300C/D Linac photon beam characteristics: a comparison between MCNP4C, GEANT3 and measurements.

Authors:  Asghar Mesbahi; Michael Fix; Mahmoud Allahverdi; Ellen Grein; Hossein Garaati
Journal:  Appl Radiat Isot       Date:  2005-03       Impact factor: 1.513

6.  Monte carlo dose calculations in conventional thorax fields for 60Co photons.

Authors:  Asghar Mesbahi; Mahmoud Allahverdi; Hossein Gheraati
Journal:  Radiat Med       Date:  2005-08

7.  Corrections to absorbed dose calculations for tissue inhomogeneities.

Authors:  M R Sontag; J R Cunningham
Journal:  Med Phys       Date:  1977 Sep-Oct       Impact factor: 4.071

8.  Effect of a femoral head prosthesis on megavoltage beam radiotherapy.

Authors:  P J Biggs; M D Russell
Journal:  Int J Radiat Oncol Biol Phys       Date:  1988-03       Impact factor: 7.038

9.  Primary mandibular reconstruction: analysis of 64 cases and evaluation of interface radiation dosimetry on bridging plates.

Authors:  P J Gullane
Journal:  Laryngoscope       Date:  1991-06       Impact factor: 3.325

10.  Monte Carlo verification of IMRT dose distributions from a commercial treatment planning optimization system.

Authors:  C M Ma; T Pawlicki; S B Jiang; J S Li; J Deng; E Mok; A Kapur; L Xing; L Ma; A L Boyer
Journal:  Phys Med Biol       Date:  2000-09       Impact factor: 3.609

View more
  7 in total

1.  Effects of Siemens TT-D carbon fiber table top on beam attenuation, and build up region of 6 MV photon beam.

Authors:  Asma Sheykhoo; Sara Abdollahi; Mohammad Hadi Hadizadeh Yazdi; Mahdi Ghorbani; Mohammad Mohammadi
Journal:  Rep Pract Oncol Radiother       Date:  2016-10-17

2.  Volumetric modulated arc therapy for prostate cancer patients with hip prosthesis.

Authors:  Ramachandran Prabhakar; Milind Kumar; Suja Cheruliyil; Silpa Jayakumar; Satheesan Balasubramanian; Jim Cramb
Journal:  Rep Pract Oncol Radiother       Date:  2013-04-28

3.  Estimation of organs doses and radiation-induced secondary cancer risk from scattered photons for conventional radiation therapy of nasopharynx: a Monte Carlo study.

Authors:  Asghar Mesbahi; Farshad Seyednejad; Amir Gasemi-Jangjoo
Journal:  Jpn J Radiol       Date:  2010-06-30       Impact factor: 2.374

4.  Consensus guidelines for postoperative stereotactic body radiation therapy for spinal metastases: results of an international survey.

Authors:  Kristin J Redmond; Simon S Lo; Scott G Soltys; Yoshiya Yamada; Igor J Barani; Paul D Brown; Eric L Chang; Peter C Gerszten; Samuel T Chao; Robert J Amdur; Antonio A F De Salles; Matthias Guckenberger; Bin S Teh; Jason Sheehan; Charles R Kersh; Michael G Fehlings; Moon-Jun Sohn; Ung-Kyu Chang; Samuel Ryu; Iris C Gibbs; Arjun Sahgal
Journal:  J Neurosurg Spine       Date:  2016-11-11

5.  Evaluation of the radiotherapy treatment planning in the presence of a magnetic valve tissue expander.

Authors:  Débora M Trombetta; Simone C Cardoso; Victor G L Alves; Alessandro Facure; Delano V S Batista; Ademir X da Silva
Journal:  PLoS One       Date:  2015-02-13       Impact factor: 3.240

6.  Dose comparison between Gafchromic film, XiO, and Monaco treatment planning systems in a novel pelvic phantom that contains a titanium hip prosthesis.

Authors:  Nicholas Ade; F C P du Plessis
Journal:  J Appl Clin Med Phys       Date:  2017-07-25       Impact factor: 2.102

7.  A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac.

Authors:  Mohammad Taghi Bahreyni Toossi; Marziyeh Behmadi; Mahdi Ghorbani; Hamid Gholamhosseinian
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.