Monte Carlo dose calculations in the treatment of a pelvis with implant and comparison with pencil-beam calculations.

Wolfram U Laub1, Fridtjof Nüsslin.   

Abstract

In the present paper, dose distribution calculated with the Monte Carlo code EGS4 and with a pencil-beam algorithm are compared for the treatment of a pelvis with an implant. Overestimations of dose values inside the target volume by the pencil-beam algorithm of up to 10% were found, which are attributed to the underestimation of the absorption of photons by the implant. The differences in dose distributions are also expressed by comparing the tumor control probability (TCP) of the Monte Carlo dose calculations with the TCP of the pencil-beam calculations. A TCP reduction of order of 30% was found.

Entities:  

Mesh:

Year:  2003        PMID: 14684187     DOI: 10.1016/j.meddos.2003.05.001

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  5 in total

1.  Dose attenuation effect of hip prostheses in a 9-MV photon beam: commercial treatment planning system versus Monte Carlo calculations.

Authors:  Asghar Mesbahi; Farshad Seyed Nejad
Journal:  Radiat Med       Date:  2007-12-25

2.  Dose distribution near thin titanium plate for skull fixation irradiated by a 4-MV photon beam.

Authors:  Tomohiro Shimozato; Keisuke Yasui; Ryota Kawanami; Kousaku Habara; Yuichi Aoyama; Katsuyoshi Tabushi; Yasunori Obata
Journal:  J Med Phys       Date:  2010-04

3.  Establishing radiation therapy treatment planning effects involving implantable pacemakers and implantable cardioverter-defibrillators.

Authors:  Michael S Gossman; Alison R Graves-Calhoun; Jeffrey D Wilkinson
Journal:  J Appl Clin Med Phys       Date:  2009-12-23       Impact factor: 2.102

4.  The accuracy of Acuros XB algorithm for radiation beams traversing a metallic hip implant - comparison with measurements and Monte Carlo calculations.

Authors:  Jarkko Ojala; Mika Kapanen; Petri Sipilä; Simo Hyödynmaa; Maunu Pitkänen
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

5.  A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac.

Authors:  Mohammad Taghi Bahreyni Toossi; Marziyeh Behmadi; Mahdi Ghorbani; Hamid Gholamhosseinian
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.