Volumetric modulated arc therapy for prostate cancer patients with hip prosthesis.

Ramachandran Prabhakar1, Milind Kumar2, Suja Cheruliyil2, Silpa Jayakumar2, Satheesan Balasubramanian2, Jim Cramb1.   

Abstract

AIM: To study the use of RapidArc techniques in the treatment of prostate cancer patients with hip prosthesis.
BACKGROUND: An important aspect of treatment planning is to achieve dose homogeneity inside the planning target volume (PTV). Especially for those patients presenting with hip prosthesis, it becomes a challenging task to achieve dose uniformity inside the PTV.
MATERIALS AND METHODS: Five prostate patients presenting with hip prosthesis who had undergone radical radiotherapy were selected for this study. Depending on the composition of prosthesis, a predefined set of Hounsfield values were assigned to each study set. RapidArc plans were generated on an Eclipse treatment planning system. Two arcs that include clockwise and counter-clockwise arcs were used in all these cases. To avoid beams passing through the prosthesis, a simple structure was defined around it with 1 cm margin and a strict dose constraint applied to the block during VMAT optimization.
RESULTS: The mean D2/D98 ratio of PTV for all the patients was 1.06 ± 0.01. The mean percentage rectum volume receiving 50 Gy, 60 Gy, 70 Gy and 75 Gy for all the patients were 33.1 ± 5.9, 21.7 ± 5.5, 13.8 ± 4.4 and 9.5 ± 3.0, respectively.
CONCLUSIONS: This study shows that using a double arc RapidArc technique is a simple and effective treatment method of treating prostate cancer in patients presenting with a hip prosthesis. The definition of a beam avoidance structure encompassing the prosthesis and applying strict dose constraints to it reduces the beam contribution to the prosthesis.

Entities:  

Keywords:  Prostate cancer; Prosthesis; Radiotherapy; Treatment planning; VMAT

Year:  2013        PMID: 24416555      PMCID: PMC3863326          DOI: 10.1016/j.rpor.2013.03.006

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  21 in total

1.  Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit.

Authors:  Sanford L Meeks; Joseph F Harmon; Katja M Langen; Twyla R Willoughby; Thomas H Wagner; Patrick A Kupelian
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

2.  An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: experimental and clinical studies.

Authors:  Mehran Yazdi; Mehran Yazdia; Luc Gingras; Luc Beaulieu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-07-15       Impact factor: 7.038

3.  Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media.

Authors:  A Ahnesjö
Journal:  Med Phys       Date:  1989 Jul-Aug       Impact factor: 4.071

4.  Radiotherapy doses of 80 Gy and higher are associated with lower mortality in men with Gleason score 8 to 10 prostate cancer.

Authors:  Niraj Pahlajani; Karen J Ruth; Mark K Buyyounouski; David Y T Chen; Eric M Horwitz; Gerald E Hanks; Robert A Price; Alan Pollack
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-07-15       Impact factor: 7.038

5.  High dose radiation delivered by intensity modulated conformal radiotherapy improves the outcome of localized prostate cancer.

Authors:  M J Zelefsky; Z Fuks; M Hunt; H J Lee; D Lombardi; C C Ling; V E Reuter; E S Venkatraman; S A Leibel
Journal:  J Urol       Date:  2001-09       Impact factor: 7.450

6.  Experimental validation of deterministic Acuros XB algorithm for IMRT and VMAT dose calculations with the Radiological Physics Center's head and neck phantom.

Authors:  Tao Han; Firas Mourtada; Kelly Kisling; Justin Mikell; David Followill; Rebecca Howell
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

7.  Intensity-modulated radiation therapy with noncoplanar beams for treatment of prostate cancer in patients with bilateral hip prosthesis-a case study.

Authors:  Chris Brooks; Rex Min Cheung; Rajat J Kudchadker
Journal:  Med Dosim       Date:  2009-04-18       Impact factor: 1.482

8.  Metal artifact reduction from reformatted projections for hip prostheses in multislice helical computed tomography: techniques and initial clinical results.

Authors:  Lifeng Yu; Hua Li; Jan Mueller; James M Kofler; Xin Liu; Andrew N Primak; Joel G Fletcher; Luis S Guimaraes; Thanila Macedo; Cynthia H McCollough
Journal:  Invest Radiol       Date:  2009-11       Impact factor: 6.016

9.  Comparison of dose distribution in IMRT and RapidArc technique in prostate radiotherapy.

Authors:  Wojciech Leszczyński; Krzysztof Slosarek; Marta Szlag
Journal:  Rep Pract Oncol Radiother       Date:  2012-08-10

10.  An opposed matched field IMRT technique for prostate cancer patients with bilateral prosthetic hips.

Authors:  Shahin Fattahi; Orest Z Ostapiak
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

View more
  7 in total

1.  Analyzing the performance of ArcCHECK diode array detector for VMAT plan.

Authors:  Rajesh Thiyagarajan; Arunai Nambiraj; Sujit Nath Sinha; Girigesh Yadav; Ashok Kumar; Vikraman Subramani
Journal:  Rep Pract Oncol Radiother       Date:  2015-12-02

Review 2.  Moderately hypofractionated prostate external-beam radiotherapy: an emerging standard.

Authors:  Jarad M Martin; Stephane Supiot; Paul J Keall; Charles N Catton
Journal:  Br J Radiol       Date:  2018-01-31       Impact factor: 3.039

3.  Volumetric modulated arc therapy in prostate cancer patients with metallic hip prostheses in a UK centre.

Authors:  Wee Loon Ng; John Brunt; Simon Temple; Mohammed Saipillai; Anoop Haridass; Helen Wong; Zafar Malik; Chinnamani Eswar
Journal:  Rep Pract Oncol Radiother       Date:  2015-04-14

4.  Dosimetry Effects Caused by Unilateral and Bilateral Hip Prostheses: A Monte Carlo Case Study in Megavoltage Photon Radiotherapy for Computed Tomography Data without Metal Artifacts.

Authors:  Courage Mahuvava; Frederik Carl Phillipus Du Plessis
Journal:  J Med Phys       Date:  2018 Oct-Dec

5.  Radiotherapy-induced toxicity in prostate cancer patients with hip prostheses.

Authors:  Andrea M Fischer; Peter J Hoskin
Journal:  Radiat Oncol       Date:  2022-01-17       Impact factor: 3.481

6.  Evaluation of VMAT Planning Strategies for Prostate Patients with Bilateral Hip Prosthesis.

Authors:  David To; Ilma Xhaferllari; Manju Liu; Jian Liang; Cory Knill; Sirisha Nandalur; Gary Gustafson; Danielle Lack
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec

7.  VMAT Optimization and Dose Calculation in the Presence of Metallic Hip Prostheses.

Authors:  Holly M Parenica; Panayiotis Mavroidis; William Jones; Gregory Swanson; Niko Papanikolaou; Sotirios Stathakis
Journal:  Technol Cancer Res Treat       Date:  2019 Jan-Dec
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.