Forehead reflectance oximetry: a clinical comparison with conventional digit sensors during laparotomic and laparoscopic abdominal surgery.

Andrea Casati1, Grazia Squicciarini, Marco Baciarello, Marta Putzu, Alessandra Salvadori, Guido Fanelli.   

Abstract

OBJECTIVE: This study compared the performance of forehead reflectance and conventional pulse oximetry (SpO(2)) in mechanically ventilated, anesthetized patients undergoing major abdominal surgery with either laparoscopic or laparotomic techniques.
METHODS: SpO(2) was continuously measured both with a forehead reflectance and a conventional digit sensor in 20 ASA physical status I-III, anesthetized patients undergoing either a laparotomic (group Laparotomy, n = 10) or laparoscopic (group Laparoscopy, n = 10) major abdominal surgery. SpO(2) values measured with the two sensors were continuously recorded at 10-second intervals during the entire procedure, and then analyzed for consistency.
RESULTS: In group Laparotomy finger and forehead readings remained consistently similar during the study period; on the contrary, in group Laparoscopy forehead SpO(2) readings showed a much wider variability. The mean (95% Confidence Intervals) difference between finger and forehead SpO(2) readings was 2.0% (-1.3% to +6.0%) in group Laparotomy and 2.5% (-3.5% to +8.4%) in group Laparoscopy (p = 0.001); however, in group Laparoscopy the mean difference between digit and forehead SpO(2) values increased from 0.2% (CI(95): -2.1% to +2.5%) during the first hour to 4.5% (CI(95): -2.5% to +11-5%) in the second hour (p = 0.0005), and 3.1% (CI(95): -2.4% to + 8.6%) in the third hour of surgery (p = 0.0005). Clinically relevant desaturation (decrease of SpO(2) < 89% for > or =30 s) detected with the forehead sensor in the Laparoscopy group was significantly more frequent and longer lasting than with conventional digit sensor. No differences were observed in group Laparotomy.
CONCLUSIONS: Forehead reflectance oximetry is as accurate as conventional digit based oximetry in mechanically ventilated patients undergoing laparotomic surgery in the supine position, but is significantly influenced by patient positioning and pneumoperitoneum during laparoscopic surgery.

Entities:  

Mesh:

Year:  2007        PMID: 17701387     DOI: 10.1007/s10877-007-9084-x

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  19 in total

1.  The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.

Authors:  Hartmut Gehring; Christoph Hornberger; Holger Matz; Ewald Konecny; Peter Schmucker
Journal:  Respir Care       Date:  2002-01       Impact factor: 2.258

Review 2.  Forehead oximetry in critically ill patients: the case for a new monitoring site.

Authors:  Richard D Branson; Paul D Mannheimer
Journal:  Respir Care Clin N Am       Date:  2004-09

3.  Non-invasive monitoring of finger arterial pressure in patients with Raynaud's phenomenon: effects of exposure to cold.

Authors:  T S Kurki; H I Piirainen; P T Kurki
Journal:  Br J Anaesth       Date:  1990-10       Impact factor: 9.166

4.  The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in Spo2 calculation.

Authors:  Kirk H Shelley; Doris Tamai; Denis Jablonka; Michael Gesquiere; Robert G Stout; David G Silverman
Journal:  Anesth Analg       Date:  2005-03       Impact factor: 5.108

5.  Comparison of a new forehead reflectance pulse oximeter sensor with a conventional digit sensor in pediatric patients.

Authors:  John W Berkenbosch; Joseph D Tobias
Journal:  Respir Care       Date:  2006-07       Impact factor: 2.258

6.  Standards for patient monitoring during anesthesia at Harvard Medical School.

Authors:  J H Eichhorn; J B Cooper; D J Cullen; W R Maier; J H Philip; R G Seeman
Journal:  JAMA       Date:  1986 Aug 22-29       Impact factor: 56.272

7.  Forehead pulse oximetry compared with finger pulse oximetry and arterial blood gas measurement.

Authors:  E Y Cheng; M B Hopwood; J Kay
Journal:  J Clin Monit       Date:  1988-07

8.  Accuracy of pulse oximetry in the intensive care unit.

Authors:  A Van de Louw; C Cracco; C Cerf; A Harf; P Duvaldestin; F Lemaire; L Brochard
Journal:  Intensive Care Med       Date:  2001-10       Impact factor: 17.440

9.  Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2004-01-13       Impact factor: 17.440

10.  Relationship between peripheral and central venous pressures in different patient positions, catheter sizes, and insertion sites.

Authors:  Mehmet Tugrul; Emre Camci; Kamil Pembeci; Ayman Al-Darsani; Lütfi Telci
Journal:  J Cardiothorac Vasc Anesth       Date:  2004-08       Impact factor: 2.628

View more
  1 in total

1.  Pulse oximetry and high-dose vasopressors: a comparison between forehead reflectance and finger transmission sensors.

Authors:  Nicolas Nesseler; Jean-Vincent Frénel; Yoann Launey; Jeff Morcet; Yannick Mallédant; Philippe Seguin
Journal:  Intensive Care Med       Date:  2012-08-07       Impact factor: 17.440

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.