Accuracy of pulse oximetry in the intensive care unit.

A Van de Louw1, C Cracco, C Cerf, A Harf, P Duvaldestin, F Lemaire, L Brochard.   

Abstract

OBJECTIVE: Pulse oximetry (SpO2) is a standard monitoring device in intensive care units (ICUs), currently used to guide therapeutic interventions. Few studies have evaluated the accuracy of SpO2 in critically ill patients. Our objective was to compare pulse oximetry with arterial oxygen saturation (SaO2) in such patients, and to examine the effect of several factors on this relationship.
DESIGN: Observational prospective study.
SETTING: A 26-bed medical ICU in a university hospital. PATIENTS: One hundred two consecutive patients admitted to the ICU in whom one or serial arterial blood gas analyses (ABGs) were performed and a reliable pulse oximeter signal was present.
INTERVENTIONS: For each ABG, we collected SaO2, SpO2, the type of pulse oximeter, the mode of ventilation and requirement for vasoactive drugs. MEASUREMENTS AND
RESULTS: Three hundred twenty-three data points were collected. The mean difference between SpO2 and SaO2 was -0.02% and standard deviation of the differences was 2.1%. From one sample to another, the fluctuations in SpO2 to arterial saturation difference indicated that SaO2 could not be reliably predicted from SpO2 after a single ABG. Subgroup analysis showed that the accuracy of SpO2 appeared to be influenced by the type of oximeter, the presence of hypoxemia and the requirement for vasoactive drugs. Finally, high SpO2 thresholds were necessary to detect significant hypoxemia with good sensitivity.
CONCLUSION: Large SpO2 to SaO2 differences may occur in critically ill patients with poor reproducibility of SpO2. A SpO2 above 94% appears necessary to ensure a SaO2 of 90%.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11685301     DOI: 10.1007/s001340101064

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  62 in total

1.  Pulse oximetry.

Authors:  Amal Jubran
Journal:  Intensive Care Med       Date:  2004-07-24       Impact factor: 17.440

2.  Clinical Characteristics and Outcomes Are Similar in ARDS Diagnosed by Oxygen Saturation/Fio2 Ratio Compared With Pao2/Fio2 Ratio.

Authors:  Wei Chen; David R Janz; Ciara M Shaver; Gordon R Bernard; Julie A Bastarache; Lorraine B Ware
Journal:  Chest       Date:  2015-12       Impact factor: 9.410

3.  Acute effects of hyperoxemia on dyspnoea and respiratory variables during pressure support ventilation.

Authors:  Carlo Alberto Volta; Valentina Alvisi; Sara Bertacchini; Elisabetta Marangoni; Riccardo Ragazzi; Marco Verri; Raffaele Alvisi
Journal:  Intensive Care Med       Date:  2006-01-24       Impact factor: 17.440

4.  Evaluation of a computer program for non-invasive determination of pulmonary shunt and ventilation-perfusion mismatch.

Authors:  Geoffrey G Lockwood; Nick L S Fung; J Gareth Jones
Journal:  J Clin Monit Comput       Date:  2014-01-09       Impact factor: 2.502

5.  Evaluation of a pulse oximeter sensor tester.

Authors:  Shuba Dugani; Iljaz Hodzovic; Seema Sindhakar; Aida Nadra; Clare Dunstan; Anthony Richard Wilkes; John Mecklenburgh
Journal:  J Clin Monit Comput       Date:  2011-05-18       Impact factor: 2.502

6.  The role of pulse oximetry in chiropractic practice: a rationale for its use.

Authors:  Michael W Hall; Anne M Jensen
Journal:  J Chiropr Med       Date:  2012-06

7.  Pulse oximetry and high-dose vasopressors: a comparison between forehead reflectance and finger transmission sensors.

Authors:  Nicolas Nesseler; Jean-Vincent Frénel; Yoann Launey; Jeff Morcet; Yannick Mallédant; Philippe Seguin
Journal:  Intensive Care Med       Date:  2012-08-07       Impact factor: 17.440

Review 8.  Asthma outcomes: pulmonary physiology.

Authors:  Robert S Tepper; Robert S Wise; Ronina Covar; Charles G Irvin; Carolyn M Kercsmar; Monica Kraft; Mark C Liu; George T O'Connor; Stephen P Peters; Ronald Sorkness; Alkis Togias
Journal:  J Allergy Clin Immunol       Date:  2012-03       Impact factor: 10.793

Review 9.  Sensors Capabilities, Performance, and Use of Consumer Sleep Technology.

Authors:  Massimiliano de Zambotti; Nicola Cellini; Luca Menghini; Michela Sarlo; Fiona C Baker
Journal:  Sleep Med Clin       Date:  2020-01-03

10.  The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study.

Authors:  Ben J Wilson; Hamish J Cowan; Jason A Lord; Dan J Zuege; David A Zygun
Journal:  BMC Emerg Med       Date:  2010-05-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.