The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.

Hartmut Gehring1, Christoph Hornberger, Holger Matz, Ewald Konecny, Peter Schmucker.   

Abstract

INTRODUCTION: Motion artifact and low perfusion often lead to faulty or absent pulse oximetry readings in clinical practice.
OBJECTIVE: Determine the impact of motion artifact and low perfusion on newly introduced pulse oximetry technologies during hypoxemic episodes in healthy volunteers.
METHODS: Five different pulse oximeters from 4 manufacturers (the Datex Ohmeda 3900P; the Agilent; the Nellcor N-3000; the Nellcor N-395; and the Schiller OX-1, which is the European version of the Ivy SatGuard 2000 with Masimo SET) were compared with respect to their ability (separated or in combination) to provide accurate readings in the presence of motion artifact and low perfusion. Four of these oximeters represent the latest available oximetry technology, and one (the N-3000) represents a previous generation of oximeters. Oxygen saturation values (S(pO(2))) and pulse rate from the oximeters were recorded during episodes of induced hypoxemia in 10 healthy volunteers. Standardized and repeatable motion artifacts were generated by a motion machine and by having the test subject perform tapping and scratching motions. Perfusion to the finger was reduced by an inflatable balloon impinging on the brachial artery. The pulse oximetry readings from the test oximeters were compared to readings from control pulse oximeters on the unperturbed reference hand. The pulse rates from the test oximeters were compared to the electrocardiographically-measured heart rate.
RESULTS: The frequency of faulty readings was increased by increasing motion interference and decreasing perfusion. The S(pO(2)) deviation was within +/- 3% of the reference reading > 95% of the time for all instruments during the control desaturation period in the absence of motion and with normal perfusion. With the combination of motion and low perfusion, the S(pO(2)) error was within +/- 3% less than 62% of the time for all oximeters tested. A significant difference in the frequency of large S(pO(2)) errors was observed only in the direct comparison of the N-395 and N-3000. The N-395 exhibited less frequent S(pO(2)) error exceeding 6% of S(pO(2)) in the combination of the most challenging situations (motion and motion with reduced perfusion). In the same situation the Datex-Ohmeda 3900P and Nellcor N-3000 showed significantly higher pulse rate errors than the other devices (Datex-Ohmeda 3900P 53% of the time and N-3000 37% of the time).
CONCLUSIONS: The established model of creating motion artifact and low perfusion is capable of simulating a hierarchy of severe clinical situations. With solely motion or solely reduced perfusion the percentage of errors exceeding +/- 3% of S(pO(2)) increased by 20% and 10%, respectively, compared to the control period. Simultaneous presence of motion and reduced perfusion leads to a relative incidence of > 35% of errors > 3% of S(pO(2)) for the various oximeters. In this situation the N-3000 and the Datex-Ohmeda 3900P exhibited differences between estimated pulse rate and electrocardiographically-measured heart rate > 25 beats/min > 37% of the time.

Entities:  

Mesh:

Year:  2002        PMID: 11749687

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  17 in total

1.  Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

2.  Perfusion index as a possible predictor for postanesthetic shivering.

Authors:  Chiharu Kuroki; Kohei Godai; Maiko Hasegawa-Moriyama; Tamotsu Kuniyoshi; Akira Matsunaga; Yuichi Kanmura; Tomoyuki Kuwaki
Journal:  J Anesth       Date:  2013-06-20       Impact factor: 2.078

Review 3.  Wireless and battery-free platforms for collection of biosignals.

Authors:  Tucker Stuart; Le Cai; Alex Burton; Philipp Gutruf
Journal:  Biosens Bioelectron       Date:  2021-01-23       Impact factor: 10.618

Review 4.  From system to organ to cell: oxygenation and perfusion measurement in anesthesia and critical care.

Authors:  Patrick Schober; Lothar A Schwarte
Journal:  J Clin Monit Comput       Date:  2012-03-22       Impact factor: 2.502

5.  Evaluating Innovative In-Ear Pulse Oximetry for Unobtrusive Cardiovascular and Pulmonary Monitoring During Sleep.

Authors:  Boudewijn Venema; Johannes Schiefer; Vladimir Blazek; Nikolai Blanik; Steffen Leonhardt
Journal:  IEEE J Transl Eng Health Med       Date:  2013-08-08       Impact factor: 3.316

6.  Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

Authors:  John R Feiner; Mark D Rollins; Jeffrey W Sall; Helge Eilers; Paul Au; Philip E Bickler
Journal:  Anesth Analg       Date:  2013-03-11       Impact factor: 5.108

7.  Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2004-01-13       Impact factor: 17.440

8.  Forehead reflectance oximetry: a clinical comparison with conventional digit sensors during laparotomic and laparoscopic abdominal surgery.

Authors:  Andrea Casati; Grazia Squicciarini; Marco Baciarello; Marta Putzu; Alessandra Salvadori; Guido Fanelli
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

9.  Comparison of two new generation pulse oximeters with arterial oxygen saturation in critically ill children.

Authors:  Bipin Jose; Rakesh Lodha; S K Kabra
Journal:  Indian J Pediatr       Date:  2014-03-15       Impact factor: 1.967

Review 10.  Accuracy of pulse oximeters in measuring oxygen saturation in patients with poor peripheral perfusion: a systematic review.

Authors:  Khashayar Poorzargar; Chi Pham; Jennita Ariaratnam; Kang Lee; Matteo Parotto; Marina Englesakis; Frances Chung; Mahesh Nagappa
Journal:  J Clin Monit Comput       Date:  2022-02-04       Impact factor: 1.977

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.