Accuracy, repeatability, and reproducibility of Artemis very high-frequency digital ultrasound arc-scan lateral dimension measurements.

Dan Z Reinstein1, Timothy J Archer, Ronald H Silverman, D Jackson Coleman.   

Abstract

PURPOSE: To determine the accuracy, repeatability, and reproducibility of measurement of lateral dimensions using the Artemis (Ultralink LLC) very high-frequency (VHF) digital ultrasound (US) arc scanner.
SETTING: London Vision Clinic, London, United Kingdom.
METHODS: A test object was measured first with a micrometer and then with the Artemis arc scanner. Five sets of 10 consecutive B-scans of the test object were performed with the scanner. The test object was removed from the system between each scan set. One expert observer and one newly trained observer separately measured the lateral dimension of the test object. Two-factor analysis of variance was performed. The accuracy was calculated as the average bias of the scan set averages. The repeatability and reproducibility coefficients were calculated. The coefficient of variation (CV) was calculated for repeatability and reproducibility.
RESULTS: The test object was measured to be 10.80 mm wide. The mean lateral dimension bias was 0.00 mm. The repeatability coefficient was 0.114 mm. The reproducibility coefficient was 0.026 mm. The repeatability CV was 0.38%, and the reproducibility CV was 0.09%. There was no statistically significant variation between observers (P = .0965). There was a statistically significant variation between scan sets (P = .0036) attributed to minor vertical changes in the alignment of the test object between consecutive scan sets.
CONCLUSION: The Artemis VHF digital US arc scanner obtained accurate, repeatable, and reproducible measurements of lateral dimensions of the size commonly found in the anterior segment.

Mesh:

Year:  2006        PMID: 17081860      PMCID: PMC2600896          DOI: 10.1016/j.jcrs.2006.07.017

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  9 in total

1.  Predicting sulcus size using ocular measurements.

Authors:  M Pop; Y Payette; M Mansour
Journal:  J Cataract Refract Surg       Date:  2001-07       Impact factor: 3.351

2.  Correlation between different measurements within the eye relative to phakic intraocular lens implantation.

Authors:  Liliana Werner; Andrea M Izak; Suresh K Pandey; David J Apple; Rupal H Trivedi; Josef M Schmidbauer
Journal:  J Cataract Refract Surg       Date:  2004-09       Impact factor: 3.351

3.  Magnetic resonance imaging and Orbscan assessment of the anterior chamber.

Authors:  Antonio M Fea; Federica Annetta; Stefano Cirillo; Delia Campanella; Massimo De Giuseppe; Daniele Regge; Federico M Grignolo
Journal:  J Cataract Refract Surg       Date:  2005-09       Impact factor: 3.351

Review 4.  Phakic intraocular lenses.

Authors:  Carlo F Lovisolo; Dan Z Reinstein
Journal:  Surv Ophthalmol       Date:  2005 Nov-Dec       Impact factor: 6.048

5.  Measurement error.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-09-21

6.  Measurement error proportional to the mean.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-07-13

7.  Very high frequency ultrasound analysis of a new phakic posterior chamber intraocular lens in situ.

Authors:  D Y Kim; D Z Reinstein; R H Silverman; D J Najafi; S C Belmont; A P Hatsis; G W Rozakis; D J Coleman
Journal:  Am J Ophthalmol       Date:  1998-05       Impact factor: 5.258

8.  High-frequency ultrasound measurement of the thickness of the corneal epithelium.

Authors:  D Z Reinstein; R H Silverman; D J Coleman
Journal:  Refract Corneal Surg       Date:  1993 Sep-Oct

9.  Very high frequency ultrasound biometry of the anterior and posterior chamber diameter.

Authors:  Mark J Rondeau; György Barcsay; Ronald H Silverman; Dan Z Reinstein; Rupa Krishnamurthy; Almira Chabi; Ted Du; D Jackson Coleman
Journal:  J Refract Surg       Date:  2004 Sep-Oct       Impact factor: 3.573

  9 in total
  8 in total

1.  Angle-to-angle and spur-to-spur distance analysis with high-resolution optical coherence tomography.

Authors:  Robert Montés-Micó; Pedro Tañá-Rivero; Salvador Aguilar-Córcoles; María Ruiz-Santos; María Dolores Rodríguez-Carrillo; Ramón Ruiz-Mesa
Journal:  Eye Vis (Lond)       Date:  2020-08-16

2.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

3.  Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using artemis VHF digital ultrasound.

Authors:  Dan Z Reinstein; Timothy J Archer; Ronald H Silverman; Mark J Rondeau; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-02       Impact factor: 3.573

4.  Assessment of the influence of lung inflation state on the quantitative parameters derived from hyperpolarized gas lung ventilation MRI in healthy volunteers.

Authors:  Paul J C Hughes; Laurie Smith; Ho-Fung Chan; Bilal A Tahir; Graham Norquay; Guilhem J Collier; Alberto Biancardi; Helen Marshall; Jim M Wild
Journal:  J Appl Physiol (1985)       Date:  2018-11-09

5.  Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness.

Authors:  Kelechi C Ogbuehi; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2012-05-21

6.  Water-suppression cycling 3-T cardiac 1 H-MRS detects altered creatine and choline in patients with aortic or mitral stenosis.

Authors:  Belinda Ding; Mark Peterzan; Ferenc E Mózes; Oliver J Rider; Ladislav Valkovič; Christopher T Rodgers
Journal:  NMR Biomed       Date:  2021-04-07       Impact factor: 4.478

7.  Visual Outcomes, Footplate Position and Vault Achieved with the Visian Implantable Collamer Lens for Myopic Astigmatism.

Authors:  Dan Z Reinstein; Ryan S Vida; Timothy J Archer
Journal:  Clin Ophthalmol       Date:  2021-11-20

8.  Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes.

Authors:  Haya M Al-Farhan; Wafa'a Majed Al-Otaibi
Journal:  Clin Ophthalmol       Date:  2012-07-06
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.