| Literature DB >> 34848942 |
Dan Z Reinstein1,2,3,4, Ryan S Vida1, Timothy J Archer1.
Abstract
PURPOSE: Report the outcomes of the implantable collamer lens (ICL) in myopic astigmatism using very high-frequency (VHF) digital ultrasound sizing.Entities:
Keywords: ICL; VHF digital ultrasound; footplate; implantable collamer lens; phakic IOL
Year: 2021 PMID: 34848942 PMCID: PMC8612663 DOI: 10.2147/OPTH.S330879
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Diagnostic and Examination Schedule
| Examination | Pre-Op | 1 Day | 1 Week | 1 Month | 3 Months | 12 Months |
|---|---|---|---|---|---|---|
| Subjective manifest refraction (CSV-1000) | ||||||
| UDVA | ||||||
| CDVA | ||||||
| Cycloplegic refraction | ||||||
| Slit lamp examination | ||||||
| Fundus examination | ||||||
| Intraocular pressure (Goldmann) | ||||||
| Contrast sensitivity (CSV-1000) | ||||||
| Ocular Response Analyzer intraocular pressure (ORA) | ||||||
| Dark/Scotopic pupil diameter (Procyon) | ||||||
| Corneal topography (Atlas) | ||||||
| Corneal tomography (Pentacam) | ||||||
| Corneal tomography (MS-39) | ||||||
| Anterior segment (MS-39) | ||||||
| Wavefront (Osiris) | ||||||
| Wavefront (WASCA) | ||||||
| Optical quality (HD Analyzer) | ||||||
| Endothelial cell count (SP-3000P) | ||||||
| Retinal OCT (RTVue) | ||||||
| VHF digital ultrasound (Artemis Insight 100) | ||||||
| Quality of vision questionnaire (QoV) |
Notes: CSV 1000 – Vector Vision, Greenville, OH. Procyon - Procyon Instruments Ltd, London, United Kingdom. Atlas – Carl Zeiss Meditec, Jena, Germany. Pentacam – Oculus, Wetzlar, Germany. MS-39 – CSO, Florence, Italy. Osiris – CSO, Florence, Italy. WASCA – Carl Zeiss Meditec, Jena, Germany. HD Analyzer – Visiometrics, Costa Mesa, CA. ORA – Reichert Ophthalmic Instruments Inc., Buffalo, NY. SP-3000P – Topcon, Tokyo, Japan. RTVue – Optovue, Freemont, CA. Insight 100 – ArcScan Inc, Golden, CO.
Figure 1Ultrasound B-scan using the Insight 100 (ArcScan Inc, Golden, CO). Red caliper lines are shown measuring anterior chamber depth (ACD), sulcus-to-sulcus (STS), and crystalline lens rise from the sulcus plane (STSL). Other dimensions that can be measured from this scan include anterior chamber angle, angle-to-angle (ATA), ciliary body inner diameter (CBID), zonule-to-zonule (ZTZ), and crystalline lens rise from the zonular plane (ZTZL).
Population Demographics
| Eyes (patients) | 42 (21 patients) |
| Age (years) | 31±6 (24 to 47) |
| Gender (M/F %, n) | 48% (20)/52% (22) |
| Attempted maximum myopic meridian (D) | −11.59±3.47 (−5.93 to −17.81) |
| Attempted spherical equivalent refraction (D) | −10.86±3.37 (−4.40 to −16.98) |
| Attempted refractive cylinder (D) | −1.45±1.15 (0.00 to −4.25) |
| Dark pupil diameter (mm) | 6.64±0.87 (4.80 to 8.44) |
| Scotopic pupil diameter (mm) | 5.55±0.90 (3.44 to 6.77) |
| Normal topography (%, n) | 71% (30) |
| FFKC topography (%, n) | 14% (6) |
| KC topography (%, n) | 14% (6) |
Abbreviations: FFKC, forme fruste keratoconus; KC, keratoconus.
ICL Parameters and Anterior Chamber Dimensions
| ICL maximum myopia power (D) | −12.31±3.06 (−7.00 to −17.50) |
| ICL cylinder power (D) | 1.35±1.13 (0.00 to 4.50) |
| ICL size: 12.6 mm/13.2 mm (%, n) | 52% (20)/48% (22) |
| ICL model: EVO/EVO+ (%, n) | 26% (11)/74% (31) |
| Primary/backup lens used (%, n) | 98% (41)/2% (1) |
| White-to-white horizontal diameter (mm) | 11.75±0.34 (11.12 to 12.49) |
| Sulcus-to-sulcus horizontal diameter (mm) | 11.93±0.43 (11.22 to 12.70) |
| Anterior chamber depth (mm) | 3.02±0.31 (2.56 to 3.69) |
| Crystalline lens rise from sulcus plane (µm) | 0.57±0.19 (0.10 to 0.89) |
| ICL central vault at 1 month (µm) | 506±233 (114 to 924) |
| Anterior chamber angle preop (°) | 42±7 (28 to 65) |
| Anterior chamber angle at 1 month (°) | 24±4 (14 to 36) |
Figure 2Nine standard graphs for reporting refractive surgery showing the visual and refractive outcomes for 42 phakic IOL treatments using the Visian V4c implantable collamer lens (Staar Surgical, Monrovia, CA).
Stability of Spherical Equivalent Refraction and SimK Keratometry
| Preoperative | 3 Months | 12 Months | Pre-post Change* | Pre-Post Change Within ±0.50 D | 3–12 Months Change | 3–12 Months Change Within ±0.50 D | |
|---|---|---|---|---|---|---|---|
| Eyes with refraction | 42 | 40 | 36 | 42 | 42 | 36 | 36 |
| Spherical equivalent refraction adjusted for intended target (D) | −10.86±3.37 (−4.40 to −16.98) | −0.07±0.30 (−0.77 to +0.59) | −0.19±0.36 (−1.07 to +0.55) | −10.67±3.25 (−16.63 to −3.88) | - | −0.12±0.38 (−1.00 to +0.63) | 83% |
| Refractive cylinder magnitude (D) | −1.45±1.15 (0.00 to −4.25) | −0.39±0.30 (0.00 to −1.25) | −0.38±0.35 (0.00 to −1.50) | −1.27±1.17 (−4.00 to +0.75) | - | +0.03±0.24 (−0.50 to +0.50) | 100% |
| Eyes with topography | 42 | 40 | 36 | 42 | 42 | 36 | 36 |
| Average keratometry (D) | 44.61±1.45 (42.28 to 48.01) | 44.64±1.49 (42.12 to 47.63) | 44.59±1.44 (42.17 to 47.58) | −0.03±0.21 (−0.67 to +0.30) | 98% (41) | −0.01±0.11 (−0.27 to +0.14) | 100% |
| Corneal astigmatism magnitude (D) | 1.24±0.66 (0.19 to 3.19) | 1.38±0.78 (0.18 to 3.41) | 1.50±0.74 (0.33 to 3.53) | +0.15±0.28 (−0.35 to +0.69) | 86% (36) | +0.01±0.21 (−0.45 to +0.37) | 100% |
Notes: *Using the last appointment, 12 months if available, 3 months otherwise.
Figure 3Vector analysis of refractive cylinder displayed as polar plots for target induced astigmatism vector (TIA), surgically induced astigmatism vector (SIA), difference vector (DV), and correction index (CI).
Vector Analysis of Refractive Cylinder and Corneal Astigmatism
| Toric Lenses | Spherical Lenses | Corneal Astigmatism | |
|---|---|---|---|
| Eyes | 33 | 9 | 42 |
| Target induced astigmatism vector (D) | 1.80±1.05 | – | – |
| Arithmetic mean | (0.50 to 4.25) | ||
| Summated vector mean | 0.62 Ax 5 | ||
| Surgically induced astigmatism vector (D) | 1.95±1.07 | 0.39±0.25 | 0.35±0.19 |
| Arithmetic mean | (0.27 to 4.70) | (0.22 to 0.99) | (0.04 to 0.70) |
| Summated vector mean | 0.45 Ax 5 | 0.08 Ax 123 | 0.24 Ax 100 |
| Correction index – geometric mean | 1.09 (0.53 to 3.31) | – | – |
| Difference vector (D) | 0.41±0.30 | 0.39±0.25 | – |
| Arithmetic mean | (0.12 to 1.58) | (0.22 to 0.99) | |
| Summated vector mean | 0.16 Ax 8 | 0.08 Ax 33 | |
| Index of success – geometric mean | 0.22 (0.04 to 2.39) | – | – |
| Angle of error (°) | |||
| Arithmetic mean | −1.8±6.7 | – | – |
| Absolute mean | 3.8±5.8 | ||
| Postop cylinder magnitude ≤0.25 D* | 52% (17) | 78% (7) | – |
| Postop cylinder magnitude ≤0.50 D* | 76% (25) | 100% (9) | – |
| Postop cylinder magnitude ≤1.00 D* | 97% (32) | 100% (9) | – |
| Postop cylinder magnitude ≤2.00 D* | 100% (33) | 100% (9) | – |
| Cylinder induced/SIA corneal astigmatism ≥0.50 D | 3% (1) | 0% (0) | 24% (10) |
| Cylinder induced/SIA corneal astigmatism ≥1.00 D | 0% (0) | 0% (0) | 0% (0) |
| Cylinder induced/SIA corneal astigmatism ≥2.00 D | 0% (0) | 0% (0) | 0% (0) |
Note: *Relative to zero cylinder, rather than adjusted for ICL calculated target cylinder.
Figure 4Surgically induced astigmatism vector (SIA) for corneal astigmatism displayed as a double angle vector diagram (DAVD) (A) and polar plot (B). The vector mean and standard deviation ellipse is shown on the DAVD. The graphs were generated using the ISRS Alpins ASSORT calculator.
Contrast Sensitivity
| 3 CPD | 6 CPD | 12 CPD | 18 CPD | |
|---|---|---|---|---|
| Preoperative | 1.64±0.19 (1.17 to 2.08) | 1.68±0.26 (0.91 to 2.14) | 1.28±0.30 (0.61 to 1.69) | 0.85±0.28 (0.17 to 1.40) |
| Postoperative* | 1.67±0.15 (1.34 to 1.93) | 1.82±0.20 (1.38 to 2.29) | 1.44±0.34 (0.61 to 1.99) | 1.00±0.33 (0.17 to 1.55) |
| Change | 0.04±0.21 (−0.30 to 0.76) | 0.14±0.22 (−0.30 to 0.91) | 0.14±0.31 (−0.64 to 0.91) | 0.14±0.29 (−0.64 to 0.91) |
| Contrast increase more than 0.25 log units | 15% (6) | 25% (10) | 30% (12) | 30% (12) |
| Contrast decrease more than 0.25 log units | 5% (2) | 5% (2) | 5% (2) | 5% (2) |
Note: *12 months if available, 3 months otherwise.
Figure 5Ultrasound B-scan using the Insight 100 (ArcScan Inc, Golden CO) showing position of the ICL footplate. (A) shows the footplate positioned just anterior to the zonules flexing up toward the sulcus (zonule position). (B) shows the footplate inserting directly into the ciliary body (ciliary position). (C) shows the footplate entering directly into the sulcus (sulcus position) with an associated tilt to the lens.
Endothelial Cell Count
| Preoperative | 3 Months | 12 Months | Pre-Post Change | Decreased ≥500 Cells/mm2 | Decreased ≥1000 Cells/mm2 | 3–12 Months Change | |
|---|---|---|---|---|---|---|---|
| Eyes with ECC | 38 | 40 | 36 | 36 | 36 | 36 | 36 |
| Endothelial cell count (cells/mm2) | 2646±221 (2144 to 3145) | 2458±349 (1346 to 3025) | 2490±359 (1447 to 3091) | −153±353 (−1238 to 639) | 14% (5) | 6% (2) | 40±225 (−431 to 667) |