Magnetic resonance imaging and Orbscan assessment of the anterior chamber.

Antonio M Fea1, Federica Annetta, Stefano Cirillo, Delia Campanella, Massimo De Giuseppe, Daniele Regge, Federico M Grignolo.   

Abstract

PURPOSE: To evaluate the correlation between white-to-white (WTW) distance as assessed by Orbscan II (Bausch & Lomb) and high-resolution magnetic resonance imaging (MRI) sulcus-sulcus (S-S) measures and to analyze the correlation between age and anterior chamber parameters.
SETTING: Istituto per la Ricerca e la Cura del Cancro, Candiolo, Torino, Italy.
METHODS: Eighty-eight patients had MRI with a 1.5 Tesla imager (GE Medical Systems) using a 3-inch circular coil. T1 weighted fast spin-echo scans were performed on the axial plane. Orbscan II was also used to measure WTW distance, K, anterior chamber depth (ACD), and lens curvature. One eye was randomly selected for analysis. Measures were compared using Spearman correlation, paired Student t test, and analysis of variance (ANOVA; post hoc: Scheffé). Intersession variability was performed in 10 patients. Ten MRI images were randomly selected and measured by 2 operators in a masked fashion (intrasession variability).
RESULTS: Intersession and intrasession correlation was good for MRI (r = 0.89 and r = 0.92, respectively), and intersession was good for Orbscan (r = 0.91). When comparing Orbscan II and MRI results, the ACD was well correlated and not significantly different, whereas the S-S and the WTW measures were not correlated and significantly different. Lens thickness and lens diameter were directly correlated with age; S-S, WTW, lens curvature, and ACD were inversely correlated with age.
CONCLUSIONS: Current methods of measuring the WTW were poorly correlated with anatomical measures. The S-S diameter and other anterior chamber structures significantly change with age, which could give rise to potential problems with posterior chamber phakic intraocular lens sizing position.

Entities:  

Mesh:

Year:  2005        PMID: 16246773     DOI: 10.1016/j.jcrs.2005.02.040

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  White-to-white corneal diameter, pupil diameter, central corneal thickness and thinnest corneal thickness values of emmetropic subjects.

Authors:  Juan A Sanchis-Gimeno; Daniel Sanchez-Zuriaga; Francisco Martinez-Soriano
Journal:  Surg Radiol Anat       Date:  2011-10-22       Impact factor: 1.246

2.  Accuracy, repeatability, and reproducibility of Artemis very high-frequency digital ultrasound arc-scan lateral dimension measurements.

Authors:  Dan Z Reinstein; Timothy J Archer; Ronald H Silverman; D Jackson Coleman
Journal:  J Cataract Refract Surg       Date:  2006-11       Impact factor: 3.351

3.  Ocular biometry through fully refocused steady-state magnetic resonance imaging sequence: reliability and agreement with the IOLMaster® 500.

Authors:  Lorenzo Ismael Perez-Sanchez; Julia Gutierrez-Vazquez; Maria Satrustegui-Lapetra; Francisco Ferreira-Manuel; Juan Jose Arevalo-Manso; Juan Jesus Gomez-Herrera; Juan Jose Criado-Alvarez
Journal:  Int Ophthalmol       Date:  2021-02-23       Impact factor: 2.031

4.  Angle-to-angle and spur-to-spur distance analysis with high-resolution optical coherence tomography.

Authors:  Robert Montés-Micó; Pedro Tañá-Rivero; Salvador Aguilar-Córcoles; María Ruiz-Santos; María Dolores Rodríguez-Carrillo; Ramón Ruiz-Mesa
Journal:  Eye Vis (Lond)       Date:  2020-08-16

5.  7 Tesla MR imaging of the human eye in vivo.

Authors:  Kathryn Richdale; Peter Wassenaar; Katharine Teal Bluestein; Amir Abduljalil; John A Christoforidis; Titus Lanz; Michael V Knopp; Petra Schmalbrock
Journal:  J Magn Reson Imaging       Date:  2009-11       Impact factor: 4.813

6.  Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using artemis VHF digital ultrasound.

Authors:  Dan Z Reinstein; Timothy J Archer; Ronald H Silverman; Mark J Rondeau; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-02       Impact factor: 3.573

7.  Phakic intraocular lens: Getting the right size.

Authors:  Kalyaani Deshpande; Rushad Shroff; Partha Biswas; Kamal Kapur; Naren Shetty; Ann Sarah Koshy; Pooja Khamar
Journal:  Indian J Ophthalmol       Date:  2020-12       Impact factor: 1.848

8.  Morphometric analysis of the lens in human aniridia and mouse Small eye.

Authors:  Anna Voskresenskaya; Nadezhda Pozdeyeva; Yevgeniy Batkov; Tatyana Vasilyeva; Andrey Marakhonov; Richard A West; Jeffrey L Caplan; Ales Cvekl; Yan Wang; Melinda K Duncan
Journal:  Exp Eye Res       Date:  2020-11-26       Impact factor: 3.467

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.