[Evaluation of dynamic contour tonometry in penetrating keratoplasties].

An Viestenz1, A Langenbucher, B Seitz, Ar Viestenz.   

Abstract

PURPOSE: The PASCAL dynamic contour tonometer (DCT) adapted to the corneal geometry is a new approach for digital measurement of intraocular pressure (IOP). The purpose of this study was to evaluate the DCT vs Goldmann applanation tonometry (GAT) in penetrating keratoplasties.
METHOD: We included 92 consecutive eyes after penetrating keratoplasty (57 without sutures, 32 with 2 running sutures, 3 with 1 running suture) in this prospective cross-sectional study. The mean outcome measurements were central corneal power (Zeiss ophthalmometer) and central corneal thickness (ultrasound pachymetry Tomey AL-1000). IOP was evaluated in consecutive measurements including 3x DCT and 1x GAT.
RESULTS: The mean IOP measured with DCT was 18.3+/-4.0 mmHg in contrast to GAT with 14.3+/-4.1 mmHg (p<0.001). Both methods showed a strong correlation of IOP values (r=0.74, p<0.001). The quality of measurement in keratoplasties after complete suture removal was between Q1 and Q3 in 88% with DCT in contrast to 59% with two running sutures. Three values with a reproducibility of 0.97 (Cronbach's alpha) could be obtained in 53 eyes without sutures and with a reproducibility of 0.94 (Cronbach's alpha) in 20 eyes (62.5%) with 2 running sutures. DCT or GAT did not correlate with central corneal thickness. The mean astigmatism did not differ between keratoplasties without sutures (4.86+/-3.35 D) or with two running sutures (4.2+/-2.8 D, p=0.38).
CONCLUSIONS: The DCT provides reproducible IOP measurements in penetrating keratoplasties with significantly higher values in comparison to GAT. Nevertheless, as known for GAT, postoperative intrastromal corneal sutures after penetrating keratoplasty disturb the exact DCT measurement as well.

Entities:  

Mesh:

Year:  2006        PMID: 16896684     DOI: 10.1007/s00347-006-1395-7

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  17 in total

1.  [Further contribution to applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1961-06       Impact factor: 3.250

2.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

3.  Comparison of dynamic contour tonometry with goldmann applanation tonometry.

Authors:  Claude Kaufmann; Lucas M Bachmann; Michael A Thiel
Journal:  Invest Ophthalmol Vis Sci       Date:  2004-09       Impact factor: 4.799

4.  [Reproducibility of dynamic contour tonometry. Comparison with TonoPenXL and Goldmann applanation tonometry - a clinical study on 323 normal eyes].

Authors:  A Viestenz; A Langenbucher; A Viestenz
Journal:  Klin Monbl Augenheilkd       Date:  2006-10       Impact factor: 0.700

Review 5.  Sources of error with use of Goldmann-type tonometers.

Authors:  M M Whitacre; R Stein
Journal:  Surv Ophthalmol       Date:  1993 Jul-Aug       Impact factor: 6.048

6.  Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states.

Authors:  Christoph Kniestedt; Michelle Nee; Robert L Stamper
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-10-13       Impact factor: 3.117

7.  Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis.

Authors:  Claude Kaufmann; Lucas M Bachmann; Michael A Thiel
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-09       Impact factor: 4.799

8.  [Impact of prostaglandin-F(2alpha)-analogues and carbonic anhydrase inhibitors on central corneal thickness -- a cross-sectional study on 403 eyes].

Authors:  A Viestenz; P Martus; U Schlötzer-Schrehardt; A Langenbucher; C Y Mardin
Journal:  Klin Monbl Augenheilkd       Date:  2004-09       Impact factor: 0.700

9.  Applanation tonometry and correction according to corneal thickness.

Authors:  R Stodtmeister
Journal:  Acta Ophthalmol Scand       Date:  1998-06

10.  Dynamic contour tonometry for post-LASIK intraocular pressure measurements.

Authors:  I Duba; A C Wirthlin
Journal:  Klin Monbl Augenheilkd       Date:  2004-05       Impact factor: 0.700

View more
  4 in total

1.  [A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann applanation tonometry].

Authors:  W Schreiber; C K Vorwerk; A Langenbucher; W Behrens-Baumann; A Viestenz
Journal:  Ophthalmologe       Date:  2007-04       Impact factor: 1.059

2.  [Reduction of intraocular pressure and ocular pulse amplitude during general anesthesia].

Authors:  H Zuche; E Morinello; A Viestenz; M Fiorentzis; B Seitz; T Volk; A Viestenz
Journal:  Ophthalmologe       Date:  2015-09       Impact factor: 1.059

3.  [Evaluation of dynamic contour tonometry in keratoconus].

Authors:  N Schädle; J D Unterlauft; T Klink; G Geerling
Journal:  Ophthalmologe       Date:  2009-06       Impact factor: 1.059

4.  [Intraocular pressure elevation after vitrectomy-Goldmann applanation tonometry measures lower intraocular pressure than dynamic contour tonometry].

Authors:  Sebastian Bäurle; Anja Viestenz; Berthold Seitz; Arne Viestenz
Journal:  Ophthalmologe       Date:  2021-07-06       Impact factor: 1.059

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.