Literature DB >> 15326129

Comparison of dynamic contour tonometry with goldmann applanation tonometry.

Claude Kaufmann1, Lucas M Bachmann, Michael A Thiel.   

Abstract

PURPOSE: The dynamic contour tonometer (DCT; Pascal tonometer) is a novel tonometer designed to measure intraocular pressure (IOP) independent of corneal properties. The purpose of this study was a comparison of the DCT with the Goldmann applanation tonometer (GAT) with respect to mean of IOP readings, the influence of ocular structural factors on IOP readings, and both intra- and interobserver variability, in a large group of healthy subjects.
METHODS: In a prospective study of 228 eyes, IOP measurements by GAT and DCT were compared, and the effects of central corneal thickness (CCT), corneal curvature, axial length, and anterior chamber depth were analyzed. To evaluate intra- and interobserver variability, IOP was measured in eight eyes by four observers.
RESULTS: There was a high concordance between the IOP readings obtained by DCT and GAT. However, IOP readings were consistently higher with DCT than with GAT (median difference: +1.7 mm Hg, interquartile range [25th-75th percentile] = 0.8-2.7 mm Hg). In contrast to GAT, multivariable regression analysis showed no significant effect of corneal thickness, corneal curvature, astigmatism, anterior chamber depth, and axial length on DCT readings. For repeated measurements the intraobserver variability was 0.65 mm Hg for the DCT and 1.1 mm Hg for the GAT (P = 0.008). Interobserver variability was 0.44 mm Hg for the DCT and 1.28 mm Hg for the GAT (P = 0.017).
CONCLUSIONS: IOP measurements by DCT are highly concordant with IOP readings obtained from GAT but do not vary in CCT and have a lower intra- and interobserver variability. DCT seems to be an appropriate method of tonometry for routine clinical use. Copyright Association for Research in Vision and Ophthalmology

Entities:  

Mesh:

Year:  2004        PMID: 15326129     DOI: 10.1167/iovs.04-0018

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  65 in total

1.  Cornea plana associated with open-angle glaucoma: a case report.

Authors:  Bilge Ozturk Sahin; Goktug Seymenoglu; Esin F Baser
Journal:  Int Ophthalmol       Date:  2011-12-11       Impact factor: 2.031

2.  [Influence of corneal thickness on intraocular pressure measurements following Descemet's stripping automated endothelial keratoplasty (DSAEK)].

Authors:  M K J Klamann; A-K B Maier; J Gonnermann; N Torun; P C Ruokonen
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

3.  Intraocular pressure and ocular pulse amplitude variations during the Valsalva maneuver.

Authors:  Umit Aykan; Mesut Erdurmus; Burhan Yilmaz; Ahmet Hamdi Bilge
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-24       Impact factor: 3.117

4.  Evaluation of ocular pulse amplitude and choroidal thickness in diabetic macular edema.

Authors:  Y Totan; T K Akyüz; E Güler; F B Güragaç
Journal:  Eye (Lond)       Date:  2015-11-13       Impact factor: 3.775

5.  Age-dependency of ocular parameters: a cross sectional study of young and elderly healthy subjects.

Authors:  G Jóhannesson; P Hallberg; K Ambarki; A Eklund; C Lindén
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-08-13       Impact factor: 3.117

6.  The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

Authors:  P-A Tonnu; T Ho; T Newson; A El Sheikh; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

7.  Ocular pulse amplitude after trabeculectomy.

Authors:  Sandra R von Schulthess; Claude Kaufmann; Lucas M Bachmann; Ahmet Yanar; Michael A Thiel
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-07-23       Impact factor: 3.117

Review 8.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

9.  Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties.

Authors:  A R Ismail; M Lamont; S Perera; D Khan-Lim; R Mehta; J D A Macleod; D F Anderson
Journal:  Br J Ophthalmol       Date:  2007-07       Impact factor: 4.638

10.  [Influence of residual corneal bed thickness after myopic LASIK on intraocular pressure measurements. Goldmann applanation tonometry and dynamic contour tonometry].

Authors:  L Muller; T Kohnen
Journal:  Ophthalmologe       Date:  2009-01       Impact factor: 1.059

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.