BACKGROUND: Goldmann applanation tonometry and dynamic contour tonometry (PASCAL) are two well established slit lamp mounted tonometric methods. Intraocular pressure measurement in bedridden patients and children is often only possible using hand held tonometers (TonoPenXL, Perkins tonometer, Draeger tonometer). This study was performed to evaluate the hand held ICare tonometer, which is based on the rebound method.
METHODS: A total of 102 eyes were examined by two highly experienced ophthalmologists for: 1) ophthalmological status, 2) central corneal power (Zeiss IOL-Master), 3) central corneal thickness (Tomey ultrasound pachymetry, five successive measurements, SD<5%), 4) intraocular pressure (IOP) measurement with the Goldmann applantation tonometer (GAT) 1x, 5) TonoPenXL (1x), 6) ICare with three successive measurement series of 6 single measurements.
RESULTS: The mean IOP(GAT) was 13.2+/-3.0 mmHg compared with the mean IOP(TonoPenXL) (13.4+/-3.1 mmHg) and with the IOP(ICare) (mean value of first measurement series: 13.4+/-3.1 mmHg). The series of measurements with the ICare showed a tonography effect (decrease of IOP from 14.6 mmHg at the first measurement and 14.2 mmHg at the second to 14.0 at the third measurement). The ICare-measurements were highly reliable (Cronbach's alpha=0.974) and showed a good correlation between the measurement series (r=0.592-0.642; p<0.001). There was a great intra-individual variability of up to 17 mmHg between the GAT, TonoPenXL and ICare methods.
CONCLUSIONS: The ICare tonometer is easy to handle and high reliability. The data are comparable with those from the Goldmann tonometer. A tonography effect of 0.6 mmHg in the successive measurement series was found.

Entities:  

Mesh:

Year:  2007        PMID: 17333073     DOI: 10.1007/s00347-007-1487-z

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  19 in total

1.  [Comparison of precision of the TonoPenXL with the Goldmann and Draeger applanation tonometer in a sitting and recumbent position of the patients -- a clinical study on 251 eyes].

Authors:  Arne Viestenz; Berthold Lausen; Anselm M Jünemann; Christian Y Mardin
Journal:  Klin Monbl Augenheilkd       Date:  2002-11       Impact factor: 0.700

2.  [Further contribution to applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1961-06       Impact factor: 3.250

3.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

4.  Tonometer utilization, accuracy, and calibration under field conditions.

Authors:  I F Wessels; Y Oh
Journal:  Arch Ophthalmol       Date:  1990-12

Review 5.  [Dynamic contour tonometry].

Authors:  C Kniestedt; H E Kanngiesser
Journal:  Ophthalmologe       Date:  2006-08       Impact factor: 1.059

6.  [Reproducibility of dynamic contour tonometry. Comparison with TonoPenXL and Goldmann applanation tonometry - a clinical study on 323 normal eyes].

Authors:  A Viestenz; A Langenbucher; A Viestenz
Journal:  Klin Monbl Augenheilkd       Date:  2006-10       Impact factor: 0.700

7.  Standardizing the measurement of intraocular pressure for clinical research. Guidelines from the Eye Care Technology Forum.

Authors:  M A Kass
Journal:  Ophthalmology       Date:  1996-01       Impact factor: 12.079

8.  [Clinical Testing of New Tonometers on the Basis of the International Standard for Human eye Tonometers ISO 8612].

Authors:  T Schwenteck; J Draeger; P Ruokonen; S Lechner; M Weise
Journal:  Klin Monbl Augenheilkd       Date:  2004-12       Impact factor: 0.700

Review 9.  Sources of error with use of Goldmann-type tonometers.

Authors:  M M Whitacre; R Stein
Journal:  Surv Ophthalmol       Date:  1993 Jul-Aug       Impact factor: 6.048

10.  Automatic tonometer with exact theory: various biological applications.

Authors:  R S MACKAY; E MARG; R OECHSLI
Journal:  Science       Date:  1960-06-03       Impact factor: 47.728

View more
  10 in total

1.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

2.  [Intra-ocular pressure during and after playing high and low resistance wind instruments].

Authors:  K Kappmeyer; I M Lanzl
Journal:  Ophthalmologe       Date:  2010-01       Impact factor: 1.059

Review 3.  [Measuring intraocular pressure by different methods].

Authors:  J Lamparter; E M Hoffmann
Journal:  Ophthalmologe       Date:  2009-08       Impact factor: 1.059

4.  [Comparison of the iCare tonometer with the Goldmann tonometer in Malawi].

Authors:  J Hohmann; M Schulze-Schwering; T Chirambo Nyaka; V Moyo; P C Kayange; D Doycheva; N H Batumba; M S Spitzer
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

5.  Accommodation-induced intraocular pressure changes in progressing myopes and emmetropes.

Authors:  L Yan; L Huibin; L Xuemin
Journal:  Eye (Lond)       Date:  2014-09-05       Impact factor: 3.775

6.  Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons.

Authors:  Saulius Galgauskas; Rasa Strupaite; Ernesta Strelkauskaite; Rimvydas Asoklis
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

7.  COMPARISON OF IOP READINGS USING REBOUND I CARE TONOMETER AND PERKINS APPLANATION TONOMETER IN AN AFRICAN POPULATION.

Authors:  D S Ademola-Popoola; A F Odi; T M Akande
Journal:  J West Afr Coll Surg       Date:  2014 Jan-Mar

8.  The utility of rebound tonometer in late elderly subjects.

Authors:  Rei Sakata; Jiro Numaga
Journal:  Clin Interv Aging       Date:  2014-01-22       Impact factor: 4.458

9.  Intraocular Pressure Changes during Accommodation in Progressing Myopes, Stable Myopes and Emmetropes.

Authors:  Yan Liu; Huibin Lv; Xiaodan Jiang; Xiaodan Hu; Mingzhou Zhang; Xuemin Li
Journal:  PLoS One       Date:  2015-10-30       Impact factor: 3.240

10.  The Effect of Inferior Rectus Muscle Thickening on Intraocular Pressure in Thyroid-Associated Ophthalmopathy.

Authors:  Xiuhong Li; Xu Bai; Zhibin Liu; Ming Cheng; Jia Li; Nian Tan; Hongfeng Yuan
Journal:  J Ophthalmol       Date:  2021-01-29       Impact factor: 1.909