"Motion-resistant" pulse oximetry: a comparison of new and old models.

Steven J Barker1.   

Abstract

Several pulse oximeter manufacturers have recently developed instruments that are claimed to be resistant to the effects of patient motion. We performed a laboratory volunteer experiment to compare the performances of several of these instruments, as well as some older models, during combinations of motion and hypoxemia. Twenty oximeters were studied. A motorized table produced different hand motions, and each motion was studied during both room air breathing and hypoxemia. Pulse oximeters on the nonmoving hand were used to provide control measurements for comparison. The Masimo SET((R)) pulse oximeter exhibited the best overall performance, with a performance index (percentage of time in which the SpO(2) reading is within 7% of control value) of 94%. The Agilent Viridia 24C was next, with an 84% index, followed by the Agilent CMS (80%), the Datex-Ohmeda 3740 (80%), and the Nellcor N-395 (69%). For comparison with older oximeter technology, the Criticare 5040 had an index of 28%. Recent technology changes have significantly improved pulse oximeter performance during motion artifact, with the Masimo oximeter leading the way. IMPLICATIONS. New improvements in pulse oximeter technology have resulted in significantly better accuracy and reliability during patient motion. The Masimo pulse oximeter demonstrated the best performance of the 20 instruments tested.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12351278     DOI: 10.1097/00000539-200210000-00033

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  22 in total

1.  Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

2.  Analysis of the ear pulse oximeter waveform.

Authors:  Aymen A Awad; Robert G Stout; M Ashraf M Ghobashy; Hoda A Rezkanna; David G Silverman; Kirk H Shelley
Journal:  J Clin Monit Comput       Date:  2006-04-13       Impact factor: 2.502

Review 3.  A review of signals used in sleep analysis.

Authors:  A Roebuck; V Monasterio; E Gederi; M Osipov; J Behar; A Malhotra; T Penzel; G D Clifford
Journal:  Physiol Meas       Date:  2013-12-17       Impact factor: 2.833

Review 4.  Wireless and battery-free platforms for collection of biosignals.

Authors:  Tucker Stuart; Le Cai; Alex Burton; Philipp Gutruf
Journal:  Biosens Bioelectron       Date:  2021-01-23       Impact factor: 10.618

5.  Oxygen saturation response to exercise in healthy pregnant women: a simple protocol and normal range.

Authors:  Edward Langford; Ahmad Khwanda; Kate Langford
Journal:  Obstet Med       Date:  2010-06-03

6.  The incidence of hypoxemia during surgery: evidence from two institutions.

Authors:  Jesse M Ehrenfeld; Luke M Funk; Johan Van Schalkwyk; Alan F Merry; Warren S Sandberg; Atul Gawande
Journal:  Can J Anaesth       Date:  2010-07-31       Impact factor: 5.063

7.  Sleep disordered breathing does not predict acute severe pain episodes in children with sickle cell anemia.

Authors:  Shaina M Willen; Mark Rodeghier; Carol L Rosen; Michael R DeBaun
Journal:  Am J Hematol       Date:  2018-01-25       Impact factor: 10.047

8.  Temporal quantification of oxygen saturation ranges: an effort to reduce hyperoxia in the neonatal intensive care unit.

Authors:  M J Bizzarro; F Y Li; K Katz; V Shabanova; R A Ehrenkranz; V Bhandari
Journal:  J Perinatol       Date:  2013-09-26       Impact factor: 2.521

9.  The reliability of a single pulse oximetry reading as a screening test for congenital heart disease in otherwise asymptomatic newborn infants.

Authors:  J D Reich; B Connolly; G Bradley; S Littman; W Koeppel; P Lewycky; M Liske
Journal:  Pediatr Cardiol       Date:  2008-09       Impact factor: 1.655

10.  [Reliability of home monitoring with event-recording compared with polysomnography in infants].

Authors:  Sabine de Nardi; Ekkehart Paditz; Thomas Erler; Andreas Gruntzke
Journal:  Wien Klin Wochenschr       Date:  2003-07-15       Impact factor: 1.704

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.