Transesophageal versus surface pulse oximetry in intensive care unit patients.

M N Vicenzi1, H Gombotz, H Krenn, C Dorn, P Rehak.   

Abstract

OBJECTIVES: To compare oximetric readings from the esophagus (STEO2) and the skin (finger, SSO2) with those obtained from arterial blood samples (SaO). In addition, to compare the influences of mean arterial pressure (MAP) and body temperature to the accuracy of STEO2 and SSO2 readings.
DESIGN: Prospective, single-center study.
SETTING: Surgical intensive care unit of an academic, teaching, and community hospital in Austria. PATIENTS: A total of 40 consecutive, severely traumatized or diseased, intensive care unit patients requiring mechanical ventilatory support and deep analgosedation. Patients had to be nonpregnant, > or =19 yrs of age, and without a disease or a trauma of the esophagus.
INTERVENTIONS: Placement of an esophageal and a finger-pulse oximetry probe and a radial artery catheter.
MEASUREMENTS AND MAIN RESULTS: STEO2, SSO2, MAP, and esophageal temperature were recorded continuously during a 4-hr period, and SaO2 was measured every 30 mins. The first outcome variable was the deviation of STEO2 and SSO2 from SaO2. The second outcome variable was the influence of MAP and body temperature on STEO2 and SSO2 regression analysis and repeated measures. Analysis of variance was used for statistics (p < .05 was accepted as significant). In patients with a MAP ranging from 29 to 111 mm Hg and a temperature ranging from 33.4 degrees C (92.1 degrees to 39.2 degrees C (102.6 degrees F), SSO2 measurements underestimate SaO2 by 2% to 4%. Whereas STEO2 matches SaO2, STEO2 was not dependent on MAP or temperature, but increased temperature or low MAP were associated with falsely low SSO2 readings.
CONCLUSIONS: Assuming correct positioning of the probe, readings from the esophagus are more consistent with arterial oxygen saturation than readings from surface pulse oximetry. MAP or temperature changes do not influence STEO2, but they do affect SSO2. In critically ill patients, STEO2 appears to be a more reliable variable than SSO2.

Entities:  

Mesh:

Year:  2000        PMID: 10921551     DOI: 10.1097/00003246-200007000-00014

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  8 in total

1.  Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

2.  Can the descending aortic stroke volume be estimated by transesophageal descending aortic photoplethysmography?

Authors:  Peng Ling; Gong Quan; Yu Siyuan; Gao Bo; Wei Wei
Journal:  J Anesth       Date:  2017-03-27       Impact factor: 2.078

3.  Esophageal pulse oximetry is more accurate and detects hypoxemia earlier than conventional pulse oximetry during general anesthesia.

Authors:  Guo Chen; Zhaoqiong Zhu; Jin Liu; Wei Wei
Journal:  Front Med       Date:  2012-10-02       Impact factor: 4.592

4.  A comparison of response time to desaturation between tracheal oximetry and peripheral oximetry.

Authors:  Li Wang; Wei Wei; Ming Gong; Ling Mu
Journal:  J Clin Monit Comput       Date:  2010-03-12       Impact factor: 2.502

5.  Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2004-01-13       Impact factor: 17.440

6.  Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing.

Authors:  Mona Ascha; Anirban Bhattacharyya; Jose A Ramos; Adriano R Tonelli
Journal:  Med Sci Sports Exerc       Date:  2018-10       Impact factor: 5.411

7.  A preliminary study on the monitoring of mixed venous oxygen saturation through the left main bronchus.

Authors:  Xiang-rui Wang; Yong-jun Zheng; Jie Tian; Zheng-hong Wang; Zhi-ying Pan
Journal:  Crit Care       Date:  2006-02       Impact factor: 9.097

8.  Impact of COVID-19 on the association between pulse oximetry and arterial oxygenation in patients with acute respiratory distress syndrome.

Authors:  Lee S Nguyen; Marion Helias; Lisa Raia; Estelle Nicolas; Paul Jaubert; Sarah Benghanem; Zakaria Ait Hamou; Pierre Dupland; Julien Charpentier; Frédéric Pène; Alain Cariou; Jean-Paul Mira; Jean-Daniel Chiche; Mathieu Jozwiak
Journal:  Sci Rep       Date:  2022-01-27       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.