Literature DB >> 23991388

Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Mohammad Soleimani1, Hassan Hashemi, Shiva Mehravaran, Mehdi Khabazkhoob, Mohammad Hassan Emamian, Mohammad Shariati, Akbar Fotouhi.   

Abstract

AIM: To compare central corneal thickness (CCT) and anterior chamber depth (ACD) measurements using rotating Scheimpflug imaging and partial coherence interferometry.
METHODS: As part of the first phase of Shahroud Eye Cohort Study with 5 190 subjects of 40 to 64 years of age, CCT and ACD were measured using Scheimpflug imaging with the Pentacam (Oculus, Inc., Lynnwood, WA, USA) and partial coherence interferometry with the Allegro BioGraph (Wavelight, Erlangen, Germany).
RESULTS: After applying exclusion criteria, we had data of 4 387 subjects with a mean age of 50.7±6.2 years. Mean CCT with Pentacam and BioGraph were 528.6±33.2µm and 525.6±32µm respectively; the difference was statistically significant (P<0.001), but the correlation was high (R=0.920). Mean ACD measurements using Pentacam and BioGraph were 2.68±0.35mm and 2.62±0.33mm respectively; the inter-device difference was significant (P<0.001) with high correlation (R=0.944). The 95% limits of agreements between devices were -22.65µm to 28.61µm and -0.16mm to 0.29mm for CCT and ACD measurements, respectively.
CONCLUSION: For both CCT and ACD, the BioGraph gave significantly lower values than the Pentacam (P<0.05). Despite the high inter-device correlation, the 95% limits of agreements were wide, and this may limit their interchangeability in measuring the CCT and ACD.

Entities:  

Keywords:  BioGraph; Pentacam; agreement; anterior chamber depth; central corneal thickness

Year:  2013        PMID: 23991388      PMCID: PMC3755313          DOI: 10.3980/j.issn.2222-3959.2013.04.19

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  34 in total

1.  Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system.

Authors:  Georg Rainer; Oliver Findl; Vanessa Petternel; Barbara Kiss; Wolfgang Drexler; Christian Skorpik; Michael Georgopoulos; Leopold Schmetterer
Journal:  Ophthalmology       Date:  2004-05       Impact factor: 12.079

2.  Reliability of the corneal thickness measurements with the Pentacam HR imaging system and ultrasound pachymetry.

Authors:  László Módis; Eszter Szalai; Gábor Németh; András Berta
Journal:  Cornea       Date:  2011-05       Impact factor: 2.651

3.  Anterior chamber depth measurement with a-scan ultrasonography, Orbscan II, and IOLMaster.

Authors:  Hassan Hashemi; Kamran Yazdani; Shiva Mehravaran; Akbar Fotouhi
Journal:  Optom Vis Sci       Date:  2005-10       Impact factor: 1.973

4.  Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan.

Authors:  Birgit Lackner; Gerald Schmidinger; Christian Skorpik
Journal:  Optom Vis Sci       Date:  2005-09       Impact factor: 1.973

5.  Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  J Cataract Refract Surg       Date:  2007-10       Impact factor: 3.351

6.  Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements.

Authors:  Hema Shankar; Deepa Taranath; Chandramalar T Santhirathelagan; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2008-01       Impact factor: 3.351

7.  Accuracy of a new partial coherence interferometry analyser for biometric measurements.

Authors:  M P Holzer; M Mamusa; G U Auffarth
Journal:  Br J Ophthalmol       Date:  2009-03-15       Impact factor: 4.638

8.  Collagen crosslinking with riboflavin and ultraviolet-A light in keratoconus: long-term results.

Authors:  Frederik Raiskup-Wolf; Anne Hoyer; Eberhard Spoerl; Lutz E Pillunat
Journal:  J Cataract Refract Surg       Date:  2008-05       Impact factor: 3.351

9.  Comparison of three methods of measuring corneal thickness and anterior chamber depth.

Authors:  Wolf Buehl; Danijela Stojanac; Stefan Sacu; Wolfgang Drexler; Oliver Findl
Journal:  Am J Ophthalmol       Date:  2006-01       Impact factor: 5.258

10.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

View more
  1 in total

1.  Corneal topography analysis of stromal corneal dystrophies.

Authors:  Yusuf Kocluk; Zuleyha Yalniz-Akkaya; Ayse Burcu; Firdevs Ornek
Journal:  Pak J Med Sci       Date:  2015 Jan-Feb       Impact factor: 1.088

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.