Anterior chamber depth measured by two methods in myopic and hyperopic phakic IOL implant.

M Vetrugno1, N Cardascia, L Cardia.   

Abstract

AIM: To evaluate the accuracy and reliability of the optical versus ultrasonic measurement of anterior chamber depth (ACD) in a group of patients affected by high myopia or hyperopia, before phakic IOL implant.
METHODS: 18 consecutive patients (34 eyes) were enrolled in this study, and asked to undergo phakic IOL implant to correct high myopia or hyperopia. The mean age was 29.5 (SD 3.4) years, the male/female ratio was 10/8. 13 patients (24 eyes) were myopic (mean myopia -16.17 (4.39) D, range -9 to -24 D), whereas five patients (10 eyes) were hyperopic (mean hyperopia 7.4 (2.01) D, range 5.5-11 D). For each patient, ACD was evaluated comparing an optical system (Orbscan topograph system) with a standard A-scan ultrasound system. To evaluate the reliability of the two methods, the average of three optical and 10 ultrasonic consecutive measurements were considered. Statistical analysis was performed by means linear regression.
RESULTS: The average difference between optical and ultrasound values was 0.17 (0.1) mm (4.68% (2.52%)). There was a constant underestimation of the ACD with the optical system compared with the ultrasound measurement, more evident in the hyperopic eyes (5.20% (1. 95%)) than in myopic ones (4.46% (2.72%)). The reliability of measurements, as showed by the standard deviation values, was higher in the optical system (0.03 (0.02)) than in the ultrasound (0.1 (0. 05)), with a statistical difference (p<0.001). The linear regression analysis between optical and ultrasound measurements was 0.8992 (p<0. 0001).
CONCLUSIONS: The optical measurement of ACD is an accurate and reliable technique in high myopic and hyperopic eyes. It gives a small underestimation of the ACD values, but it could be preferable to the ultrasound technique, because it demonstrates more repeatability and has the advantages of a non-contact technique.

Entities:  

Mesh:

Year:  2000        PMID: 11004094      PMCID: PMC1723277          DOI: 10.1136/bjo.84.10.1113

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  32 in total

1.  Iris-claw lens in phakic eyes to correct hyperopia: preliminary study.

Authors:  P U Fechner; D Singh; K Wulff
Journal:  J Cataract Refract Surg       Date:  1998-01       Impact factor: 3.351

2.  [Comparison of simple methods to screen predisposing eye of primary angle-closure glaucoma].

Authors:  T Ye; W Mao; D Lu
Journal:  Zhonghua Yan Ke Za Zhi       Date:  1995-09

3.  Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry.

Authors:  V Yaylali; S C Kaufman; H W Thompson
Journal:  J Cataract Refract Surg       Date:  1997-11       Impact factor: 3.351

4.  Measuring anterior chamber depth with Orbscan Topography System.

Authors:  G U Auffarth; M R Tetz; Y Biazid; H E Völcker
Journal:  J Cataract Refract Surg       Date:  1997-11       Impact factor: 3.351

5.  Ocular components measured by keratometry, phakometry, and ultrasonography in emmetropic and myopic optometry students.

Authors:  D A Goss; H G Van Veen; B B Rainey; B Feng
Journal:  Optom Vis Sci       Date:  1997-07       Impact factor: 1.973

6.  Biometry and primary angle-closure glaucoma among Chinese, white, and black populations.

Authors:  N G Congdon; Q Youlin; H Quigley; P T Hung; T H Wang; T C Ho; J M Tielsch
Journal:  Ophthalmology       Date:  1997-09       Impact factor: 12.079

7.  Comparison of corneal versus through-the-lid A-scan ultrasound biometry.

Authors:  J D Twelker; S Kirschbaum; K Zadnik; D O Mutti
Journal:  Optom Vis Sci       Date:  1997-10       Impact factor: 1.973

8.  Effect of decreasing surface and interface irregularities after photorefractive keratectomy and laser in situ keratomileusis on optical and functional outcomes.

Authors:  P Vinciguerra; M Azzolini; P Airaghi; P Radice; V De Molfetta
Journal:  J Refract Surg       Date:  1998-04       Impact factor: 3.573

9.  Posterior chamber phakic intraocular lens for hyperopia.

Authors:  D R Sanders; R G Martin; D C Brown; J Shepherd; M R Deitz; M DeLuca
Journal:  J Refract Surg       Date:  1999 May-Jun       Impact factor: 3.573

10.  Ultrasound biomicroscopy of Baikoff anterior chamber phakic intraocular lenses.

Authors:  J J Saragoussi; M Puech; M Assouline; O Berges; G Renard; Y J Pouliquen
Journal:  J Refract Surg       Date:  1997 Mar-Apr       Impact factor: 3.573

View more
  5 in total

1.  Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode.

Authors:  B Meinhardt; O Stachs; J Stave; R Beck; R Guthoff
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-09-15       Impact factor: 3.117

2.  Comparison of preoperative and postoperative ocular biometry in eyes with phakic intraocular lens implantations.

Authors:  Joo Youn Shin; Jae Bum Lee; Kyoung Yul Seo; Eung Kweon Kim; Tae-im Kim
Journal:  Yonsei Med J       Date:  2013-09       Impact factor: 2.759

3.  Measurement of Angle Kappa Using Ultrasound Biomicroscopy and Corneal Topography.

Authors:  Joon Hyung Yeo; Nam Ju Moon; Jeong Kyu Lee
Journal:  Korean J Ophthalmol       Date:  2017-04-24

4.  Anterior chamber dimensions, angles and pupil diameter in patients with Down syndrome: A comparative population-based study.

Authors:  Soheila Asgari; Hassan Hashemi; Akbar Fotouhi; Shiva Mehravaran
Journal:  Indian J Ophthalmol       Date:  2020-05       Impact factor: 1.848

5.  Study of distribution of white-to-white corneal diameter and anterior chamber depth in study population obtained with optical biometry using intraocular lens (IOL) master.

Authors:  Kritika Singh; Sandeep Gupta; P S Moulick; Neeraj Bhargava; Alok Sati; Gagandeep Kaur
Journal:  Med J Armed Forces India       Date:  2018-08-14
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.