Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.

H T Ho1, W K Y Yeung, B W Y Young.   

Abstract

BACKGROUND: Low blood glucose in newborns is difficult to detect clinically. Hence a reliable "point of care" device (glucometer) for early detection and treatment of low glucose is needed.
OBJECTIVE: To evaluate the performance of five readily available glucometers for the detection of low blood glucose in newborn infants.
METHOD: Glucostix measurements were taken for newborns with risk factors using a Reflolux S (Boehringer) glucometer. If the initial reading was low (< 2.6 mmol/l), further measurements were taken with two other glucometers (phase I, Advantage and Glucotrend (Roche); phase II, Elite XL (Bayer) and Precision (Abbott)), and plasma glucose was measured in the laboratory (Aeroset; Abbott).
RESULTS: Over 10 months, 101 specimens were collected from 71 newborns (57 in phase I; 44 in phase II). The Advantage glucometer usually overestimated blood glucose with a mean difference of 1.07 mmol/l (p < 0.01) at all low glucose ranges. The Glucotrend, Precision, and Elite XL glucometers performed better; the mean differences were not significantly different from the laboratory measured value (0.17 mmol/l (p = 0.37); -0.12 mmol/l (p = 0.13), and 0.24 mmol/l (p = 0.13) respectively). For detection of glucose concentrations < 2.6 mmol/l, the Precision glucometer had the highest sensitivity (96.4%) and negative predictive value (90%). For lower glucose concentrations (< 2.0 mmol/l), the Glucotrend glucometer performed even better (sensitivity 92.3%, negative predictive value 96.3%).
CONCLUSION: Point of care devices should have good precision in the low glucose concentration range, sensitivity, and accuracy for early detection of neonatal hypoglycaemia. None of the five glucometers was satisfactory as the sole measuring device. The Glucotrend and Precision glucometers have the greatest sensitivity and negative predictive value. However, confirmation with laboratory measurements of plasma glucose and clinical assessment are still of the utmost importance.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15210675      PMCID: PMC1721720          DOI: 10.1136/adc.2003.033548

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  18 in total

1.  Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates.

Authors:  J Girouard; J C Forest; J Massé; M Leroux; N C Bradburn; T C Noblet; J O Joynes; J Baum
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  Comparison of EML 105 and advantage analysers measuring capillary versus venous whole blood glucose in neonates.

Authors:  P J McNamara; N Sharief
Journal:  Acta Paediatr       Date:  2001-09       Impact factor: 2.299

3.  Comparison of two strip test methods of whole blood glucose measurement in the neonatal period.

Authors:  M Papp; N Sharief
Journal:  Acta Paediatr       Date:  2001-09       Impact factor: 2.299

4.  Adverse neurodevelopmental outcome of moderate neonatal hypoglycaemia.

Authors:  A Lucas; R Morley; T J Cole
Journal:  BMJ       Date:  1988-11-19

5.  Neural dysfunction during hypoglycaemia.

Authors:  T H Koh; A Aynsley-Green; M Tarbit; J A Eyre
Journal:  Arch Dis Child       Date:  1988-11       Impact factor: 3.791

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Transcutaneous glucose measurement using near-infrared spectroscopy during hypoglycemia.

Authors:  I Gabriely; R Wozniak; M Mevorach; J Kaplan; Y Aharon; H Shamoon
Journal:  Diabetes Care       Date:  1999-12       Impact factor: 19.112

8.  Glucose monitoring with long-term subcutaneous microdialysis in neonates.

Authors:  F A Baumeister; B Rolinski; R Busch; P Emmrich
Journal:  Pediatrics       Date:  2001-11       Impact factor: 7.124

9.  Long-term effects of neonatal hypoglycemia on brain growth and psychomotor development in small-for-gestational-age preterm infants.

Authors:  C B Duvanel; C L Fawer; J Cotting; P Hohlfeld; J M Matthieu
Journal:  J Pediatr       Date:  1999-04       Impact factor: 4.406

Review 10.  Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds.

Authors:  M Cornblath; J M Hawdon; A F Williams; A Aynsley-Green; M P Ward-Platt; R Schwartz; S C Kalhan
Journal:  Pediatrics       Date:  2000-05       Impact factor: 7.124

View more
  14 in total

1.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

2.  Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycaemia.

Authors:  Andreas Michel; Helmut Küster; Alexander Krebs; Ingrid Kadow; Wilfried Paul; Matthias Nauck; Christoph Fusch
Journal:  Eur J Pediatr       Date:  2005-07-22       Impact factor: 3.183

Review 3.  The screening and management of newborns at risk for low blood glucose.

Authors:  Michael R Narvey; Seth D Marks
Journal:  Paediatr Child Health       Date:  2019-12-09       Impact factor: 2.253

Review 4.  Challenges to glycemic measurement in the perioperative and critically ill patient: a review.

Authors:  Andrew D Pitkin; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

5.  The Accutrend sensor glucose analyzer may not be adequate in bedside testing for neonatal hypoglycemia.

Authors:  Mark Rosenthal; Bernhard Ugele; Gerd Lipowsky; Helmut Küster
Journal:  Eur J Pediatr       Date:  2005-10-19       Impact factor: 3.183

6.  Comparison of Point of Care (POC) Testing of Glucose by B Braun Glucometer and Hemocue Glucose 201+ Analyser Versus Centralised Testing in Neonatal Intensive Care Unit (NICU).

Authors:  Sudha Reddy V R; Sumathi M E; Beere Gowda Y C; Mohamed Suhail S
Journal:  J Clin Diagn Res       Date:  2014-07-20

7.  Continuous glucose monitoring in newborn infants: how do errors in calibration measurements affect detected hypoglycemia?

Authors:  Felicity Thomas; Mathew Signal; Deborah L Harris; Philip J Weston; Jane E Harding; Geoffrey M Shaw; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2014-02-27

8.  ABM clinical protocol #1: guidelines for blood glucose monitoring and treatment of hypoglycemia in term and late-preterm neonates, revised 2014.

Authors:  Nancy Wight; Kathleen A Marinelli
Journal:  Breastfeed Med       Date:  2014-05       Impact factor: 1.817

9.  Cost Analysis of Cot-Side Screening Methods for Neonatal Hypoglycaemia.

Authors:  Matthew J Glasgow; Jane E Harding; Richard Edlin
Journal:  Neonatology       Date:  2018-06-12       Impact factor: 4.035

10.  Accuracy of the StatStrip versus SureStep Flexx glucose meter in neonates at risk of hypoglycemia.

Authors:  Ratchada Kitsommart; Sopapan Ngerncham; Pimol Wongsiridej; Tharatip Kolatat; Kriang-Sak Jirapaet; Bosco Paes
Journal:  Eur J Pediatr       Date:  2013-05-05       Impact factor: 3.183

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.