Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates.

J Girouard1, J C Forest, J Massé, M Leroux, N C Bradburn, T C Noblet, J O Joynes, J Baum.   

Abstract

OBJECTIVE: To evaluate the clinical performance of the Glucometer Elite XL Diabetes Care System in neonatal settings using a multicenter study RESEARCH DESIGN AND METHODS: A total of 388 blood specimens from 333 neonates were included in the study. A capillary or arterial sample was analyzed for determination of glucose with the Glucometer Elite XL system by an attending trained nurse. Through the same sampling site, a specimen was collected and sent to the laboratory for measurement of plasma glucose, bilirubin, and hematocrit.
RESULTS: The regression analysis between the results of the Glucometer Elite XL system and comparative methods resulted in the following: Glucometer Elite XL meter = 1.01 x laboratory method + 0.02 mmol/l (n = 388). For the 1.1-4.0 mmol/l plasma glucose range, the regression was Glucometer Elite XL meter = 1.07 x laboratory method + 0.12 mmol/l (n = 150). A difference plot indicated a mean bias of 0.04 mmol/l (95% CI -0.01 to 0.10). No relationship was found between meter glucose biases and hematocrit levels (r = 0.10, P = 0.14). Although a statistically significant correlation existed between bilirubin levels and the glucose meter biases (r = 0.14, P = 0.005), the predicted mean biases were of little clinical significance.
CONCLUSIONS: The Glucometer Elite XL system showed a good performance when used in neonatal settings.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10937513     DOI: 10.2337/diacare.23.8.1149

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  7 in total

1.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

2.  Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycaemia.

Authors:  Andreas Michel; Helmut Küster; Alexander Krebs; Ingrid Kadow; Wilfried Paul; Matthias Nauck; Christoph Fusch
Journal:  Eur J Pediatr       Date:  2005-07-22       Impact factor: 3.183

3.  Accuracy and Precision of Measured Blood Sugar Values by Three Glucometers Compared to the Standard Technique.

Authors:  Shadi Khakpour Kermani; Alireza Khatony; Rostam Jalali; Mansur Rezaei; Alireza Abdi
Journal:  J Clin Diagn Res       Date:  2017-04-01

4.  The Accutrend sensor glucose analyzer may not be adequate in bedside testing for neonatal hypoglycemia.

Authors:  Mark Rosenthal; Bernhard Ugele; Gerd Lipowsky; Helmut Küster
Journal:  Eur J Pediatr       Date:  2005-10-19       Impact factor: 3.183

5.  Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.

Authors:  H T Ho; W K Y Yeung; B W Y Young
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-07       Impact factor: 5.747

6.  Accuracy of the StatStrip versus SureStep Flexx glucose meter in neonates at risk of hypoglycemia.

Authors:  Ratchada Kitsommart; Sopapan Ngerncham; Pimol Wongsiridej; Tharatip Kolatat; Kriang-Sak Jirapaet; Bosco Paes
Journal:  Eur J Pediatr       Date:  2013-05-05       Impact factor: 3.183

7.  Screening for hypoglycemia at the bedside in the neonatal intensive care unit (NICU) with the Abbott PCx glucose meter.

Authors:  Cynthia Balion; Vijaylaxmi Grey; Afisi Ismaila; Susan Blatz; Wendy Seidlitz
Journal:  BMC Pediatr       Date:  2006-11-03       Impact factor: 2.125

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.