Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycaemia.

Andreas Michel1, Helmut Küster, Alexander Krebs, Ingrid Kadow, Wilfried Paul, Matthias Nauck, Christoph Fusch.   

Abstract

UNLABELLED: To prevent persistent neurodevelopment and physical growth deficits in neonatal care, it is mandatory to determine blood glucose levels as quickly and precisely as possible, preferably using micro-methods. However, most commercially available instruments have not been validated and approved for this purpose. The aim of this study was to validate the Glucometer Elite XL, a newly developed device for point-of-care testing (POCT). In samples from 869 newborn infants, glucose levels were simultaneously measured by the Glucometer Elite XL in whole blood and by an accepted clinical laboratory method in haemolysed blood using the ECA 2000 device. An acceptable method agreement was found between the POCT and the ECA 2000 method (mean difference 0.013 mmol/l, SD 0.69). As determined by regression analysis (Passing-Bablok), the slope was 1.086 with a y-intercept of -0.4 mmol/l ( r =0.959, P <0.05). The differences between measurement pairs of both assays versus the haematocrit were negligible. With a cut-off for hypoglycaemia at 2.6 mmol/l glucose in haemolysed blood, the sensitivity of the POCT device was 0.63 and specificity was 0.98. Raising the cut-off of the Glucometer Elite XL to 3.2 mmol/l, the sensitivity and specificity incremented to 1.0 and 0.89, respectively.
CONCLUSION: The Glucometer Elite XL instrument can be recommended for point-of-care blood glucose measurement in newborn infants if its character as a screening method is taken into account. To compensate deviating results, we advise to shift its cut-off for hypoglycaemia recognition to a safe threshold of 3.2 mmol/l. However, hypoglycaemia has to be confirmed by a valid glucose measurement in the clinical laboratory.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16041524     DOI: 10.1007/s00431-005-1733-9

Source DB:  PubMed          Journal:  Eur J Pediatr        ISSN: 0340-6199            Impact factor:   3.183


  17 in total

1.  Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates.

Authors:  J Girouard; J C Forest; J Massé; M Leroux; N C Bradburn; T C Noblet; J O Joynes; J Baum
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  A comparison of chromogen test strip (Chemstrip bG) and serum glucose values in newborns.

Authors:  P C Holtrop; K A Madison; F L Kiechle; R E Karcher; D G Batton
Journal:  Am J Dis Child       Date:  1990-02

3.  Adverse neurodevelopmental outcome of moderate neonatal hypoglycaemia.

Authors:  A Lucas; R Morley; T J Cole
Journal:  BMJ       Date:  1988-11-19

4.  Accuracy of glucose reflectance testing for detecting hypoglycemia in term newborns.

Authors:  L Meloy; G Miller; M H Chandrasekaran; C Summitt; G Gutcher
Journal:  Clin Pediatr (Phila)       Date:  1999-12       Impact factor: 1.168

Review 5.  The definition of hypoglycaemia.

Authors:  J W Gregory; A Aynsley-Green
Journal:  Baillieres Clin Endocrinol Metab       Date:  1993-07

Review 6.  Prevention and management of neonatal hypoglycaemia.

Authors:  A Mehta
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1994-01       Impact factor: 5.747

Review 7.  Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds.

Authors:  M Cornblath; J M Hawdon; A F Williams; A Aynsley-Green; M P Ward-Platt; R Schwartz; S C Kalhan
Journal:  Pediatrics       Date:  2000-05       Impact factor: 7.124

8.  Determining the best method for first-line assessment of neonatal blood glucose levels.

Authors:  C L Thomas; L Critchley; M W Davies
Journal:  J Paediatr Child Health       Date:  2000-08       Impact factor: 1.954

Review 9.  Neonatal glucose metabolism: differential diagnoses, evaluation, and treatment of hypoglycemia.

Authors:  Richard M Cowett; Jeffrey L Loughead
Journal:  Neonatal Netw       Date:  2002-06

10.  Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.

Authors:  H T Ho; W K Y Yeung; B W Y Young
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-07       Impact factor: 5.747

View more
  3 in total

Review 1.  Hypoglycemia in critically ill children.

Authors:  E Vincent S Faustino; Eliotte L Hirshberg; Clifford W Bogue
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

Review 3.  Continuous glucose monitoring in neonates: a review.

Authors:  Christopher J D McKinlay; J Geoffrey Chase; Jennifer Dickson; Deborah L Harris; Jane M Alsweiler; Jane E Harding
Journal:  Matern Health Neonatol Perinatol       Date:  2017-10-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.