Literature DB >> 16235054

The Accutrend sensor glucose analyzer may not be adequate in bedside testing for neonatal hypoglycemia.

Mark Rosenthal1, Bernhard Ugele, Gerd Lipowsky, Helmut Küster.   

Abstract

UNLABELLED: The aim of this prospective observational study was to compare a bedside test with the reference laboratory method in routine postnatal glucose monitoring. Term newborns with increased risk or clinical signs of hypoglycemia were screened with a bedside test. In case of a glucose value below 2.25 mmol/L, a second blood sample was taken and a duplicate glucose measurement done in the laboratory using a bedside test (Accutrend sensor) and the reference laboratory method (hexokinase method) at the same time and from the same sample. From 110 term newborns, 122 blood samples were obtained for duplicate measurements (median 1.69 mmol/L, SD 0.45 mmol/L). Of these 122, Accutrend correctly identified 97% as being <2.25 mmol/L by the laboratory method. A Bland-Altman plot revealed a mean underestimation of the Accutrend of only -0.09 mmol/L. However, due to high scattering, the maximal over- and underestimation was 0.89 and 1.39 mmol/L, respectively. Only 75% of the results from the Accutrend were within +/-20% of the result of the laboratory method. If the cut-off for low glucose concentrations was set 0.6 mmol/L higher for the bedside test as compared to the laboratory method, all patients except one would have been correctly identified as hypoglycemic.
CONCLUSION: When using the Accutrend sensor, single infants with even marked hypoglycemia might be missed. Some delay in receiving accurate measurements might be more helpful for clinical decisions and long-term outcome than immediate but potentially misleading results.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16235054     DOI: 10.1007/s00431-005-0013-z

Source DB:  PubMed          Journal:  Eur J Pediatr        ISSN: 0340-6199            Impact factor:   3.183


  16 in total

1.  Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates.

Authors:  J Girouard; J C Forest; J Massé; M Leroux; N C Bradburn; T C Noblet; J O Joynes; J Baum
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  The Hemocue B glucose analyser and neonatal blood glucose monitoring.

Authors:  K Raile; G Simbruner
Journal:  Ann Clin Biochem       Date:  1998-03       Impact factor: 2.057

3.  The use of a Hemocue blood glucose analyser in a neonatal unit.

Authors:  M Leonard; M Chessall; D Manning
Journal:  Ann Clin Biochem       Date:  1997-05       Impact factor: 2.057

4.  Adverse neurodevelopmental outcome of moderate neonatal hypoglycaemia.

Authors:  A Lucas; R Morley; T J Cole
Journal:  BMJ       Date:  1988-11-19

5.  Evaluation of Glucocard Memory 2 and Accutrend sensor blood glucose meters.

Authors:  S Arens; V Moons; P Meuleman; F Struyf; Z Zaman
Journal:  Clin Chem Lab Med       Date:  1998-01       Impact factor: 3.694

Review 6.  The definition of hypoglycaemia.

Authors:  J W Gregory; A Aynsley-Green
Journal:  Baillieres Clin Endocrinol Metab       Date:  1993-07

Review 7.  Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds.

Authors:  M Cornblath; J M Hawdon; A F Williams; A Aynsley-Green; M P Ward-Platt; R Schwartz; S C Kalhan
Journal:  Pediatrics       Date:  2000-05       Impact factor: 7.124

Review 8.  Neonatal glucose metabolism: differential diagnoses, evaluation, and treatment of hypoglycemia.

Authors:  Richard M Cowett; Jeffrey L Loughead
Journal:  Neonatal Netw       Date:  2002-06

9.  Clinical utility and accuracy of a blood glucose meter for the detection of neonatal hypoglycemia.

Authors:  Muhammad Haroon Hamid; Akmal Laeeq Chishti; Sajid Maqbool
Journal:  J Coll Physicians Surg Pak       Date:  2004-04       Impact factor: 0.711

10.  Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.

Authors:  H T Ho; W K Y Yeung; B W Y Young
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-07       Impact factor: 5.747

View more
  6 in total

Review 1.  Hypoglycemia in critically ill children.

Authors:  E Vincent S Faustino; Eliotte L Hirshberg; Clifford W Bogue
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

Review 3.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 4.  Challenges to glycemic measurement in the perioperative and critically ill patient: a review.

Authors:  Andrew D Pitkin; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

Review 5.  Hypoglycemia: When to Treat?

Authors:  Venkat Reddy Kallem; Aakash Pandita; Girish Gupta
Journal:  Clin Med Insights Pediatr       Date:  2017-12-15

Review 6.  Continuous glucose monitoring in neonates: a review.

Authors:  Christopher J D McKinlay; J Geoffrey Chase; Jennifer Dickson; Deborah L Harris; Jane M Alsweiler; Jane E Harding
Journal:  Matern Health Neonatol Perinatol       Date:  2017-10-17
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.