Accuracy of the StatStrip versus SureStep Flexx glucose meter in neonates at risk of hypoglycemia.

Ratchada Kitsommart1, Sopapan Ngerncham, Pimol Wongsiridej, Tharatip Kolatat, Kriang-Sak Jirapaet, Bosco Paes.   

Abstract

The study was performed to evaluate the accuracy of the StatStrip (SS) and SureStep Flexx (SF) glucose meters compared to plasma glucose in infants at risk for neonatal hypoglycemia and to determine the effect of bilirubin and hematocrit on the results. A prospective cross-sectional study was conducted on 172 venous blood glucose samples from infants who had initial low point-of-care (POC) glucose tests measured simultaneously by SS and SF. Plasma glucose levels were compared to both POC instruments, and the effect of bilirubin and hematocrit levels on mean glucose differences were analysed. Mean (SD) plasma glucose was 2.12 (0.45) mmol/L; (range, 1.11-3.06 mmol/L). Mean (1.96SD) glucose differences of the SS versus SF were 0.21 (0.70) mmol/L and -0.04 (0.78) mmol/L, respectively. SS sensitivity was 94.7 % with an 86.1 % negative predictive value (NPV) at 2.8 mmol/L, while the SF had a 100 % sensitivity and NPV at the same cut-off level. No correlations were identified between mean glucose differences and either hematocrit or bilirubin levels in both glucose meters. Both the SS and SF glucose meters have limited use when compared to plasma glucose. Hence, they can only be employed as screening tools in at-risk neonates with an appropriate, predetermined cut-off level. Hematocrit and bilirubin levels did not affect the accuracy of both devices.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23644649     DOI: 10.1007/s00431-013-2019-2

Source DB:  PubMed          Journal:  Eur J Pediatr        ISSN: 0340-6199            Impact factor:   3.183


  25 in total

1.  Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates.

Authors:  J Girouard; J C Forest; J Massé; M Leroux; N C Bradburn; T C Noblet; J O Joynes; J Baum
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  Incidence of neonatal hypoglycemia in babies identified as at risk.

Authors:  Deborah L Harris; Philip J Weston; Jane E Harding
Journal:  J Pediatr       Date:  2012-06-23       Impact factor: 4.406

3.  Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.

Authors:  Brad S Karon; Laurie Griesmann; Renee Scott; Sandra C Bryant; Jeffrey A Dubois; Terry L Shirey; Steven Presti; Paula J Santrach
Journal:  Diabetes Technol Ther       Date:  2008-04       Impact factor: 6.118

4.  Performance of a new interference-resistant glucose meter.

Authors:  Somlak Vanavanan; Pitak Santanirand; Upsorn Chaichanajarernkul; Anchalee Chittamma; Jeffrey A Dubois; Terry Shirey; Mark Heinz
Journal:  Clin Biochem       Date:  2009-09-19       Impact factor: 3.281

Review 5.  Postnatal glucose homeostasis in late-preterm and term infants.

Authors:  David H Adamkin
Journal:  Pediatrics       Date:  2011-02-28       Impact factor: 7.124

Review 6.  Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds.

Authors:  M Cornblath; J M Hawdon; A F Williams; A Aynsley-Green; M P Ward-Platt; R Schwartz; S C Kalhan
Journal:  Pediatrics       Date:  2000-05       Impact factor: 7.124

7.  Transient neonatal hypoglycemia--long-term effects on neurodevelopmental outcome.

Authors:  Nazan Dalgiç; Ebru Ergenekon; Sebnem Soysal; Esin Koç; Yildiz Atalay; Kivilcim Gücüyener
Journal:  J Pediatr Endocrinol Metab       Date:  2002-03       Impact factor: 1.634

Review 8.  Knowledge gaps and research needs for understanding and treating neonatal hypoglycemia: workshop report from Eunice Kennedy Shriver National Institute of Child Health and Human Development.

Authors:  William W Hay; Tonse Nk Raju; Rosemary D Higgins; Satish C Kalhan; Sherin U Devaskar
Journal:  J Pediatr       Date:  2009-11       Impact factor: 4.406

9.  Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.

Authors:  H T Ho; W K Y Yeung; B W Y Young
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-07       Impact factor: 5.747

Review 10.  Overview on self-monitoring of blood glucose.

Authors:  Martina Montagnana; Marco Caputo; Davide Giavarina; Giuseppe Lippi
Journal:  Clin Chim Acta       Date:  2009-01-09       Impact factor: 3.786

View more
  3 in total

1.  Evaluation of Blood Glucose Monitoring System in Screening for Neonatal Hypoglycemia: Tighter Accuracy Standard.

Authors:  Mei-Fen Tsao; Hui-Wen Chang; Chien-Hsi Chang; Chi-Hsuan Cheng; Hsiu-Chen Lin
Journal:  J Diabetes Sci Technol       Date:  2016-09-14

2.  The StatStrip glucose monitor is suitable for use during hyperinsulinemic euglycemic clamps in a pediatric population.

Authors:  Kara A Lindquist; Kelsey Chow; Amy West; Laura Pyle; T Scott Isbell; Melanie Cree-Green; Kristen J Nadeau
Journal:  Diabetes Technol Ther       Date:  2014-01-28       Impact factor: 6.118

3.  Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy.

Authors:  Ata Mahmoodpoor; Hadi Hamishehkar; Kamran Shadvar; Sarvin Sanaie; Afshin Iranpour; Vahid Fattahi
Journal:  Indian J Crit Care Med       Date:  2016-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.